摘要 目的观察家庭监测(HM)功能对安置起搏器患者监测报警事件的应用。方法植入带有HM系统的双腔或三腔起搏器患者28例,随访15.71±9.73个月,记录系统传送的事件及返院随访的情况。结果 17例(61%)患者传送了报警事件。临床心血管事件报警8例,其中心房颤动报警最为多见,较常规随访提早发现32.2±19.0天。起搏系统报警事件14例(34次),其中以P波感知不良报警次数最多,为17次,较常规随访提早发现42.5±25.3天。结论 HM系统可在下一次常规随访前提早发现报警事件,有助于起搏器术后患者的实时监测和随访。 Objective To observe the monitoring of the warning events in patients after implantation of pacemaker with home-monitoring(HM) function. Methods Twenty-eight patients underwent dual chamber or three chamber pacemakers with HM was included. During a mean follow-up of 15.71 ± 9.73 months, events delivered by HM and the follow-up initia- ted by that was recorded. Result Seventeen(61% ) patients had warning events,8 of which were clinical cardiovascular warning events. It turned out that atrial fibrillation(AF) events were the most clinical warning events. AF events were de- tected 32.2 ± 19.0 days before the next conventional follow-up. 14(34 times) patients were found with pacing system e- vents in which P wave sensitivity disability played a main role( 17 times). It was detected 42.5± 25.3 days before the next conventional follow-up. Conclusion HM system is able to detect warning events at an earlier time before the next con- ventional follow-up and beneficial to the real-time monitoring as well as a more efficient follow-up.
出处 《中国心脏起搏与心电生理杂志》 2012年第6期492-494,共3页 Chinese Journal of Cardiac Pacing and Electrophysiology
关键词 心血管病学 起搏 家庭监测系统 随访 远程监控 Cardiology Pace Home-monitoring Follow-up Remote monitoring
  • 相关文献

参考文献10

  • 1Jung W, Rillig A, Birkemeyer R,et al. Advances in remote monito- ring of implantable pacemakers, cardioverter defibrillators and cardi- ac resynchronization therapy systems[ J]. J Interv Card Electrophys- iol, 2008,23( 1 ) :73.
  • 2Calcagnini G, Censi F, Floris M,et al. Evaluation of electromagnet- ic interference of GSM mobile phones with pacemakers featuring re- mote monitoring functions [ J ]. Pacing Clin Electrophysiol, 2006,29 (4) :380.
  • 3吴文利,梅霞,李正恭,郑向清,张源,宋波.具备家庭监测功能起搏器临床应用[J].中国心脏起搏与心电生理杂志,2011,25(6):493-495. 被引量:4
  • 4Kossaify A, Zoghbi S, Milliez P. Assessment of ventricular pacing in the setting of an institutional improvement program: insights into physiological pacing[ J]. Clin Med Insights Cardiol,2012 ,6 :79.
  • 5Varma N. Rationale and design of a prospective study of the efficacy of a remote monitoring system used in implantable cardioverter defib- rillator follow-up: the Lumos-T Reduces Routine Office Device Fol- low-Up Study (TRUST) study[J]. Am Heart J, 2007,154(6) :1 029.
  • 6Lazarus A. Remote, wireless, ambulatory monitoring of implantable pacemakers, cardioverter defibrillators, and cardiac resynchroniza- tion therapy systems : analysis of a worldwide database [ J ]. Pacing Clin Electrophysiol, 2007,30( Suppl 1 ) :S2.
  • 7McAlister FA, Ezekowitz J, Hooton N, et al. Cardiac resynchroniza- tion therapy for patients with left ventricular systolic dysfunction: a systematic review[ J]. JAMA, 2007,297 (22) :2 502.
  • 8Ricci RP, Morichelli L, Santini M. Home monitoring remote control of pacemaker and implantable cardioverter defibrillator patients in clinical practice: impact on medical management and health-care re- source utilization [ J ]. EuroDace .2008.10 ( 2 ) : 164.
  • 9Elsner C, Sommer P, Piorkowski C,et al. A prospective multicenter comparison trim of Home Monitoring against regular follow-up in MADIT II patients: additional visits and cost impact[J]. Computers in Cardiology, 2006,33, 241.
  • 10Nielsen JC, Kottkamp H, Zabel M, et al. Automatic home monito- ring of implantable cardioverter defibrillators [ J 1. Europace, 2008, 10(6) :729.

二级参考文献1

  • 1Antzelevitch C. From Cell to Bedside[ J ]. Curt Probl Cardiol, 2005, 30(9) :34.

共引文献3

同被引文献12

  • 1赵龙.一种具有实时监测功能的电动助力转向试验系统的研究[D].重庆大学,2012.
  • 2王振东,凌峰,于忠,等.起搏器预防病窦综合征患者阵发性房颤[C]//浙江省医学会心电生理与起搏分会.2006年浙江省心电生理与起搏学术年会论文汇编[G].浙江省医学会心电生理与起搏分会,2006:2.
  • 3王宇星.慢性心力衰竭患者心脏再同步治疗临床获益与左心室起搏位点的相关性研究[D].浙江大学,2013.
  • 4孙欣.综合指标提高超声对心脏再同步化治疗的预测价值[D].北京协和医学院,2013.
  • 5Jung W , Rillig A, Birkemeyer R,et al. Advances in remote monito- ring of implantable pacemakers,cardioverter defibrillators and cardi- aeresynchmnization therapy systems[ J]. J lnterv Card Electrophysi- ol,2008,23( 1 ) :73.
  • 6Nielsen JC,Kottkmnp H,Zabel M,et al. Automatic home monitoring of implmltable eardioverter defibrillators [ J ]. Europace,2008,10 ( 6 ) :729.
  • 7庄颖珠,晋群,张吉新.CRT/CRTD治疗心力衰竭的研究进展[J].医学综述,2011,17(19):2964-2967. 被引量:5
  • 8戴研,杨杰孚,周玉杰,梁兆光,李康,丁燕生,张澍.家庭监测系统在双腔起搏器植入患者中的多中心注册研究[J].中国循环杂志,2013,28(1):29-32. 被引量:24
  • 9宿燕岗.植入性心脏电子装置术后心房颤动的一些临床问题[J].中国心脏起搏与心电生理杂志,2013,27(5):377-380. 被引量:2
  • 10解玉泉,李毅刚.家庭监测系统在心脏电子植入装置患者随访中的应用[J].中国介入心脏病学杂志,2015,23(8):471-473. 被引量:5

引证文献3

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部