期刊文献+

4.75keV能点四通道Kirkpatrick-Baez显微镜 被引量:8

4.75 keV four-channel Kirkpatrick-Baez microscope
下载PDF
导出
摘要 基于CH柱形靶压缩实验的时间分幅诊断需求,研制了4.75keV能点四通道Kirkpatrick-Baez(KB)显微镜。进行了四通道KB系统的光学设计、像质模拟和分析。采用支撑锥芯的方式解决了四通道KB物镜的集成装配问题,并通过"4.75keV+8keV"双能点多层膜完成了实验室内的KB系统装调。该套系统在神光Ⅱ装置成功进行了像质考核实验,获得了高分辨的四象限网格图像,成像间距符合设计要求,具备了开展时间分幅惯性约束聚变物理实验的条件。 To meet the requirement of framing diagnosis of CH cylindrical target implosion experiments, a four-channel 4.75 keV KB microscope was developed. The optical structure and imaging quality of the four-channel KB microscope were ana- lyzed. With the size limit of framing camera taken into account, the problem of assembly of the four-channel KB microscope has been solved using reference cone core. The assembly and alignment experiments of the four-channel KB microscope were per- formed in the laboratory. The clear image of four-quadrant grid with a spatial resolution of 3 μm in central field was obtained in Shenguang-Ⅱ laser facility, and the image spacing meets the design requirements, indicating that inertial confinement fusion framing experiments could be carried out.
出处 《强激光与粒子束》 EI CAS CSCD 北大核心 2013年第4期903-907,共5页 High Power Laser and Particle Beams
基金 国家自然科学基金项目(11027507,11105098) 上海市青年科技启明星计划项目(10QA1406900)
关键词 惯性约束聚变 等离子体诊断 KB显微镜 X射线多层膜 分幅相机 inertial confinement fusion plasma diagnostics Kirkpatrick Baez microscope X-ray multilayer framingcamera
  • 相关文献

参考文献30

  • 1Koch J A, Landen O L, Jr Barbee T W, et al. High-energy X-ray microscopy techniques for laser-fusion plasma research at the National Ignition Facility[J].Applied Optics, 1998, 37(10) : 1784-1795.
  • 2Meyerhofer D D. Diagnostics for high-energy-density physics[R].Los Angeles: University of California, 2009.
  • 3Drake R P. Introduction to high energy density physics (HEDP)[G]. Los Angeles: Summer School on High Energy Density Physics. 2009.
  • 4Oertel J A, Archuleta T N, Schrank L S. The large format X ray imager[J]. Review of Scientific Instruments. 2001. 72(1) :701-704.
  • 5Kimbrough J R, Bell P M. Christianson G B. et al. National Ignition Facility core X-ray streak camers[J].Review of Scientific Instruments,2001,72(1):748-750.
  • 6Hilsabeek T J, Hares J D, Kilkenny J D, et al. Pulse-dilation enhanced gated optical imager with 5 ps resolution[J]. Reviewof Scientific Instruments, 2010, 81:10E317.
  • 7Kilkenny J D. The role of experimental science in ICF examples from X-ray diagnoslics and targets[J]. European Physical Journal D,2007, 44(2):289-292.
  • 8Schneider M B, Holder J P, James D I., et al. Time resolved soft X-ray imaging (SXRI) diagnostic for use at the NIF and OMEGA lasers [C]//16th Topical Conference on High Temperature Plasma Diagnostics. 2006.
  • 9Stafford D. Time integrated soft X-ray imaging in high intensity laser experiments[D]. Davis: University of California at Davis, 2009.
  • 10Beck J B. The effects of convergent geometry on the ablative Rayleigh-Taylor instability in cylindrical implosions[D]. West Lafayette: Purdue University, 1996.

二级参考文献22

  • 1顾春时,王占山,穆宝忠,王风丽,张众,秦树基,陈玲燕.基于非周期多层膜的X射线成像研究[J].光子学报,2006,35(6):881-885. 被引量:10
  • 2Gotchev O V, Jaanimagi P A, Knauer J P, et al. High throughput, high-resolution Kirkpatrick-Baez microscope for advanced streaked imaging of ICF experiments on OMEGA[J]. Rev Sci Instrum, 2003,74(3): 2178-2181.
  • 3Gotchev O V, Hayes L J, Jaanimagi P A, et al. Large grazing-angle, multi-image Kirkpatrick Baez microscope as the front end to a high- resolution streak camera for OMEGA[J]. Rev Sci Instrum ,2003,74(12):5065-5069.
  • 4Rosch R, Boutin J Y, le Breton J P, et al. X-ray imaging with grazing-incidence microscopes developed for the LIL program[J].Rev Sci Instrum, 2007,78 : 033704.
  • 5Kirkpatrick P, Baez A V. Formation of optical images by X-rays[J].J Opt Soc Am,1948,38(9) :766-774.
  • 6Underwood J H. High-energy X-ray microscopy with multilayer reflectors[J]. Rev Sci Instrum, 1986,57(8) :2119-2123.
  • 7Gotchev O V, Jaanimagi P A, Knauer J P, et al. KB-PJX A streaked imager based on a versatile X-ray microscope coupled to a high-current streak tube[J]. Rev Sci Instrum ,2004,75(10) :4063-4068.
  • 8Bridou F, Mercier R, Raynal A, et al. Large field double Kirkpatrick Baez microscope with nonperiodic multilayers for laser plasma imaging[J]. Rev Sci Instrum, 2002,73(11) : 3789-3795.
  • 9Marshall F J, Allen M M, Knauer J P, et al. A high-resolution X-ray microscope for laser driven planar foil experiments[J].Rev Sci Instrum, 1998,75(4) : 1118-1124.
  • 10Harvey J E. X-ray optics[M]// Handbook of optics(Vol. 2). 2nd ed. New York: McGraw Hill, 1995.

共引文献27