期刊文献+

基于LabVIEW的内窥镜手柄工效学评价系统设计

Design of Ergonomic Evaluation System for Endoscope Handle Based on LabVIEW
下载PDF
导出
摘要 设计了一个能够对内窥镜手柄进行工效学评价的系统,在模拟内窥镜手术操作实验的过程中采集人体手部与内窥镜手柄的表面接触压力以及手臂的表面肌电信号。通过对数据的分析与处理得到了内窥镜手术过程中手柄的表面压力分布情况,以及人体局部肌电信号的特征值,为内窥镜手柄设计的工效学评价提供了客观、有效的实验数据。 A system which can make ergonomic evaluation of the endoscope handle is designed. During the sim- ulated endoscope surgical procedures, the surface contact pressure between the human hand and the endoscope handle and the SEMG of the forearm are collected. Through data analysis and processing, the curve surface pressure distribution of the endoscopy and the local SEMG characteristics of human are obtained, which provides an objective and valid experiment data for the ergonomic evaluation of the design of an endoscope handle.
出处 《测控技术》 CSCD 北大核心 2013年第3期32-34,44,共4页 Measurement & Control Technology
关键词 内窥镜 工效学评价 表面肌电信号 表面接触压力 endoscope ergonomic evaluation SEMG surface contact pressure
  • 相关文献

参考文献15

  • 1Berguer R, Gerber S, Kilpatrick G, et al. An ergonomic com- parison of in-line vs pistol-grip handle configuration in a lap- aroscopic grasper [ J ]. Surgical Endoscopy, 1998,12 (6).
  • 2Matern U, Kuttler G, Giebmeyer C, et al. Ergonomic aspects of five different types of laparoscopic instrument handles un- der dynamic conditions with respect to specific laparoscopic tasks : an electromyographic-based study [ J ]. Surgical Endos- copy ,2004,18 ( 8 ) : 1231 - 1241.
  • 3American Dental Association, Survey Center. Survey of cur- rent issues in dentistry : repetitive motion injuries [ M ]. Chi- cago:American Dental Association, 1997:2 - 8.
  • 4Bramson J B, Smith S, Romagnoli G. Evaluating dental office ergonomic : risk factors and hazards [ J ]. Journal of the Ameri- can Dental Association, 1998,129 (2) : 174 - 183.
  • 5Trejo A E, Done K N, DiMartino A A, et al. Articulating vs conventional laparoscopic grasping tools-urgeons' opinions [ J ]. International Journal of Industrial Ergonomics, 2006,36 (1) :25 -35.
  • 6Herring S R, Trejo A E, Hallbeck M S. Evaluation of four cursor control devices during a target acquisition task for lap- aroscopic tool control [J]. Applied Ergonomics, 2010, 41 (1) :47 -57.
  • 7王健,金小刚.表面肌电信号分析及其应用研究[J].中国体育科技,2000,36(8):26-28. 被引量:132
  • 8Cifrek M, Medved V, Tonkovic S, et al. Surface EMG based muscle fatigue evaluation in biomechanics [ J ]. Clinical Bio- mechanics, 2009,24 (4) : 327 - 340.
  • 9Xie H B, Wang Z Z. Mean frequency derived via Hilbert- Huang transform with application to fatigue EMG signal anal-ysis[ J ]. Computer Methods and Programs in Biomedicine, 2006,82(2) :114 - 120.
  • 10蔚二文,陈维毅.表面肌电图在肌肉功能评估中的应用[J].大众科技,2007,9(7):120-122. 被引量:32

二级参考文献21

  • 1杨坚,张颖.表面肌电图在神经肌肉病损功能评估中的应用[J].中国临床康复,2004,8(22):4580-4581. 被引量:62
  • 2王奎,刘建红,宋刚.sEMG技术在评价运动性疲劳方面的方法及应用[J].安徽体育科技,2004,25(3):49-51. 被引量:57
  • 3吴毅,俞晓杰.等速技术和表面肌电图技术在骨关节炎中的应用[J].继续医学教育,2006,20(30):47-52. 被引量:11
  • 4童诗白.模拟电子技术基础[M].北京:高等教育出版社,2000..
  • 5HAGG G M. Interpretation of EMG spectral alternations and alternation indexes at sustained contraction [J]. J. Appl.Physiol. , 1992,73 (4) : 1211-1217.
  • 6WEBBER C L, et al. Influence of isometric loading on biceps EMG dynamics as assessed by linear and nonlinear tools[J]. J.Appl. Physio. , 1995, 78 (3): 814-822.
  • 7BILODEAU M, et al. EMG power spectra of elbow extensors during ramp and step isometric contractions[J]. Eur. J. Appl.Physiol. ,1991, 63:24-28.
  • 8SOKOMONOW M, et al. Electromyogram power spectra frequencies associated with motor unit recruitment strategies[J].J. Appl. Physiol. ,1990,68(3),1177-1185.
  • 9MASUDA T, et al. Recruitment threshold and muscle fiber conduction velocity of single motor unites[J]. J. Electromyogram Kinesiol, 1991, (1): 116-123.
  • 10BOUISSOU P, et al. Surface EMG power spectrum and intramuscular pH in human vastus lateralis muscle during dynamic exercise[J]. J. Appl. Physiol. ,1989,67(3):1245-1249.