床面高度作为摆位参数对放射治疗精确度的影响

Impact of Bed Height as Setup Parameter on Accuracy of Radiation Therapy
下载PDF
导出
摘要 目的:探究在放射治疗患者中通过床面高度行垂直摆位参数的实际临床意义。方法:截取某院2010年3月至2012年1月共68例通过予真空袋固定的体部肿瘤患者,利用EPID拍摄系统以首次摆位侧位片为标准,在放疗末段以床高、源皮距为参数行摆位及拍摄侧位片,测量图像中心点垂直位移情况。结果:源皮距组:>3 mm的中心点偏差共26例,精确度为62.61%;床高组:>3 mm中心点偏差共12例,精确度为82.35%,床高组精确度比源皮距组高出19.74个百分点,进行统计分析,P<0.05,差异具有统计学意义。结论:在胸腹部放疗患者中如将床面高度定位垂直方向摆位参数,可以减少治疗中心移位误差性,提高精准摆位,提升精确治疗质量,保证患者生命安全。 Objootive To explore the actual clinical significance of bed height as setup parameter on the radiation therapy patients. Mothods Sixty-eight cases of patients with body tumors fixed to the vacuum bag in the Affiliated Cancer Hospital of Tianjin Medical University from March 2010 to January 2012 were selected, and EPID shooting system was used with the first position of lateral radiographs as the standard. Bed height and source-skin distance were adopted as the parameters to perform setup and lateral radiography and measurement of the vertical displacement of the image center. Results In the group of source-skin distance, there were 26 ones whose centers had a displacement more than 3 mm, with an accuracy of 62.61%, and in the group of bed height, there were 12 ones whose centers had a displacement more than 3 mm, with an accuracy of 82.35%. Irs significantly different between the two groups with P less than 0.05. Conclusion To set the bed height as the posture parameter in the vertical direction can decrease the error of center displacement and increase the precision of setup, and then the therapy quality can be enhanced and the patient safety be ensured.
作者 贾国峰
出处 《医疗卫生装备》 CAS 2012年第11期124-125,共2页 Chinese Medical Equipment Journal
关键词 床面高度 放射治疗 摆位参数 bed height radiation therapy setup parameter
  • 相关文献

参考文献2

二级参考文献6

  • 1Catton C,Lebar L,Warde P,et al.Improvement in total positioning error for lateral prostate fields suing a soft immobilization device[].Radiotherapy and Oncology.1997
  • 2Mubata CD,Bidmead AM,Ellingham LM,et al.Portal imaging protocol for radical dose-escalated radiotherapy treatment of prostate cancer[].International Journal of Radiation Oncology Biology Physics.1998
  • 3Bijhold J,Lebesque JV,Hart AAM,et al.Maximizing set-up accuracy using portal images as applied to a conformal boost technique for prostate cancer[].Radiotherapy and Oncology.1992
  • 4Bentel GC,Marks LB,Sherouse GW,et al.The effectiveness of immobilization during prostate irradiation[].International Journal of Radiation Oncology Biology Physics.1995
  • 5Bel A,Vos PH,Rodrigus PTR,et al.High-precision prostate cancer irradiation by clinical application of an off-line patient set-up verification procedure using portal imaging[].International Journal of Radiation Oncology Biology Physics.1996
  • 6de Neve W,van del Heuvel F,Coghe M,et al.Interactive use of online portal imaging in pelvic radiation[].International Journal of Radiation Oncology Biology Physics.1993

共引文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部