摘要
目的通过不同检测系统分析同一批患者标本,分析检测结果在不同线性范围内的相关性及临床符合情况。方法在罗氏cobas e411与贝克曼DXI 800上检测40例不同浓度水平患者标本,进行线性回归分析。结果人绒毛膜促性腺激素(β-HCG)在罗氏cobas e411线性范围内与贝克曼DXI 800线性范围内相关系数分别为0.998 3和0.995 5,两个检测系统检测结果临床符合率为100%。结论β-HCG在罗氏cobas e411与贝克曼DXI800上检测结果具有良好的相关性,临床诊断符合性一致。
Objective To discuss correlation between different analysis systems and clinical compatibility by examination of 40 patients′ specimens at different concentrations.Methods 40 patients′ samples were detected between Roche cobas e411 and Beckman DXI 800,results were statistically analyzed referring to EP9-2A.Results The corresponding correlation coefficient in Roche and DXI 800 linearity range were 0.998 3 and 0.995 5,results between two systems completely confirmed to clinical diagnostics.Conclusion β-HCG testing results have a good correlation and clinical compatibility between Roche cobas e411 and Beckman DXI 800 systems.
出处
《检验医学与临床》
CAS
2013年第2期165-166,共2页
Laboratory Medicine and Clinic