多排螺旋CT自适应统计迭代重建算法的客观评价 被引量:5

Evaluation of adaptive statistical iterative reconstruction algorithm filter on image quality in multislice CT scanning
下载PDF
导出
摘要 目的:对GE公司最新研发的自适应统计迭代重建算法(ASIR)进行物理测试和体模验证。方法:采用国际通用CT成像测试体模扫描图像,测量不同算法重建图像的图像噪声、低对比分辨力(CNR)和系统调制传递函数(MTF)。结果:采用ASIR重建算法可以明显降低图像噪声,提高CNR而MTF保持不变(t=-1.706,P>0.05)。结论:ASIR算法可以在显著降低图像噪声的同时保证图像空间分辨力不变,在提高CT图像质量和降低辐射剂量方面具有较高的临床应用价值。 Objective:To evaluate the effect of adaptive statistical iterative reconstruction algorithm(ASIR)on improving image quality of multislice CT.Methods:Catphan phantom (Phantom Lab,USA)was examined on Discovery CT750 HD (GE Healthcare,Waukesha,USA).Noise,CNR and MTF in the images processed by ASIR were compared with the data without using ASIR Results:MTF did not change significantly by the application of 3D ORA filtering (t=-1.706,P0.05) while the noise was reduced.CNR was increased by using ASIR.Conclusion:The ASIR shows high potential to reduce image noise without effecting image quality attributes.By applying this approach,the same image quality can be achieved whilst gaining a significant dose reduction.
出处 《中国医学装备》 2013年第1期2-4,共3页 China Medical Equipment
基金 国家自然科学基金(30870751) 国际原子能机构(IAEA)合作研究项目(CRPNO.16155/R1) 北京市卫生系统高层次卫生技术人才培养计划(2009-3-57)"多排螺旋CT临床成像质量与受检者受照剂量优化匹配及评价方法研究"
关键词 图像质量 螺旋CT 重建算法 Image quality Spiral CT Reconstruction algorithm
  • 相关文献

参考文献10

  • 1Singh S,Kalra MK,Hsieh J. Abdominal CT comparison of adaptive statistical iterative and filtered back projection reconstruction techniques[J].Radiology,2010,(02):373-383.
  • 2Miéville FA,Gudinchet F,Brunelle F. Iterative reconstruction methods in two different MDCT scanners:Physical metrics and 4-alternative forced-choice detectability experiments–A phantom approach[J].Physica Medica,2012.
  • 3Alvin C,Silva1,Holly J. Innovations in CT Dose Reduction Strategy:Application of the Adaptive Statistical Iterative Reconstruction Algorithm[J].American Journal of Roentgenology,2010,(01):191-199.
  • 4Prakash P,Kalra MK,Ackman JB. Diffuse lung disease:CT of the chest with adaptive statistical iterative reconstruction technique[J].Radiology,2010,(01):261-269.
  • 5Park EA,Lee W,Kim KW. Itera-tive reconstruction of dual-source coronary CT angiography:assessment of image quality and radiation dose[J].International Journal of Cardiovascular Imaging,2012,(07):1775-1786.
  • 6胡熙芳,刘文涛.数字化摄影(DR)图像质量控制[J].中国医学装备,2009,6(7):39-40. 被引量:12
  • 7周小兵,王德杭.CT低剂量扫描的探讨[J].中国医学装备,2011,8(7):36-38. 被引量:10
  • 8白玫,彭明辰.X线CT成像质量评估中的调制传递函数的测试与分析[J].医疗设备信息,2006,21(5):11-13. 被引量:15
  • 9Verdun FR,Denys A,Valley JF. Detection of low-contrast objects:experimental comparison of single–and multi–detector row CT with a phantom[J].Radiology,2002,(02):426-431.
  • 10Efstathopoulos EP,Costaridou L,Kocsis O. A protocol-based evaluation of medical image digitizers[J].British Journal of Radiology,2001,(885):841-846.

二级参考文献15

  • 1唐文祥,朱志贤,郑钧正,韩发明,何韦川.X射线CT所致受检者辐射剂量探讨[J].中华放射医学与防护杂志,2001,21(S1):85-87. 被引量:7
  • 2朱昆明,刘培进.数字摄影DR应用探讨[J].医用放射技术杂志,2004(11):45-45. 被引量:7
  • 3黄少英.全数字化X线影像技术在放射科的应用价值[J].中国基层医药,2005,12(1):72-73. 被引量:14
  • 4杨燕敏,曹厚德.CT检查中辐射剂量的研究[J].医学影像学杂志,2006,16(8):865-867. 被引量:12
  • 5Goodenough D,Atkins F,Dyer S.Automated techniques for quality assurance of radiological image modalities.SPIE.Image Capture,Formatting,and Display,1991,1444:87-99.
  • 6Dainty JC,Shaw R.Image Science[M].New York:Academic Press,1974.
  • 7Goodenough D,Atkins F.Theoretical and practical aspects of automated quality assurance approaches,particularly for CT[C].Proceedings of the ICR 89,editors Silve,et.Al,1990.
  • 8Fujita H,Tsai D-Y,Itoh T,et al.A simple method for determining the modulation transfer function in digital radiography[J].IEEE Transactions on Medical Imaging,1992,11:34-39.
  • 9Cunningham IA,Reid BK.Signal and noise in modulation transfer function determinations using the slit,wire,and edge techniques[J].Medical Physics,1992,19:1037-1044.
  • 10Samei E,Flynn MJ,Reimann DA.A method for measurement the presampled MTF of digital radiographic systems using an edge test device[J].Medical Physics,1998,25:102-113.

共引文献34

同被引文献48

  • 1Brenne D J,Hall E J.Computed tomography:an increasing source of radiation exposure[J].N Engl J Med,2007,357(22):2 277-2 284.
  • 2Fuchs T A,Fiechter M,Gebhard C,et al.CT coronary angiography:impact of adapted statistical iterativerecon-struction(ASi R)on coronary stenosis and plaque compo-sition analysis[J].Int J Cardiovasc Imaging,2013,29:719-724.
  • 3Marin D,Nelson R C,Schindera S T,et al.Low-tube-voltage,high-tube-current multidetector abdominal CT:improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm-initial clinical experience[J].Radiology,2010,254:145-153.
  • 4Mulkens T H,Bellinck P,Baeyaert M,et al.Use of an automatic exposure control mechanism for dose optimization in multidetector row CT examinations:clinical evaluation[J].Radiology,2005,237:213-223.
  • 5Lee C H,Goo J M,Ye H J,et al.Radiation dose modulation techniques in the multidetector CT era:from basics to practice[J].Radio Graphics,2008,28(5):1 451-1 459.
  • 6Matsuda I,Hanaoka S,Akahane M,et al.Adaptive statistical iterative reconstruction for volume-rendered computed tomography portovenography:improvement of image quality[J].Jpn J Radiol,2010,28:700-706.
  • 7Silva A C,Lawder H J,Hara A,et al.Innovations in CT dose reduction strategy:application of the adaptive statistical iterative reconstruction algorithm[J].AJR Am J Roentgenol,2010,194(1):191-199.
  • 8王革,俞恒永,勃鲁努.德.曼,庄天戈,赵俊.X线CT研究与发展之展望[J].中国医疗器械杂志,2008,32(3):157-169. 被引量:42
  • 9孟悛非,范淼.重视CT低剂量研究[J].中华放射学杂志,2009,43(7):679-680. 被引量:48
  • 10周旭辉,彭振鹏,郑丽丽,李树荣,杨智云,孟悛非,陈星.64层螺旋CT肺动脉成像低管电压设置结合个体化对比剂应用的对照研究[J].中华放射学杂志,2009,43(7):704-707. 被引量:48

引证文献5

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部