日志文件在静态调强计划MLC到位精度检测中的应用 被引量:2

Application of Log File in Detection of Position Accuracy of Multi-leaf Collimator with Static IMRT
下载PDF
导出
摘要 目的利用软件分析Dynalog日志文件,检测静态调强计划(SIMRT)的多叶光栅(MLC)到位精度。方法由于每执行一个SIMRT会产生多个日志文件,而Varian DynalogFileViewer(DFV,Version 7.0)软件不具有批量处理功能,并且它的分析结果与本院其他日常检测结果相矛盾,为此,自行开发软件VarianMLC-DlgFile(VMD),用于分析日志文件,检测SIMRT的MLC到位精度。结果应用VMD软件分析81个SIMRT连续执行3天的MLC到位精度,结果有80个计划的MLC到位偏差符合限定标准(<1 mm),仅有1个计划的MLC到位偏差>1 mm、且<2 mm,此计划的A组第25个叶片出现偏差>1 mm的次数最多;DFV软件的统计结果综合了机器出束和出束结束瞬间两种状态的MLC到位精度,故给出了错误的MLC到位偏差数据。结论 VMD软件能区分对待机器出束和出束结束瞬间两种状态的MLC到位精度,其分析结果可指导物理师和工程师有针对性地开展质量管理(QA&QC)工作。 Objective Using a software to analyze Dynalog log files, and detect the position accuracy of multileaf collimator (MLC) with static IMRT (SIMRT). Methods Because a lot of log files would produced after one SIMRT is done, and Varian DynalogFileViewer (DFV, Version7.0) could not finish batch process, besides, the analyzed results of DFV dose were not agree with the detected daily results, so Varian MLC-DlgFile (VMD) is designed, then VMD is applied to analyze the MLC position accuracy of 81 SIMRT, which were continuously done in 3 days. Results In VMD, the MLC position errors of 80 SIMRT are all 〈1 mm, one SIMRT is 〉1 mm, 〈 2 mm, and the 25th blade of A group in this SIMRT takes most errors of〉l mm. The statistical results of DFV integrate the position accuracy of MLC in two conditions of bearn on and beam end, and get the wrong position errors of MLC. Conclusion VMD could separate beam on from beam end, so its result would be helpful for the QA&QC work of physicists and engineers.
作者 余辉 张书旭
出处 《中国医疗设备》 2013年第1期24-26,54,共4页 China Medical Devices
基金 广东省科技计划项目(2010B031600146)
关键词 直线加速器 多叶光栅 日志文件 放射治疗 静态调强计划 linear accelerator multi-leaf collimator log file radiotherapy static IMRT
  • 相关文献

参考文献12

二级参考文献40

  • 1Klein EE, Tepper J, Sontag M, et al. Technology assessment of muhileaf collimation: A North American users survey. Int .J Radiat Oncol Bio1,1999,44(3) :705-710.
  • 2Pasquino M, Borca VC, Catuzzo P, et al. Transmission, penumbra and leaf positional accuracy in commissioning and quality assurance program of a multileaf collimator for step-and-shoot IMRT treatments. Tumori,2006,92(6) :511-516.
  • 3Sun Nuclear Corporation. MapCheck Model 1175 Manual. Melbourne (FL) : Sun Nuclear Corporation. 2004.
  • 4Yang Y, Xing L. Quantitative measurement of MLC leaf displacements using an electronic portal image device. Phys Med Biol,2004,49(8) : 1521-1533.
  • 5Jursinic P, Kainz K, Ahunbay E, et al. Use of a 2D array of diodes to test the accuracy of MLC leaf position and gap width. Med Phys, 2006,33(6) : 2147.
  • 6Yang Y, Xing L. Using the volumetric effect of a finite-sized detector for routine quality assurance of multileaf collimator leaf positioning. Med Phys,2003,30(3) :433-441.
  • 7Mohammadi M, Bezak E. Evaluation of MLC leaf positioning using a scanning liquid ionization chamber EPID. Phys Med Biol, 2007,52 (1) :21-33.
  • 8Chang J, Obcemea CH, Sillanpaa J, et al. Use of EPID for leaf position accuracy QA of dynamic multi-leaf collimator (DMLC) treatment. Med Phys, 2004,31 (7) : 2091-2096.
  • 9Woo MK, Nico A. Impact of muhileaf collimator leaf positioning accuracy on intensity modulation radiation therapy quality assurance ion chamber measurements. Med Phys,2005,32(5) :1440-1445.
  • 10Ezzell G A,Galvin J M, Low D,et al.Guidance document on delivery,treatment planning, and clinical implementation of IMRT:report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee[J].Med.Phys,2003, 30:2089-2115.

共引文献36

同被引文献18

  • 1胡逸民 张红志 戴建荣.肿瘤放射物理学[M].北京:原子能出版社,1996.614.
  • 2中华人民共和国卫生部.GBZ126-2011电子加速器放射治疗放射防护要求[S].北京:中国质检出版社,2011.
  • 3国家技术监督局.GB15213-94医用电子加速器性能和试验方法[S].北京:中国标准出版社,1995.
  • 4国家质量监督检验检疫总局.GB/T19046-2003医用电子加验收试验和周期检验规程[S].北京:中国标准出版社,2003.
  • 5Rangaraj D, Zhu M, Yang D, et al. Catching errors with patient-specific pretreatment machine log file analysis [ J ]. Pract Radiat Oncol, 2013, 3(2) : 80-90.
  • 6Luo W, Li J, Price RA Jr, et al. Monte Carlo based IMRT dose verification using MLC log files and R/V output[ J]. Med Phys, 2006, 33(7) : 2557-2564.
  • 7Bai S, Li G, Wang M, et al. Effect of MLC leaf position, collimator rotation angle, and gantry rotation angle errors on intensity-modulated radiotherapy plans for nasopharyngea| earcinoma[J]. Med Dosim, 2012, 38(2): 143-147.
  • 8Siebers JV, Keall PJ, Wu Q, et al. Effect of patient setup errors on sirnulataneously integrated boost head and neck IMRT treament plans [ J]. Int J Radiat Oncol Biol Phys, 2005, 63 (2) : 422-433.
  • 9Mu G, Ludlurn E, Xia P. Impact of MLC leaf position errors on simple and complex IMRT plans for head and neck cancer[ J]. Phys Med Biol, 2008, 53(1) : 77-88.
  • 10Rangel A, Dunscombe P. Tolerance on MLC leaf position accuracy for IMRT delivery with a dynamic MLC [ J]. Med Phys, 2009, 36 (7) : 3304-3309.

引证文献2

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部