综合测试台在滤棒吸阻检测中的应用研究 被引量:7

The Research of Test Station in Filter Pressure Drop Testing
下载PDF
导出
摘要 为寻找公司配置的德国Blogwaldt KC公司DT和北京欧美利华公司OM-Ⅰ综合测试台在滤棒吸阻项目检测车间和实验室的检测结果存在偏差的原因,在考察了综合测试台的偏倚和稳定性后,发现OM-Ⅰ综合测试台存在漂移现象明显的问题,并通过更换A/D转换模块、修改吸阻计算公式进行改造。结果显示,OM-Ⅰ综合测试台在改造后可以将车间和实验室的检测结果的偏差控制在50 Pa范围内;校准频次可以由每班1次延长至每天1次。因此,改造后的OM-Ⅰ综合测试台检测的准确性和稳定性得到了提升,检测结果更加可靠,适合车间和质检室环境温湿度条件不一致情况下的滤棒吸阻检测。 In order to find the reason of the deviation between DT and OM-Ⅰ test station in filter pressure drop,we compare the difference of the bias and stability of DT and OM-Ⅰ test station.There is an obvious drift phenomenon in OM-Ⅰ test station.In order to solve that problem,we change the A/D conversion module and modify the formula of the pressure drop.The results showed that: the bias of pressure drop test result can control within 50 Pa by using OM-Ⅰ test station;the frequency of calibration can be extended to once a day.Therefore,the accuracy and stability of the OM-Ⅰ test station is improved after upgrade,the test results are more reliable.And it's suitable for the pressure drop test in difference environment conditions between the workshop and QC room.
作者 伊恒伟
出处 《安徽农业科学》 CAS 2013年第5期2266-2267,共2页 Journal of Anhui Agricultural Sciences
关键词 综合测试台 滤棒吸阻 准确性 稳定性 Test station Filter pressure drop Accuracy Stability