期刊文献+

Mark V 高压注射器参数间关系探讨 被引量:1

Research on the Parameters of Mark V Injector
下载PDF
导出
摘要 Mark V高压注射器参数的正确选择是决定心血管造影成功的关键。其中流率的选择更需恰当和准确。本文主要讨论了影响流率的相关因素 ,以及流率与压力限制之间的关系。总结出了一套典型的注射参数予置值。实验表明 ,使用浓度为 75 %左右造影剂 ,压力限制设置在 2 0 6 8— 5 5 15 k This study sought to correctly preset the parameters of Mark v injection.The emphasis was laid on the flow rate in that,it ought to be set more carefully and more accurately for ensuring success in angiography.The authors analyzed the influential factors of flow rate and the relationship between flow rate and pressure limiting and hence proposed a set of typical presetting injection parameters.Experimental results demonstrated that with the contrast media concentration around 75%,high quality images could be obtained by setting the pressure limiting within 2068 5515kPa.
出处 《生物医学工程学杂志》 EI CAS CSCD 北大核心 2000年第1期114-116,共3页 Journal of Biomedical Engineering
关键词 流率 MarkV 高压注射器 心血管造影 Flow rate Pressure limiting Concentration
  • 相关文献

参考文献1

  • 1[1]MEDRAD.Mark V/Mark V Plus Injection System.Service Manual KMP,826-s,1992

引证文献1

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部