血细胞计数仪与手工HiCN法测定血红蛋白结果比较 被引量:1

A COMPARATIVE STUDY OF BOTH HEMOCYTOMETER AND ARTIFICIAL HiCN Hb DETERMINATION
下载PDF
导出
摘要 目的 :探讨仪器测定 H b结果与手工法测定 H b结果是否存在差异。方法 :采用血细胞计数仪与手工 Hi CN法分别对随机人群、健康人群、贫血病人同时进行 H b测定。结果 :两法测定 H b结果存在显著性差异 ,贫血病人组差异尤为显著 ,P<0 .0 1。结论 :血细胞计数仪测定 Hb结果稳定 ,但两法结果存在差异 ,不可忽视。 Objective: To investigate whether there are differences in the results of both hemocytometer and artificial HiCN Hb determination. Method: Simultaneous Hb determination was done both by hand and by a hemocytometer in random samples, healthy samples and anemic samples. Results: There isa significant differences between the 2 approaches. The anemic group showed greater significance than the first 2 groups(P<0.01). Conclusion: The results from hemocytometer are more reliable and stable. There isa significant difference between the two approaches that cannot be ignored.
作者 文燕
出处 《海南医学院学报》 CAS 2000年第2期88-89,共2页 Journal of Hainan Medical University
关键词 血红蛋白 高铁血红蛋白 血细胞计数 HICN法 Hemoglobin determination hemocytometry instruments and equipment methemoglobin
  • 相关文献

参考文献7

  • 1叶应妩.王毓三.全国临床检验操作规程.第2版.南京;东南大学出版社,1988.289
  • 2张建中,贺福如.氰化高铁血红蛋白测定中若干问题的探讨[J].临床检验杂志,1996,14(2):112-112. 被引量:6
  • 3周道银.酸碱对HiCN法测定Hb的影响[J].临床检验杂志,1995,13(1):41-41.
  • 4吕先航.自动血细胞分析仪各参数的临床应用[J].实用医学检验杂志,1998,5(1):1-1.
  • 5郭新荣,曹兴建.血液分析仪现场室间质量测评[J].临床检验杂志,1996,14(4):208-208. 被引量:1
  • 6顾可果.血液细胞分析仪的临床应用[J].临床检验杂志,1993,16(1):50-50.
  • 7床保军.关于血液分析仪测定RBC、Hb、HCT的评价[J].陕西医学检验,1995,10(1):26-26.

二级参考文献4