眼轴长度和IOL在囊袋内的对齐方式对Toric IOL旋转稳定性的影响 被引量:1

Influence of axial length and alignment in the capsular bag on rotational stability of a Toric intraocular lens
下载PDF
导出
摘要 目的:使用特定软件(Adobe Photoshop)评估人工晶状体(intraocular lens,IOL)的旋转稳定性,确定眼轴长度和IOL在囊袋内的对齐方式对ToricIOL旋转稳定性的影响。方法:研究对象眼植入AcrySof Toric IOL。使用光学相干或浸润式A型超声生物测量仪测量眼轴(axial length,AL)。手动角膜曲率计和角膜地形图测定角膜散光。IOL的对齐方式包括垂直、水平和倾斜。使用特定软件测定IOL的旋转程度并获得平均绝对差。观察术后6mo眼轴长度和IOL在囊袋内的对齐方式对旋转稳定性的影响。结果:该研究评估了168例168眼。AL为19.33~29.04(平均24.12±1.62)mm。平均IOL旋转在术后1wk为0.4°,1wk~1mo为1.1°,1~3mo为0.3°,3~6mo为0.2°。最大旋转发生在1wk~1mo。术后6moAL和晶状体的旋转有很强的相关性(r=0.92,P<0.01),而三种对齐方式(垂直、水平、斜向)对旋转的影响没有显著差异(P=0.102)。结论:眼轴越长ToricIOL的旋转程度越大。IOL在囊袋内的对齐方式对旋转程度没有影响。 · AIM: To evaluate the rotational stability of a Toric intraocular lens (IOL) using purpose-designed software and to determine the influence of axial length (AL) and in- the-bag IOL alignment on IOL rotation.· METHODS: This study enrolled eyes that had AcrySof toric IOL implantation. The AL was measured using optical coherence biometry or immersion A-scan biometry. Corneal astigmatism was determined by manual keratometry and topography. The IOL alignment was vertical, horizontal, or oblique. Rotational stability was measured using the purpose-designed software, and the mean absolute difference was determined. The effect of AL and IOL alignment on rotational stability was determined 6 months postoperatively. · RESULTS. The study evaluated 168 eyes (168 patients).The mean AL was 24.12±1.62 ( range 19.33 to 29.04) mm. The median IOL rotation was 0.4 degree from baseline to 1 week, 1.1 degree from 1 week to 1 month, O. 3 degree from 1 to 3 months, and 0.2 degree from 3 to 6 months.The maximum rotation occurred between 1 week and 1 month. There was a strong correlation between AL and IOLrotation at 6 months (r = 0.92, P〈0.01). The mean absolute difference at 6 months was not significantly different between the 3 axis placement categories when correlated with the rotation ( P= O. 102, analysis of variance).· CONCLUSION : Toric IOL rotation is greater in eyes with a longer AL. Alignment of the IOL in the capsular bag have no influence on rotation.
作者 彭婷婷 张庆
出处 《国际眼科杂志》 CAS 2013年第3期474-477,共4页 International Eye Science
关键词 TORIC 旋转稳定性 眼轴长度 对齐方式 Toric rotational stability axial length alignment
  • 相关文献

参考文献19

  • 1Ruhswurm I, Scholz U, Zehetmayer M, et al. Astigmatism correction with a foldable toric intraocular lens in cataract patients. J Cataract Refract Surg 2000 ; 26 (7) : 1022 -1027.
  • 2Chang DF. Early rotational stability of the longer Staar toric intraocular lens; fifty consecutive cases. J Cataract Refract Surg 2003 ; 29 ( 5 ) : 935-940.
  • 3De Silva DJ, Ramkissoon YD, Bloom PA. Evaluation of a toric intraocular lens with a Z-haptic. J Cataract Refract Surg 2006;32(9): 1492-1498.
  • 4Ohmi S. Decentration associated with asymmetric capsular shrinkage and intraocular lens size. J Cataract Refract Surg 1993 ;19(5) :640-643.
  • 5Chang DF. Comparative rotational stability of single-piece open-loop acrylic and plate - haptic silicone toric intraocular lenses. J Cataract Refract Surg 2008 ;34( 11) : 1842-1847.
  • 6Patel CK, Ormonde S, Rosen PH, et al. Postoperative intraocular lens rotation; a randomized comparison of plate and loop haptic implants. Ophthalmology 1999; 106 ( 11 ) : 2190 - 2195 discussion by DJ Apple, 2196.
  • 7Shimizu K, Misawa A, Suzuki Y. Toric intraocular lenses: correcting astigmatism while controlling. axis shift. J Cataract Refract Surg 1994; 20 (5) :523-526.
  • 8Weinand F, Jung A, Stein A, et al. Rotational stability of a single?piece hydrophobic acrylic intraocular lens: new method for high?precision rotation control. J Cataract Refract Surg 2007 ;33(5) :800-803.
  • 9Viestenz A, Seitz B, Langenbucher A. Evaluating the eye's rotational stability during standard photography; effect on determining the axial orientation of toric intraocular lenses. J Cataract Refract Surg 2005; 31 (3) :557-561.
  • 10Vie stenz A, Seitz B, Langenbucher A. Bifokalfotographie mit dem HRA - eine optimierte Methode zur Beurteilung der Achslage torischer Intraokularlinsen. [Bifocal photography using the HRA - an optimized method to evaluate !lOL's axial orientation] Klin Monatsbl Augenheilkd 2004;221(9) :788-790.

同被引文献4

引证文献1

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部