摘要
目的 :探讨仪器测定 H b结果与手工法测定 H b结果是否存在差异。方法 :采用血细胞计数仪与手工 Hi CN法分别对随机人群、健康人群、贫血病人同时进行 H b测定。结果 :两法测定 H b结果存在显著性差异 ,贫血病人组差异尤为显著 ,P<0 .0 1。结论 :血细胞计数仪测定 Hb结果稳定 ,但两法结果存在差异 ,不可忽视。
Objective: To investigate whether there are differences in the results of both hemocytometer and artificial HiCN Hb determination. Method: Simultaneous Hb determination was done both by hand and by a hemocytometer in random samples, healthy samples and anemic samples. Results: There isa significant differences between the 2 approaches. The anemic group showed greater significance than the first 2 groups(P<0.01). Conclusion: The results from hemocytometer are more reliable and stable. There isa significant difference between the two approaches that cannot be ignored.
出处
《海南医学院学报》
CAS
2000年第2期88-89,共2页
Journal of Hainan Medical University
关键词
血红蛋白
高铁血红蛋白
血细胞计数
HICN法
Hemoglobin determination
hemocytometry
instruments and equipment