连续流动法测定烟草中磷酸盐含量的不确定度评定

Evaluation of Uncertainty for the Determination of Phosphate Content in Tobacco Using Continuous Flow Method
下载PDF
导出
摘要 依据测量不确定度评定方法,建立数学模型,对连续流动法测定烟草中磷酸盐含量的各测量不确定度分量进行评定,得出测量不确定度的结果为0.012%;水分测量和样品浓度测量是测量误差的最大影响因素,实验中更应重视以确保测量结果的准确可信。 Depending on the method for evaluating uncertainty, all the component uncertainty of the determination for phosphate content in tobacco using continuous flow method was evaluated according to the established mathematical model. The expanded uncertainty was 0.012%; and the measure for water and concentration of sample played on the most important influence in the measurement error, which should be paid more attention for the accuracy and credibility of the determination results.
出处 《福建分析测试》 CAS 2012年第5期33-37,共5页 Fujian Analysis & Testing
关键词 连续流动法 磷酸盐 不确定度 Continuous Flow Method Phosphate Uncertainty
  • 相关文献

参考文献6

  • 1CNAS-CL01:2006检测和校准实验室能力认可准则[s].2006.
  • 2CNAS-GL06:2006,化学分析中不确定度的评估指南[S].中国合格评定国家认可委员会,2006.
  • 3JJFl059—1999.测量不确定度评定与表示【S].北京:中国计量出版社,1999.
  • 4YC/T343-2010,烟草及烟草制品磷酸盐的测定连续流动法[S].
  • 5YC/T 31-1996.烟草及烟草制品试样的制备和水分测定烘箱法[S].
  • 6JJF(烟草)4.3-2012,烟草及烟草制品连续流动法测定常规化学成分测量不确定度评定指南第3部分:总氮[S].