期刊文献+

基于蒙特卡罗方法的6MV Truebeam剂量计算 被引量:7

Dose calculation of 6 MV Truebeam using Monte Carlo method
下载PDF
导出
摘要 使用蒙特卡罗方法快速准确地模拟6MV Varian Truebeam医用电子直线加速器射野剂量特性,探究厂家提供相位空间源的可用性及模拟方法的准确性。以Varian公司提供的出束窗口位置处相位空间文件作为Beamnrc/EGSnrc输入源,模拟射野形成结构并计算10cm×10cm射野下的均匀水体模中的剂量分布,将计算结果与相同条件下的实验测量数据进行比较。同一计算机模拟时,传统完整机头模拟方法需4~5h,文中所述方法模拟时间可缩减至48min,剂量计算结果与相同条件下所得实际测量数据可较好地吻合,两者的百分深度剂量差异低于3%,且建成区吻合良好;不同深度处80%等剂量线所包含的射野区域内百分离轴剂量比差异均低于3%,且半影区效果好。使用厂家提供相位空间文件作为EGSnrc入射源,能快速模拟治疗头,得到剂量计算结果与实际测量值的差异满足剂量计算精度要求。 The purpose of this work is to simulate 6 MV Varian Truebeam linac dosimeter characteristics using Monte Carlo method and to investigate the availability of phase space file and the accuracy of the simulation.With the phase space file at linac window supplied by Varian to be a source,the patient-dependent part was simulated.Dose distributions in a water phantom with a 10 cm×10 cm field were calculated and compared with measured data for validation.Evident time reduction was obtained from 4-5 h which a whole simulation cost on the same computer to around 48 minutes.Good agreement between simulations and measurements in water was observed.Dose differences are less than 3% for depth doses in build-up region and also for dose profiles inside the 80% field size,and the effect in penumbra is good.It demonstrate that the simulation using existing phase space file as the EGSnrc source is efficient.Dose differences between calculated data and measured data could meet the requirements for dose calculation.
作者 宋婷 周凌宏
出处 《强激光与粒子束》 EI CAS CSCD 北大核心 2012年第12期2975-2978,共4页 High Power Laser and Particle Beams
基金 国家自然科学基金项目(30970866)
关键词 Truebeam EGSNRC 剂量计算 蒙特卡罗 Truebeam EGSnrc dose calculation Monte Carlo
  • 相关文献

参考文献14

  • 1陈秋秋,鱼红亮,孙亮.肿瘤放射治疗中的剂量计算[J].中国辐射卫生,2009,18(3):376-378. 被引量:3
  • 2Verhaegen F, Seuntjens J. Monte Carlo modeling of external radiotherapy photon beamsQj]. Phys Med Biol ,2003 ,48(21) : 107-164.
  • 3Kawrakow I. Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version[J]. Med Phys ,2000, 27(3):485-498.
  • 4Rogers D W,Ma C M, Walter S B,et al. Beamnrc user manual[R]. NRCC Report PIRS-0509A, 2001.
  • 5Rogers D W, Faddegon D A,Ding G X,et al. BEAM: a Monte Carlo code to simulate radiotherapy treatment units[J]. Med Phys * 1995,22(5):503-524.
  • 6Kawrakow I,Rogers D W. The EGSnrccode system: Monte Carlo simulation of electron and photon transport[R], NRCC Report PIRS-701, 2006.
  • 7Lin S,Chu T,Lin J. Monte Carlo simulation of a clinical linear accelerator[J]. Applied Radiation and Isotopes, 2001,55(.6) : 759-765.
  • 8Walters B R, Kawrakow I,Rogers D W. DOSXYZnrc users manual[R]. NRCC Report P1R&-794 revB, 2007.
  • 9Constantin M, Perl J, LoSasso T,et al. Modeling the Truebeam linac using a CAD to Geant4 geometry implementation : dose and IAEA-compliant phase space calculations[J]. Med Phys , 2011,38(7):4018-4024.
  • 10ICRU. Prescribing, recording and reporting photon beam therapy[R]. ICRU-78,1993.

二级参考文献29

  • 1吴宜灿,李国丽,陶声祥,吴爱东,孔令玲,刘伯学,林大全,陈义学,宋钢,赵攀,林辉,陈朝斌,黄群英,吴李军.精确放射治疗系统ARTS的研究与发展[J].中国医学物理学杂志,2005,22(6):683-690. 被引量:44
  • 2周凌宏,金浩宇,陈超敏.基于蒙特卡罗模拟的快速剂量计算方法研究[J].仪器仪表学报,2007,28(4):699-702. 被引量:8
  • 3邱学军.我国肿瘤放射治疗技术、应用及产业发展展望与对策[J].中国医疗器械信息,2007,13(3):18-29. 被引量:9
  • 4朱广迎.放射肿瘤学[M].北京:科学出版社,2004:7.
  • 5王圣忠 刘守业 朱龙祥 等.早期阴茎癌的镭模放射治疗.肿瘤,1986,6(5):221-221.
  • 6ICRU,Report 24, Deternination of Absorbed Dose in a Patient Irradiated by Beams of X or Gamma Rays in Radiotherapy Procedures[ R]. International Commission on Radiation Units and Measurements. Washington DC, 1976.
  • 7Tsai JS,Wazer D E,Ling M N,et al. Dosimetric verification of the dynamic intensity - modulated radiation therapy of 92 patients[Jl. Int J Radiat Oncol Biol Phys, 1998,40:1 213 - 1 230.
  • 8Bently R E, Milan J. An interactive digital computer system for radiotherapy treatment planning[ J ]. Br J Radiol, 1971,44:307 -317.
  • 9Milan J, Se B, Benteley R E,et al. The storage and manipulation of radiation dose date in a small digital computer [ J ]. British Journal of Radiology, 1974,47 : 115 - 121.
  • 10Pascal Storchi, Evert Woudstra. Calculation models for determining the absorbed dose in water phantoms in off - axis planes of rectangular fields of open and wedged photon beam [ J ]. Phys Med Biol, 1995,40:511 - 527.

共引文献2

同被引文献49

  • 1金浩宇,周凌宏,陈光杰,陈超敏.一种基于蒙特卡罗模拟的快速剂量计算模型[J].第四军医大学学报,2006,27(21):2002-2004. 被引量:5
  • 2周凌宏,金浩宇,陈超敏.基于蒙特卡罗模拟的快速剂量计算方法研究[J].仪器仪表学报,2007,28(4):699-702. 被引量:8
  • 3Varian Medical Systems. TruebeamIM Ins-tallation Product Acceptance[M]. California,2011:3.
  • 4Fontenot JD,Alkhatib H,Garrett JA,et al.AAPM Medical Physics Practice Guideline2. a : Commissioning and quality assuranceof X-ray-based image-guided radiotherapysystems[J].J Appl Clin Med Phys,2014,15⑴:4528.
  • 5ICRU Report 24. Determination of absorbed dose in a patient irradiated by beams of X or Gamma rays in ra- diotherapy Procedures [ M ]. 1976.
  • 6Verhaegen F, Seuntjens J. Monte Carlo modeling of external Radiotherapy photon beams [ J]. Phys Med Biol. 2003,48(21) :107 - 164.
  • 7Reynaert N, Van der Marck S C, Schaart D R, et al. Monte Carlo treatment plarming for photon and elec- tron Beams [ J ]. Radiation Physics and Chemistry. 2007,76(4) :643 - 686.
  • 8Kawrakow I, Waiters BR B. Efficient photon beam dose calculations using DOSXYZnrc with BEAMnrc [J]. Medical Physics. 2006,33 (8) : 3046 - 3056.
  • 9Rogers D W, Faddegon D A, Ding G X, et al. BEAM : a Monte Carlo code to simulate radiotherapy treatment units [ J]. Medical Physics. 1995,22(5) :503 - 524.
  • 10Rogers D W O, Wahers B, Kawrakow I. BEAMnrc us- ers manual, NRCC Report PIRS - 0509 (A) revK [ R ]. Ottawa: National Research Council of Canada. 2009.

引证文献7

二级引证文献15

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部