期刊文献+

反应堆铝合金样品中^(63)Ni的测量技术研究 被引量:2

Measurement of ^(63)Ni in Aluminium Alloy Samples in Reactor
下载PDF
导出
摘要 为建立反应堆铝合金样品中63 Ni活度测量的方法,在前期确定了样品前处理流程的基础上,进行了全程化学回收率实验、去污实验、镍载体量对测量效率的影响及空白实验等方面工作。全程化学回收率在70%以上,对各种杂质离子的去污因子均为102及以上,镍载体的加入量选为5mg。该方法对DPM(放射性核素每分钟衰变数)的标准偏差为6.29,相对标准偏差为15.9%,检出限为1.38Bq/g,已应用于反应堆退役样品的分析测量。 To establish a method of measuring the activity of 63Ni in aluminium alloy samples in reactor,the chemical yield,the decontamination,the effect of Ni carrier on measuring efficiency and blank experiments were studied based on the previous confirmed pre-treatment of samples.The chemical yield of this method is beyond 70%.The decontamination factor is more than 102 for every impurity ion.The amount of the Ni carrier is 5 mg.The standard deviation is 6.29 and the relative standard deviation is 15.9%.The detection limit of the method is 1.38 Bq/g.The method has been applied to the retired samples of the reactor.
出处 《原子能科学技术》 EI CAS CSCD 北大核心 2012年第B09期153-157,共5页 Atomic Energy Science and Technology
关键词 铝合金 化学回收率 去污因子 测量效率 标准偏差 检出限 aluminium alloy chemical yield decontamination factor measure efficiency standard deviation detection limit
  • 相关文献

参考文献8

  • 1SCHEUERER C, SCHUPFNER R. A very sen sitive LSC procedure to determine Ni 63 in envi ronmental samples, steel and concrete[J]. J Ra dioanal Nucl Chem: Articles, 1995, 193(1) 127-131.
  • 2MELLADO J, TARANCON A, GARCiA J F, et al. Combination of chemical separation and data treatment for 55Fe, 63Ni, 99Tc, 137Cs and 9Sr/9Y activity determination in radioactive waste by liquid scintillation[J]. Applied Radia- tion and Isotopes, 2005, 63(2): 207-215.
  • 3翟盛庭,冯静毅,孔祥蓉,吴勋群,沈峰.反应堆构件中^(63)Ni的分离和自动测量[J].核动力工程,1995,16(1):88-93. 被引量:8
  • 4RADWAN M, PRZYBY1SKA A, MYKOWS- KA E, et al. Measurement o{ nickel-63 low ac- tivity insamples of metals and alloys with the help of liquid seintillators[J]. Int J Appl RadiatIsot, 1981, 32: 97-99.
  • 5王萍,邱永梅,吴涛,谭昭怡,但贵萍,孙宇.反应堆铝合金样品中^(63)Ni测定的前处理技术研究[J].原子能科学技术,2010,44(B09):128-133. 被引量:4
  • 6卢瑛,班莹.金属材料中^(63)Ni的测定[J].核化学与放射化学,2000,22(1):45-49. 被引量:9
  • 7HOU Xiaolin, OSTERGAARD L F, NIELSEN S P. Determination of 6aNi and 5SFe in nuclear waste samples using radiochemical separation and liquid scintillation eounting[J]. Analytica Chimi- ca Acta, 2005, 535(1 2): 297-307.
  • 8余耀仙.水中63Ni含量的测定[J].辐射防护,1988,8(2):146-150.

二级参考文献18

  • 1翟盛庭,沈峰.自动绝对测量^(99)Tc活度[J].原子能科学技术,1994,28(3):241-245. 被引量:4
  • 2翟盛庭,冯静毅,孔祥蓉,吴勋群,沈峰.反应堆构件中^(63)Ni的分离和自动测量[J].核动力工程,1995,16(1):88-93. 被引量:8
  • 3SCHEUERER C,SCHUPFNER R.A very sensitive LSC procedure to determine Ni-63 in environmental samples,steel and concrete[J].J Radioanal Nucl Chem,1995,93(1):127-131.
  • 4MELLADO J,TARANCON A.Combination of chemical separation and data treatment for 55Fe,63Ni,99Tc,137Cs and 90Sr/99Y activity determination in radioactive waste by liquid scintillation[J].Applied Radiation and Isotopes,2005,63:207-215.
  • 5HOU Xiaolin,φSTERGAARD L F,NIELSEN S P.Determination of 63Ni and 55Fe in nuclear waste samples using radiochemical separation and liquid scintillation counting[J].Analytica Chimica Acta,2005,535:297-307.
  • 6余耀仙.水中镍-63的测量.辐射防护,1988,8(2):146-150.
  • 7RADWAN M,PRZYBYLSKA A,MYKOWSKA E,et al.Measurement of nickel-63 low activity in samples of metals and alloys with the help of liquid scintillators[J].Int J Appl Radiat Isot,1981,32:97-99.
  • 8赵敏,班莹.水中铁-59的测定[R].北京:原子能出版社,1993.
  • 9翟盛庭,沈峰,楼滨乔.通用自动计数与数据处理装置及应用[J]核电子学与探测技术,1992(02).
  • 10翟盛庭,王依华.Ys-01多功能液体闪烁测量装置及其高的能量分辨率特性[J]核动力工程,1987(04).

共引文献9

同被引文献10

引证文献2

二级引证文献2

相关作者