摘要 目的:通过全血标本的检测,对SYSMEX XE-5000、SYSMEX XE-2100、BEKMAN-COUL-TER AC.T 5diff血细胞仪的主要性能进行分析。方法:以卫生部提供的临检室血细胞质量评比液和我院随机病人抗凝全血标本,按说明操作SYSMEXXE-5000、XE-2100、BEKMAN-COULTERAC.T 5diff血细胞仪,进行标本检测,测定白细胞(WBC)、红细胞(RBC)、血红蛋白(HBG)、血小板(PLT)总变异系数、准确性。结果:三种仪器的WBC、RBC、HBG的总变异系数相近,随机病人检测结果统计学分析表明,差异没有统计学意义(P>0.05);结论:SYSMEX XE-5000、XE-2100、BEKMAN-COULTER AC.T 5diff血细胞分析仪均可运用于临床检测中。 Objective:To evaluate primarily SYSMEXXE-2100、SYSMEX XE-2100 and BEKMAN-COULTER AC.T 5diff hematology analyzers by anticoagulation blood samples from randomized patients.Methods:To measure total coefficient of variation and the precision of WBC、RBC、HBG and PLT according to these instructions by quality appraisal specimens from Clinical laboratory of The Ministry of Health and anticoagulation blood samples from randomized patients.Results: The statistics showed these results of randomized patients are not different from three kinds of hematology analyzers.Meanwhile total coefficient of variation of WBC、RBC and HBG are approximated to.Conclusion:SYSMEX XE-2100、ABBOTT CD-3200 and BEKMAN-COULTER AC.T 5diff hematology analyzers are suit for applying in clinical laboratory.
出处 《内蒙古医学杂志》 2012年第9期1033-1035,共3页 Inner Mongolia Medical Journal
关键词 血细胞分析仪 使用评价 总变异系数 Hematology analyzer Total coefficient of variation Evaluation
  • 相关文献

参考文献3

二级参考文献12

  • 1张广慧,赵霞,沈立松.儿科患儿使用CD-1700血液分析仪检测中间型细胞镜检的探讨[J].检验医学,2006,21(4):389-391. 被引量:2
  • 2姜波,吴红,陈世锋,黄万里,林少霞,韩英杰.全自动血液分析仪异常报警信息的分析及临床应用[J].中华检验医学杂志,2006,29(11):1013-1016. 被引量:39
  • 3王羽.全国临床检验操作规程.3版.南京:东南大学出版社,2006:11.
  • 4EP5-A2 Evaluation of precision performance of quantitative measurement methods; Approved guideline-Second edition. CLSI, 2004.24 ( 25 ).
  • 5Bull BS, Fujimoto K, Houwen B,et al. International Council for Standardization in Haematology (ICSH) recommendations for "surrogate reference" method for the packed cell volume. Lab Hematol,2003,9 : 1-9.
  • 6EP6-A Evaluation of the linearity of quantitative measurement procedures: a statistical approach; Approved guideline. CLSI, 2003.23 ( 16 ).
  • 7EP9-A2 Method comparison and bias estimation using patient samples; Approved guideline-Second edition. CLSI, 2002. 22 (19).
  • 8H20-A2 Reference leukocyte ( WBC ) differential count (proportional) and evaluation of instrumental methods;Appreved standard-Second edition. CLSI,2007.27 (4).
  • 9Szczepanski T, Orfao A, van der Velden VH, et al. Minimal residual disease in leukaemia patients. Lancet Oncol, 2001,2 : 409-417.
  • 10徐薇,倪丽芬.急性白血病的诊断与治疗近况[J].宁波医学,1999,11(3):138-139. 被引量:4