强迫振荡呼吸阻力监测仪数据分析验证准则 被引量:1

Data Analysis Validation Criterion in Respiratory Impedance Monitor Based on Forced Oscillation Technique
下载PDF
导出
摘要 研究了强迫振荡呼吸阻力监测仪的数据分析模块。以最优线性近似法为理论基础,结合互相干法等数据处理方法,将采集得到的原始压力和原始流量信号经过数据处理后得到呼吸阻力、呼吸阻抗、呼吸电抗和呼吸顺应性参数指标,以反映人体的肺部状况信息。通过改进的Pwelch算法计算得到相干系数作为测试数据是否有效的准则,相干系数的阈值为0.95,大于或等于0.95为有效数据。数据分析验证准则通过Matlab仿真实验验证了算法的准确性,实验测试误差为0.000 1~0.000 2。 The data analysis module is designed for respiratory impedance monitor which is based on Forced Oscillation Technique: combining the optimal linear approximation method with the correlation method. The lung status can be reflected by parameters of respiratory resistance, respiratory reactance and respiratory compliance which were processed by collecting original press and flow signal of the pipeline. Coherence coefficient calculated by modified Pwelch algorithm is an effect criteria in the test, the threshold is equal to O. 95. The accuracy of data analysis validation criterion can be verified by Matlab simulation. Test error is 0. 000 1 -0. 000 2.
出处 《科学技术与工程》 北大核心 2013年第10期2849-2852,共4页 Science Technology and Engineering
基金 广东省科技计划项目(2012B010100043) 东莞市高校科研机构科技计划项目(2011108102016)资助
关键词 强迫振荡技术 最优线性近似法 互相干法 改进的Pwelch算法 MATLAB仿真 forced oscillation technique the optimal linear approximation correlation method modified pwelch algorithm matlab simulation
  • 相关文献

参考文献9

  • 1Horowitz J, Siegel S D, Frank P, et al. Computation of respiratory impendence from forced sinusoidal oscillations during breathing. Computers and Biomedical Research, 1983:499-521.
  • 2Rigau J, Farre ' R, Roca J, et al. A portable forced oscillation de- vice for respiratory home monitoring. Eur Respir J, 2002 : 146-150.
  • 3Pelle G, Lorino A M, Lorino H, et al. Microcomputer-based system to calculate respiratory impedance from forced random noise data. Med & Biol Eng & Comput, 1986 ;24:541-544.
  • 4Welch P D. The use of fast fourier transform for the estimation of pow- er spectra. A method based on time averaging over short, modified periodograms. IEEE Transactions on Audio and Electro Acoustics, 1967, AU-15, No. 2,70-73.
  • 5Dellac R, Rotger M, Aliverti A, et al. Noninvasive detection of ex- piratory flow limitation in COPD patients during nasal CPAP. Eur Re- spir J, 2006:983-991.
  • 6Pimmel R L, Sunderland R A, Robinson D J, et al. Instrumentation for measuring respiratory impedance by forced oscillations. IEEE Trans. Biomed. Eng, 1977; BME-24:S9-93 Mar.
  • 7Melo P L, Werneck M M, Giannella-Neto A. Influence of the pres- sure generator non-linearities in the accuracy of respiratory input im- pedance measured by forced oscillation. Med & Biol Eng &Comput, 2000; 38 : 102-108.
  • 8Kaczka D W, Yang K, Barnas G M, et al. Use of ventilator driven forced oscillations for characterizing human respiratory mechanics at low frequencies. Bioengineering Conference, Proceeding of the 1991IEEE Seventeenth Annual Northeast, 1991:73-74.
  • 9Dellac R L, Santus P, Aliverti A, et al. Detection of expiratory flowlimitation in COPD using the forced oscillation technique. Eur Respir J, 2004:232-240.

同被引文献9

引证文献1

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部