期刊文献+

一种用于生物芯片阅读仪非均匀性校正的点光源

Spot light for non-uniformity correction of bio-chip reader
下载PDF
导出
摘要 介绍了一种用于生物芯片阅读仪非均匀性校正的点光源。该点光源由光源本身和定位模板组成。通过特殊的电路设计,光源能快速稳定、稳定性好;光源体积很小,通过巧妙的设计发光体LED在PCB板上的位置,配合使用上下完全对称的定位模板,实现了点光源能在视野的ROI内移动,达到非均匀性校正的目的。经过校正,生物芯片灰度值CV从7%降到了4%,证明了校正的有效性。 A spot light for non - uniformity correction of bio - chip reader is introduced. It contains a light and a positioning template. By special circuit design, the light can reach stable value quickly and the stability is good enough. It is very small. The lumi- nous element LED is skillfully designed on the proper position of PCB. With the help of the completely symmetrical positioning tem- plate, the spot light can be moved in the ROI to achieve the goal of non - uniformity correction. After correction, the CV value of the gray intensity of bio - chip can be reduced from 7% to 4%. It proves the validity of correction.
作者 张香琴
出处 《医疗装备》 2012年第12期1-4,共4页 Medical Equipment
关键词 生物芯片阅读仪 镜头 非均匀性校正 点光源 定位模板 bio -chip reader, lens, non -uniformity correction, spot light, positioning template
  • 相关文献

参考文献5

二级参考文献21

共引文献108

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部