摘要 目的建立用新鲜全血校准血细胞分析仪的方法,保证临床标本在不同型号的血细胞分析仪上检测结果的可比性、一致性。方法以校准好的BC-5500为参考机,用新鲜全血在参考机BC-5500和BC-5300、BF-5180、BF-6800之间进行校准传递。结果校准后各仪器之间的偏差可以接受,同一标本在各仪器上的结果具有可比性。结论用参比血细胞分析仪定值新鲜全血的方法可用于同室多台血细胞分析仪的校准及比对,既保证了各仪器结果的精密度及准确度,又降低了成本。
作者 李建华
出处 《检验医学与临床》 CAS 2013年第1期91-92,共2页 Laboratory Medicine and Clinic
  • 相关文献

参考文献4

二级参考文献16

  • 1[1]International Committee for Standardization in Hematology;Expert p anel on cytometry.Reference method for the enumeration of erythrocytes and leuk ocytes[J]. Clin Lab Heamat,1994,16∶131-138.
  • 2[2]International Committee for Standardization in Hematology;Expert panel o n hemoglobinometry.Recommendations for reference method for hemoglobinometry in human blood(ICSH standard 1986)and specifications for international hemoglobincy anide reference preparation 3rd[J].Clin Lab Haematol,1987,9∶73-79.
  • 3[3]NCCLS.Precedure for Determining Packed Cell Volume by the Microhematioc r it Method-Second Edition;Approved Standard[S].NCCLS Document H7-A2, Vol.1 3 No 9,1993,13(9).
  • 4Peng Mingting,Syst J Int,1998年,8卷,79页
  • 5ICSH Cytometry Panel.Platelet counting by RBC/Platelet ratio method.Am J Clin Pathol,2001,115:460-464.
  • 6ICSH Cytometry Panel.Recommendation for reference method for packed cell volume,Lab Hematol,2001,7:148-170.
  • 7ICSH.Reference method for the enumeration of erythrocytes and leukocytes.Clin Lab Haematol,16,3l-138.1994
  • 8ICSH HGB Expert Panel.Recommendations for reference method for haemoglobinometry in human blood and specifications for international haemiglobincyanide standard.J of Clin Pathol,49:271-274.1996
  • 9彭明婷,申子瑜,谷小林,等.红细胞和白细胞计数参考方法.WS/T,244-2005.
  • 10彭明婷,申子瑜,谷小林,等.血小板计数参考方法.WS/T,245-2005.

共引文献139

同被引文献38

  • 1Cordiano V,Miserocchi F,Storti M.Bone marrow metastases from anaplastic oligodendroglioma presenting with pancytopenia and hy- pogammaglobulinemia : a case report [J].Tumori, 2011,97 (6) : 808 - $11.
  • 2中国合格评定国家认可中心.《医学实验室质量和能力认可准则》[S].中国,IS015189:2007.
  • 3National Committee for Clinical Laboratory Standards.EP9-A2 Method Comparison and Bias Estimation UsingPatient Samples[S]. Approved Guideline,2002.
  • 40;Sullivan MB. Quality control of multichannel haematol-ogy analysers: critique of current methods and the needfor performance goals [J]. Clin Lab Haematol, 1990,12(1):3-12.
  • 5The National Committee for Clinical Lab- oratory Standards. Method comparison and bias estimation using patient sampale [S]. Approved Guideline, Second Edition, NCCLS,EP9-A2,2002.
  • 6唐吉斌,王京荣,黄明坤,等.不同血细胞分析仪检测指标评价分析[J].医学检验与临床,2008,19(6) :33-34.
  • 7袁静芝,康晓波,吴霞.同一实验室内多台血细胞分析仪的比对分析[J].医药前沿,2013(1) :120.
  • 8The National Committee for Clinical Laboratory Stand- ards. EP9-A2 Method comparison and bias estimation u- sing patient sample[S]. 2nd Edition, Wayne, PA, USA: NCCLS, 2002.
  • 9International Organization for Standardization. ISO15189 Medical laboratories-particular requirements for quality and eompetenee[S]. Geneva ISO, 2003.
  • 10中国合格评定国家认可中心.ISO15189医学实验室一质量和能力的特殊要求[s].北京:CNAS,2003.

引证文献5

二级引证文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部