期刊文献+

A light-through-side cell and co-used furnace designed for micro-Raman spectra study on volatile melts

A light-through-side cell and co-used furnace designed for micro-Raman spectra study on volatile melts
原文传递
导出
摘要 A new type of sample cell specially designed for micro-Raman spectra study on volatile melts is described. The cell is made of quartz. The lower section of the cell is circular ring-shaped and the one-end-closed hole in the cell is used for locating the heating body. The tube attached to the cell lid for placing the thermal couple is inserted into the sample, which ensures that the temperature measurement accuracy is good. The cell can be sealed for reducing the effect of the composition change caused by the in_homogeneous volatilization. During the Raman spectra scanning, the laser beam is transmitted into the cell through the side but not the top. Meanwhile, a furnace has been designed to match the cell. The heating body is closely coupled with the sample cell, and there is a hole on the side wall of the furnace for the entrance of the microlens. The assembly has performed well in the tested Raman spectrum measurement of molten NaNO3 at 602 ℃. A new type of sample cell specially designed for micro-Raman spectra study on volatile melts is described. The cell is made of quartz. The lower section of the cell is circular ring-shaped and the one-end-closed hole in the cell is used for locating the heating body. The tube attached to the cell lid for placing the thermal couple is inserted into the sample, which ensures that the temperature measurement accuracy is good. The cell can be sealed for reducing the effect of the composition change caused by the in_homogeneous volatilization. During the Raman spectra scanning, the laser beam is transmitted into the cell through the side but not the top. Meanwhile, a furnace has been designed to match the cell. The heating body is closely coupled with the sample cell, and there is a hole on the side wall of the furnace for the entrance of the microlens. The assembly has performed well in the tested Raman spectrum measurement of molten NaNO3 at 602 ℃.
出处 《Chinese Chemical Letters》 SCIE CAS CSCD 2012年第11期1291-1294,共4页 中国化学快报(英文版)
基金 supported by the National Natural Science Foundation of China(No.51004034) the Ph.D.Programs Foundation of Ministry of Education of China(No.20100042110007)
关键词 Light-through-side cell Raman spectroscopy Volatile melts Comer microlens Cell-around-heating body furnace Light-through-side cell Raman spectroscopy Volatile melts Comer microlens Cell-around-heating body furnace
  • 相关文献

参考文献10

  • 1N. Guo, G.E Zeng, S.Q. Xi, Chin. Chem. Lett. 4 (1993) 255.
  • 2S.K. Ratkje, E. Rytter, J. Phys. Chem. 78 (1974) 1499.
  • 3L.M. Toth, LP. Young, G.P. Smith, Anal. Chem. 41 (1969) 683.
  • 4J.P. Young, Anal. Chem. 36 (1964) 390.
  • 5C. Solomon, J.H. Clarke, J.O. Bockbis, J. Chem, Phys. 49 (1968) 445.
  • 6B. Gilbert, G. Mamantov, Appl. Spectrosc. 29 (1975) 276.
  • 7V. Dracopoulos, B. Gilbert, G.N. Papatheodorou, J. Chem. Soc. Faraday Trans. 94 (1998) 2601.
  • 8E. Stefanidaki, G.M. Photiadis, C.G. Kontoyannis, et al. J. Chem. Soc. Dalton Trans. 11 (2002) 2302.
  • 9R.R. Nazmutdinov, T.T. Zinkicheva, S.Y. Vassiliev, et al. Spectrochim. Acta A: Mol. Biomol. Spectrosc. 75 (2010) 1244.
  • 10J.J. George, W.J. David, J. Chem. Phys. 35 (1961) 739.

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部