期刊文献+

基于嵌入式的人体电阻抗检测系统的设计与实现 被引量:5

Based on the Embedded Body Electrical Impedance Detection System Design and Implementation
下载PDF
导出
摘要 医学上利用对人体各组织与器官的电阻抗检测,可以分析出人体的生理与病理状况,也可以检测人体的脂肪分布情况;文章设计的基于嵌入式的人体电阻抗检测系统,利用嵌入式处理器LPC2148结合低通滤波器产生一个频率为50kHz的射频恒流交变信号,将此信号加于接触人体皮肤的激励电极上,通过检测电极对地的电压就可以计算出相应的电阻抗值;通过对该系统的模拟实验表明,检测值与标准值误差小于±10%,阻抗变化曲线线性良好,这说明该系统的稳定性与准确性均能满足对于人体的电阻抗检测要求,所以该设计具有广泛的应用前景。 The medical use of electrical impedance detection of human tissues and organs that can analyze human physiological and patho logical conditions, can detect human body fat distribution. Incentive electrode designed in this paper use of the embedded processor LPC2148 electrical impedance detection system based on the embedded body, combined with the low pass filter to produce a frequency of 50 kHz RF constant current alternating signal, this signal is applied to the contact with the skin, can be calculated by detecting the voltage of the elec- trode to the electrical impedance value. By the system simulation experiments show that the detection value and standard error of less than± 10%, a good linear impedance curve, which shows the stability and accuracy of the system can meet the testing requirements of the electrical impedance of the human body, so The design has a broad application prospects.
作者 王冬艳
出处 《计算机测量与控制》 北大核心 2013年第3期604-607,共4页 Computer Measurement &Control
关键词 电阻抗 嵌入式 恒流交变信号 electrical impedance embedded constant current alternating signal
  • 相关文献

参考文献7

二级参考文献36

  • 1王育才,杨景林,李清亚,傅根铭.利用生物电阻抗测量人体脂肪的研究[J].营养学报,1994,16(1):78-82. 被引量:27
  • 2李清亚,王育才.应用人体脂肪分析仪测定老年人体脂含量[J].营养学报,1995,17(1):47-49. 被引量:5
  • 3Wang JZ, Deurenberg P. The validity of predicted body composition in Chinese adults from anthropometry and bio-electrical impedance in comparison with densitometry. Brit J Nutr, 1996,76:175--182.
  • 4Lukaski HC. Methods for the assessment of body composition: traditional and new. Am J Clin Nutr, 1987,46:437--456.
  • 5Deurenberg P, van der Kooy K, Evers P, ert al. Assessment of body composition by bio-electrical impedance in a an population aged > 60y.Am J Clin Nutr, 1990,51:3--6.
  • 6Deurenberg P. Limitations of the bio-electrical impedance method for the assessment of body fat in severe obesity. Am J Clin Nutr, 1996,64:449s--452s.
  • 7McNeill G, Fowler PA, Maughan R J, et al. Body fat in lean and overweight women estimated by six methods. Br J Nutr, 1991,65: 95--103.
  • 8van der Kooy K, Leenen R, Deurenberg P, et al. Changes in fat free mass in obese subjects after weight loss : a comparative study. Int J Obes,1992,16:675--683.
  • 9Lukaski HC. Comparison of proximal and distal placements of electrodes to assess human body composition by bio-electrical impedance. In: Eills KJ, Eestman JD, eds. Human body composition. Bosic Life Sci, 1993,40 : 39--43.
  • 10Deurenberg P, van der Kooy K, Leenen R, et al. Sex and age specific prediction formulas for estimating body composition from bio-electrical impedance: a cross validation study. Int J Obes, 1991,15:17--25.