摘要 本文对自动分析仪的"样本携带污染率"的测试方法和计算公式进行了分析,证明样本携带污染率K,是自动分析仪内部上一个样本的残留量相对于下一个样本的输入量加上一个样本的残留量的总体积比,提出了使用"样本携带污染率"计算公式修正测试值的见解。分析了在测试方法上自动分析仪的随机误差对"样本携带污染率"测试值的影响,建议采用多次测试的均值,避免仪器随机误差对测试结果的影响。 Key words In this paper, analyses the test methods and formula of the sample carryover rate, tested by an automatic analyzer. Through this, proved that the sample carryover rate K is a volume ration, the sample residues relative to the total volume of the next sample input and the previous sample resides. Proposed the use of "sample carryover rate" formula fix the value of the test result. Analyze of the test methods of random errors affect on the "sample carryover rate", the mean of several sample tests result recommended to avoid random deviation on the test results.
出处 《中国医疗器械信息》 2013年第2期49-55,共7页 China Medical Device Information
关键词 “样本携带污染率” 使用公式修正测试数据 随机偏差 sample carryover rate, formula fix result, random deviatio
  • 相关文献

参考文献8

  • 1P. M. G. Broughton.Crry-over in automatic analysers. Journal of Automatic Chemistry, Volume 6, Number 2 (April-June 1984), pages 94-95.
  • 2D. Pohland. Evaluation of the Automated Haematology Analyser Sysmex M-2000. J. Clin. Chem. Clin. Biochem. Vol. 27, 1989, pp. 41-47.
  • 3S.M. Lewis, R .M .Rowan, Evaluation of a prototype for a reference platelet counter. J Clin Pathol 1990;43:932-936.
  • 4冯仁丰教授(renfengfeng@online.sh.cn)临床实验室质量管理和检测性能评价-5分析误差的概念更新.
  • 5YYT0014-2005半自动生化分析仪.
  • 6YY-T0653-2008血液分析仪.
  • 7陆红,王明山,等,BECKMANCOULTERGen.S全自动血液分析仪分析性能评价,中国论文下载中心.
  • 8HITACHI7170型全自动生化分析仪精密度和交叉污染率评价,临床实验室仪器信息网(www.Clin-Lab.Com).