期刊文献+

高清计算机断层扫描模式对冠状动脉支架成像质量的影响 被引量:1

Comparison in different modes of high-definition computerized tomography coronary stent imaging
下载PDF
导出
摘要 目的比较高清模式和标准模式对高清计算机断层扫描(HDCT)冠状动脉支架成像质量的影响。方法纳入30例冠状动脉支架术后复查患者(共置入支架48个),行冠状动脉支架成像,分别用高清模式和标准模式进行重建,以支架原始数据为参照,对两种不同重建模式的图像质量进行评价。结果与标准模式相比,高清模式重建测得的支架内径[(2.39±0.58)mmvs.(2.08±0.51)mm]和平均长度[(22.38±5.99)mm vs.(22.48±5.87)mm]均更接近于实际值,两组数据存在统计学差异(P<0.05)。结论 HDCT成像空间能有效提高冠状动脉支架测量的准确性。 Objective To compare the influences of high-definition mode and standard mode on the quality of high-definition computerized tomography(HDCT) coronary stent imaging.Methods A total of 30 rechecked patients after coronary stenting(48 stents) were given HDCT coronary stent imaging,and then the images were reconstructed with high-definition mode and standard mode respectively.The imaging quality of two modes was reviewed taken stent original data as the reference.Results Compared with the standard mode,the stent internal diameter(2.3854 mm vs.2.0813mm) and average stent length(22.3750 mm vs.22.4771mm) obtained by high-definition mode were closer to actual values,and there was statistical difference in data between two groups(P〈0.05).Conclusion HDCT image space can effectively improve the accuracy of coronary stent measuring.
出处 《中国循证心血管医学杂志》 2013年第1期53-54,共2页 Chinese Journal of Evidence-Based Cardiovascular Medicine
关键词 体层摄影术 X线计算机 冠状动脉 支架 Tomography X-ray computer Coronary artery Stents
  • 相关文献

参考文献8

  • 1Leber AW,Becker A,Knez A,et al. Accuracy of 64-slice computed tomography to classify and quantify plaque volumes in the proximal coronary system: a comparative study using intravascular ultrasound[J]. J Am Coil Cardiol,2006,47(3):672-7.
  • 2刘新,李颖,杨立,赵锡海,王新疆,赵绍宏,张玉霄,蔡祖龙.64层螺旋CT评价冠状动脉支架内再狭窄的价值[J].中华放射学杂志,2006,40(8):808-811. 被引量:44
  • 3王妍焱,吴国庚,周诚,高建华,焦晟,曹会志.64层螺旋CT前门控冠状动脉横断面扫描低剂量技术的初步研究[J].中华放射学杂志,2008,42(10):1018-1021. 被引量:79
  • 4Seifarth H,Raupach R,Schaller S,et al. Assessment of coronary artery stents using 16-slice MDCT angiography:evaluation of a dedicated reconstruction kernel and a noise reduction filter[J]. Eur Radiol,2005,15(4):721-6.
  • 5吕滨主译.心脏CT血管造影手册[M].第1版,北京:人民军医出版社.2009:115-6.
  • 6于淑靖,李忠信,马国军.64MSCT对冠状动脉支架通畅性评价的准确性与支架材料的关系[J].中国临床医学影像杂志,2007,18(10):716-718. 被引量:8
  • 7Gilard M,Cornily JC,Rioufol G,et al. Noninvasive assessment of left main coronary stent patency with 16-slice computed tomography[J]. Am J Cardiol,2005,95(1): 110-2.
  • 8Maintz D,Seifarth H,Raupach R,et al. 64-slice muhidetector coronary CT angiography:in vitro evaluation of 68 different stents[J]. Eur Radiol,2006,6(4): 818-26.

二级参考文献21

共引文献125

同被引文献4

引证文献1