依儿童年龄设置胸部管电压优化辐射剂量和图像质量的前瞻性临床研究 被引量:5

Children Age Dependent Chest Tube Voltage Settings for Optimizing Radiation Dose and Image Quality: A Prospective Clinical Study
原文传递
导出
摘要 目的探讨依儿童年龄设置DR胸部管电压优化儿童辐射剂量和图像质量的可行性。方法将2008年1月至2010年12月在新疆医科大学第一附属医院就诊的0至14岁儿童分成0~、1~、3~、7~、11~14岁五个年龄段,每个年龄段再采用计算机随机分成优化组与对照组进行DR摄影。优化组根据年龄设置不同千伏值进行自动控制曝光DR摄影,对照组采用常规千伏值进行自动控制曝光DR摄影。辐射剂量计量用剂量面积乘积(DAP),图像质量用可视化分级分析评分(VGAS)。采用SPSS 17.0软件进行统计分析。结果最初纳入2 450例受试儿童,排除因严重基础疾病而导致某些解剖特征不能显示者35例(1.43%)后,最终纳入2 415例儿童胸部正位图像进行分析,其中优化组1 202例,对照组1 213例。结果显示:优化组DAP均值小于对照组(t=–4.967,P=0.008),而VGAS均值大于对照组(t=23.738,P=0.000)。结论依儿童年龄设置DR胸部管电压能降低儿童辐射剂量,同时改善图像质量。 Objective To explore the feasibility of children age dependent chest tube voltage settings for optimizing radiation dose and image quality in digital radiography (DR). Methods Children aged 0 to 14 who visited the First Affili- ated Hospital of Xinjiang Medical University from January, 2008 to December, 2010, were divided into the following 5 age groups: 0-, 1-, 3-, 7-, and 11-14-year-old, and each group was then randomly assigned to the optimization and control groups to take DR by computer. DR in automatic exposure control with different tube voltage depending on children age; the control group: DR in automatic exposure control with conventional fixed tube voltage. The dose area product (DAP) was adopted for the measurement of radiation dose, while the visual grading analysis score (VGAS) was for image quality. Then SPSS 17.0 was applied for statistical analysis. Results A total of 2 450 children were initially included. Finally there were 2 415 children included in the anteroposterior chest image in this study, and the other 35 children were excluded for their anatomical characteristics could not be showed due to serious primary diseases. There were 1 202 patients in the optimization group and 1 213 patients in the controlled group. The mean DAP in the optimization group was lower than that in the control group with a significant difference (t= -4.967, P=0.008). The mean VGAS in the optimization group was higher than that in the control group with a significant difference (t= 23.738, P=0.000). Conclusion Children age dependent tube voltage settings can effectively reduce radiation dose and improve image quality at the same time,
出处 《中国循证医学杂志》 CSCD 2013年第4期409-412,共4页 Chinese Journal of Evidence-based Medicine
基金 新疆医科大学第一附属医院循证医学基金(编号:2007-X2-02)
关键词 数字化X线摄影 儿童胸部 辐射剂量 图像质量 随机对照试验 Digital radiography Children chest Radiation dose Image quality Randomized controlled trial
  • 相关文献

参考文献16

  • 1Willis CE. Optimizing digital radiography of children. Eur J Radiol, 2009, 72(2): 266-273.
  • 2Slovis TL, Frush DP. Biological effects of diagnostic radiation on children. In Slovis TL, ed. Caffey's Pediatric Diagnostic Imaging. Philadephia: Elsevier 2007: 3-12.
  • 3International Commission on Radiological Protection 60:1990 Rec- ommendations of the International Commission on Radiological Protection. Annals of the ICRP, 1991, 21(1-3): 1-201.
  • 4Vano E. ICRP recommendations on "managing patient dose in digi- tal radiology"(invited paper). Radiat Prot Dos, 2005, 114(1-3): 126- 130.
  • 5Busch HP, Faulkner K. Image quality and dose management in digi- tal radiography: a new paradigm for optimization. Radiat Prot Dos, 2006, 117(1-3): 143-147.
  • 6Donnelly LF. Reducing radiation dose associated with pediatric CT by decreasing unnecessary examinations. AIR Am J Poentgenol, 2005, 184(2): 655-657.
  • 7Dougeni ED, Delis HB, Karatza AA, et al. Dose and image quality optimization in neonatal radiography. Brit J Radiol, 2007, 80(958): 807-815.
  • 8孙建忠,王志康,章伟敏,揭丽勇.直接X线成像受检者照射剂量及图像质量的实时监测[J].中华放射学杂志,2009,43(12):1264-1267. 被引量:21
  • 9Bath M, Mansson LG. Visual grading characteristics (VGC) analysis:a non-parametric rank-invariant statistical method for im- age quality evaluation. Br J Radiol, 2007, 80(951): 169-176.
  • 10Fink C, Hallscheidt PJ, Noeldge G, et al. Clinical comparative study with a large-area amorphous silicon fiat-panel detector: image qual- ity and visibility of anatomic structures on chest radiography. AIR, 2002, 178(2): 481-486.

二级参考文献18

  • 1包丽霞,刘铁峰,李湘义.儿童X线诊断检查的放射防护[J].齐齐哈尔医学院学报,2004,25(7):754-755. 被引量:8
  • 2沈晓勇,毛定立,柴春华,鲍永华.胸部X线摄影照射野外防护的研究[J].浙江预防医学,2007,19(4):29-30. 被引量:3
  • 3陈华 刘志钦.通过调整婴幼儿CT头颅扫描参数改善辐射剂量的尝试.医疗卫生装备,2008,29(10):281-282.
  • 4McCullough CH, Bruesewitz MR, Kofler JM. CT dose reduction and dose management tools: Overview of available options[J]. Radio Graphics, 2006, 26: 503-512.
  • 5Kalra MK, Maher MM, Toth TL, et al. Strategies for CT radiation dose optimlzation[J]. Radiology,2004, 230:. 619-628.
  • 6Brenner D J, Hall EJ.Computed tomography an increasing source of radiation exposure[J]. N Engl J Med, 2007, 357: 2277-2284.
  • 7Brenner DJ,Hall EJ.Current concepts-Computed tomography an increasing source of radiation exposure.New Engl J Med,2007,357(22):2277-2284.
  • 8Billinger J,Nowotny R,omolka P.Diagnostic reference levels in pediatric radiology in Austria.EurRadiol,2010,20:1572-1579.
  • 9黄奋尧,宋玉全,丘红英,陈智杰,曾庆思.膝关节摄影中照射野大小对DDR系统影像质量的影响[J].实用医学影像杂志,2008,9(1):49-51. 被引量:4
  • 10彭谦,范淼,严超贵,张晋昕,孟悛非,陈星.模拟青少年脊柱多层CT低剂量扫描的动物实验[J].中山大学学报(医学科学版),2008,29(4):431-434. 被引量:6

共引文献33

同被引文献38

引证文献5

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部