非接触性眼压计与Icare回弹式眼压计的临床应用 被引量:1

下载PDF
导出
摘要 目的比较分析非接触性眼压计与Icare回弹式眼压计在健康体检中测量结果分析。探讨非接触性眼压计与Icare回弹式眼压计在临床中的应用与性能。方法对100例(200只眼)分别用Icare回弹式眼压计与非接触性眼压计测量眼压,对所得数据进行t检验及相关分析等方法比较两种眼压计结果的一致性。结果非接触性眼压计所测眼压值为(15.8±1.8)mmHg,Icare回弹式眼压计所测眼压平均值为(15.8±1.8)mmHg,差异无统计学意义(P=0.40)。结论在正常人群健康体检及高眼压的筛查中,非接触性眼压计与Icare回弹式眼压计可互补,均值得推广。
作者 吕茜
出处 《河北医药》 CAS 2012年第18期2769-2770,共2页 Hebei Medical Journal
  • 相关文献

参考文献5

  • 1Knies T, Lin S, Choe J, et al. Clinical comparison of contourand applana- tion tonometry And their relationship to pachymetry. Arch Ophthalmol, 2005,123 : 1523-1537.
  • 2Kanngiesser HE, Knjestedt C, Robert YC. Dynamic contour tonometry : presentation of a new tonometer. J Glaucos ,2005,14:344-350.
  • 3李娜,段宣初.回弹式眼压计的临床应用价值[J].中国实用眼科杂志,2010,28(7):689-693. 被引量:4
  • 4Pakrou N, Gray T, Mills R, et al. Clinical comparison of the Icare tonome- ter and goldmann applanation tonometry. J Glaucoma,2008 17:43-47.
  • 5Poosichi A, Mitchell R, Nicholas S, et al. The icare rebound tonometer: eomparisons with goldmann tonometry, and influence of central corneal thickness. Clin Exp Ophthalmol,2009 ,37 :687-691.

二级参考文献33

  • 1乔荣华,刘丽娜,李美玉,任泽钦.Tono-Pen眼压计临床应用评价[J].中国实用眼科杂志,2004,22(12):980-982. 被引量:12
  • 2韩英军,梁天蔚,勇志鹏,张满红,李文芳,胡利华,王立华,陈建丽,王婕,孙挥宇.Tono-pen眼压计与Goldmann眼压计对穿透性小梁切除术后患者眼压测量的比较[J].国际眼科杂志,2006,6(6):1424-1426. 被引量:12
  • 3Kass MA,Heuer DK,Higginbotham E,et al.,The Ocular Hypertension Treatment Study:a randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma[J].Arch Ophthalmol,2002,120:701-713.
  • 4Wessels IF,Oh Y.Tonometer utilization,accuracy and calibration under field conditions[J].Arch Ophthalmol,1990,108(12):1709-1712.
  • 5Kass MA.Standardizing the measurement of intraocular pressure for clinical research Guidelines from the Eye Care Technology Forum[J].Ophthalmology,1996,103(1):183-185.
  • 6Kontiola AI.A new electromechanical method for measuring intraocular pressure[J].Doc Ophthalmol,1997,93(3):265-276.
  • 7Moreno-Montanes J,Garcia N,Fernandez-Hortelano A,et al.Rebound Tonometer Compared With Goldmann Tonomeler in Normal and Pathologic Corneas[J].Cornea,2007,26(4):427-430.
  • 8Pakrou N,Gray T,Mills R,et al.Clinical Comparison of the Icare Tonometer and Goldmann Applanation Tonometry[J].J Glaucoma,2008,17(1):43-47.
  • 9Queir ó s A,Gonz ález-M éijome JM,Fernandes P,et al.Technical note:a comparison of central and peripheral intraocular pressure using rebound tonometry[J].Ophthalmic Physiol Opt,2007,27(5):506-511.
  • 10Sahin A,Basmak H,Niyaz L,et al.Reproducibility and tolerability of the ICare rebound tonometer in school children[J].J Glaucoma,2007,16(2):185-188.

共引文献3

同被引文献11

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部