不同CT扫描条件模拟定位对放射治疗计划的影响 被引量:6

Evaluation of scanning parameters in CT simulation on radiotherapy planning
原文传递
导出
摘要 目的:利用不均匀等效模体,探讨不同CT扫描条件对CT值及照射跳数(MU)的影响。方法:采用大小2种CT孔径(80和70cm)、2种模体几何摆放顺序及2种扫描电压(120和140KV),测量不同组合,比较各自的CT值,建立相应的CT-电子密度(ED)转换曲线,选取盆腔、胸部、头颈部各10例患者的CT图像模拟适形计划(CRT)和调强计划(IMRT),分析照射MU数值的偏差。结果:对于小孔径CT,无论扫描电压、模体几何位置如何变化,其MU数值相差均≤0.1%;大孔径CT扫描电压改变对MU值无影响,模体几何摆放顺序的改变有影响,但<0.3%;CRT计划和IMRT计划各自偏差值相同。结论:对于精确放疗计划系统,CT扫描条件改变和测量模体位置改变均可引起照射MU数值误差。若使用大孔径CT进行模拟定位,需根据其特性建立合理的电子密度曲线,并应用在计划制定中。 OBJECTIVE:To study the influence of different CT scanning conditions on the CT value and radiation do ses by using heterogeneous equivalent phantoms. METHODS: Two kinds of CT bore (80 cm and 70 cm),geometrical ar rangements of phantom and scanning voltage (120 KV and 140 KV) were adopted respectively. The CT values of different combinations were measured and compared respectively. The CT-electron density (CT-ED) conversion curve was estab- lished. The CT images from thirty patients composed of ten cases of pelvic cavity, ten cases of thorax and ten cases of head and neck region respectively were used tO make the CRT treatment planning and the IMRT planning respectively. The MU deviations of the treatment plannings were analyzed. RESULTS: For the smaller bore CT with any changes of scanning voltage and geometrical arrangements of phantom,the MU deviations of was less than 0.1%. For the larger bore CT, the change of scanning voltage had not influence on the MU value,the arrangements of phantom affected the MU value and the MU deviations was less than 0.3 %. Both the CRT treatment planning and the IMRT planning had the same MU devi ations. CONCLUSIONS: For the accurate radiotherapy planning system, the radiation MU value deviation may be caused by the CT-ED conversion which is caused by the changes of CT scanning condition and measuring phantom placement. In the simulation with big core CT,comprehensive ED curves should be built in TPS.
出处 《中华肿瘤防治杂志》 CAS 北大核心 2012年第20期1579-1581,共3页 Chinese Journal of Cancer Prevention and Treatment
关键词 放射疗法 计算机辅助 体层摄影术 X线计算机 立体定位技术 radiotherapy, computer-assisted tomography, X-ray computed stereotaxic techniques
  • 相关文献

参考文献9

  • 1Constantinou C, Harrington JC, DeWerd LA. An electron density calibration phantom for CT-based treatment planning computers [J]. Med Phys,1992,19(2) :325-327.
  • 2张富利,陈建平,郑明民.PHILIPS大孔径CT模拟机层敏感度曲线的测试研究[J].中国医疗设备,2008,23(8):132-134. 被引量:3
  • 3祁振宇,黄劭敏,邓小武.放疗计划CT值的校准检测及其影响因素分析[J].癌症,2006,25(1):110-114. 被引量:53
  • 4陈立新,李文杰,黄晓延,卢杰,钱剑扬.治疗计划系统中CT值及图像质量保证的若干问题讨论[J].中国医学物理学杂志,2003,20(4):193-196. 被引量:12
  • 5Guan H, Yin FF,Kim JH. Accuracy of inhomogeneity correction in photon radiotherapy from CT scans with different settings[J]. Phys Med Biol,2002,47(17) :223-231.
  • 6Ahmad N, Belal M, Nada T, et al. Influence of electron density spatial distribution and X-ray quality during CT simulation of dose calculation accuracy[J].J Applied Clinical Medical Physics, 2011,12(3) :80-89.
  • 7Cozzi L, Fogliata A, Buffa F. Dosimetric impact of computed tomo- graphy calibration on a commercial treatment planning system for ex- ternal radiation therapy[J]. Radiother Onco1,1998,48(3) :335-338.
  • 8Skrzynski W, Zielinska S, Wachowicz M, et al. Computed tomo- graphy as a sourse of electron density information for radiation treatment planning[J]. Strahlenther onkol, 2010, 186 (6) : 327-322.
  • 9邓小武,黄劭敏,祁振宇.CT模拟机的质量控制和质量保证检验[J].中国肿瘤,2004,13(9):546-550. 被引量:33

二级参考文献36

  • 1邓小武,黄劭敏,祁振宇.CT模拟机的质量控制和质量保证检验[J].中国肿瘤,2004,13(9):546-550. 被引量:33
  • 2胡丽霞,李先明,吴冬,何慧玲,徐坚民.多层螺旋CT模拟定位技术在调强放疗计划设计中的应用[J].医学影像学杂志,2006,16(2):159-161. 被引量:7
  • 3刘传亚,秦维昌,王巍,卢传友.64层螺旋CT的层敏感度曲线测试[J].中华放射学杂志,2006,40(11):1210-1212. 被引量:11
  • 4[1]AAPM.AAPM report No.83.Quality assurance for computed-tomography simulators and the computed-tomographysimulation process[J].Med Phys,2003,30:2762-2792.
  • 5[2]Curry TS,Dawdey JE,Murry RC.Christensen's physics of diagnostic radiology [M].4th ed.PA,Malvern:Lea & Febiger,1990.
  • 6[3]Cozzi L,Fogliata A,Buffa F.Dosimetric impact of computed tomography calibration on a commercial treatment planning system for external radiation therapy [J].Radiother Oncol,1998,48:335-338.
  • 7[4]Williams TR,Thwaites TI.Radiotherapy physics in practice [M].2nd ed.Oxford:Oxford University press,2000.118-149.
  • 8[5]Constantinouc,Harrington JC,DeWerd LA.An electron density calibration phantom for CT-based treatment planning computers [J].Med Phys,1992,19(2):325-332.
  • 9[6]Cozzi L,Fogliata A,Buffa F,et al.Dosimetric impact of computed tomography calibration on a commercial treatment planning system for external radiation therapy [J].Radiother Oncol,1998,48:335-338.
  • 10[7]Beneventis,Chionne F,Gobbi G,et al.Quantitative CT to mography for radiotherapy treatment planning:calibration phantom and sources of error[J].Radiother Oncol,1995,37 (Suppl):41.

共引文献82

同被引文献58

引证文献6

二级引证文献17

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部