摘要 目的评定水中总α放射性测量实验中的不确定度。方法通过分析计算公式中的有关参数,分析测量实验中的不确定度。结果该分析实验中由测量仪器带来的不确定度为10.4%,仪器探测效率带来的不确定度为6.5%、由铺样的分散性带来的不确定度为3.43%。结论水中总α放射性主要受到测量仪器的不确定度、仪器探测效率和铺样的分散性等3个方面的不确定度因素影响。 [Objective] To evaluate the uncertainty in measurement of gross α radioactivity in water.[Methods]By analyzing the relevant parameters in the formula,the uncertainty in measurement was studied.[Results]The uncertainty caused by instrument was 10.4%,the uncertainty caused by instrument detection efficiency was 6.5%,and the uncertainty caused by thickness of samples was 3.43%.[Conclusion]In measurement of gross α radioactivity in water,the uncertainties mainly include instrument uncertainty,instrument detection efficiency uncertainty,and thickness of samples uncertainty.
作者 刁春娜
出处 《职业与健康》 CAS 2012年第22期2746-2747,2750,共3页 Occupation and Health
关键词 放射性 不确定度 总α 评估 Radioactivity Uncertainty Gross α Evaluation
  • 相关文献

参考文献5

  • 1CNAS-GL06:2006化学分析中不确定度的评估指南[M].北京:中国合格评定国家认可委员会,2006.
  • 2ISOTC147,SC7N-45.分析方法的校准和评价及操作特征的计算[S].
  • 3上海市计量测试技术研究院.常用测量不确定度评定方法及应用实例[M].北京:中国计量出版社,2002.
  • 4中华人民共和国核行业标准.水中总a放射性浓度的测定厚源法[S].EJ/T1075-1998.
  • 5复旦大学 清华大学 北京大学.原子核物理实验方法[M].北京:原子能出版社,1997.93-96,109-117,301-304.