Zero drift of intraventricular and subdural intracranial pressure monitoring systems

Zero drift of intraventricular and subdural intracranial pressure monitoring systems
原文传递
导出
摘要 Objective: To assess zero drift of intra- ventricular and subdural intracranial pressure (ICP) moni- toring systems. Methods: A prospective study was conducted in pa- tients who received Codman ICP monitoring in the neuro- surgical department from January 2010 to December 2011. According to the location of sensors, the patients were ca- tegorized into two groups: intraventricular group and sub- dural group. Zero drift between the two groups and its as- sociation with the duration of ICP monitor were analyzed. Results: Totally, 22 patients undergoing intraven- tricular ICP monitoring and 27 receiving subdural ICP moni- toring were enrolled. There was no significant difference in duration of ICP monitoring, zero drift value and its absolute value between intraventricular and subdural groups (5.38 d± 2.58 d vs 4.58 d.±2.24 d, 0.77 mmHg±2.18 mmHg vs 1.03 mmHg±2.06mmHg, 1.68 mmHg~.l.55 mmHgvs 1.70mmHg.t_l.53 mmHg, respectively; all P〉0.05). Absolute value of zero drift in both groups significantly rose with the increased duration of ICP monitoring (P〈0.05) while zero drift value did not. Moreover, daily absolute value in the intraventricular group was sig- nificantly smaller than that in the subdural group (0.27 mm Hg_+ 0.32 mm Hg vs 0.29 mm Hg_-_*0.18 mm Hg, P〈0.05). Conclusion: This study demonstrates that absolute value of zero drift significantly correlates with duration of both intraventricular and subdural ICP monitoring. Due to the smaller daily absolute value, ICP values recorded from intraventficular system may be more reliable than those from subdural system.
出处 《Chinese Journal of Traumatology》 CAS CSCD 2013年第2期99-102,共4页 中华创伤杂志(英文版)
关键词 Intracranial pressure Equipment failure Cerebral ventricles Subdural space 监测系统 零点漂移 颅内压 膜下 毫米汞柱 ICP 持续时间
  • 相关文献

参考文献16

  • 1Bratton SL, Chestnut RM, Ghajar J, et al. Guidelines for the management of severe traumatic brain injury. VI. Indications for intracranial pressure monitoring. J Neurotrauma 2007;24(Suppl 1):S37-44.
  • 2Slavin KV, Misra M. Infratentorial intracranial pressure monitoring in neurosurgicai intensive care unit. Neurol Res 2003; 25(8):880-4.
  • 3Czosnyka M, Czosnyka Z, Pickard JD. Laboratory testing of three intracranial pressure microtransducers: technical report. Neurosurgery 1996;38(1):219-24.
  • 4Signorini DF, Shad A, Piper IR, et al. A clinical evaluation of the Codman MicroSensor for intracraniai pressure monitoring. Br J Neurosurg 1998;12(3):223-7.
  • 5Oshorov AV, Savin IA, Goriachev KA, et al. Post-monitor-ing changes in zero drift of"Codman" intracranial pressure sensors. Anesteziol Reanimatol 2011 (3):62-6.
  • 6A1-Tamimi YZ, Helmy A, Bavetta S, et al. Assessment of zero drift in the Codman intracranial pressure monitor: a study from 2 neurointensive care units. Neurosurgery 2009;64(1):94-8.
  • 7Fernandes HM, Bingham K, Chambers IR, et al. Clinical evaluation of the Codman microsensor intracranial pressure moni- toring system. Acta Neurochir Suppl 1998;71:44-6.
  • 8Morgalla MH, Krasznai L, Dietz K, et al. Methods of experimental and clinical assessment of the relative measurement accuracy of an intracranial pressure transducer. Technical note. J Neurosurg 2001;95(3):529-32.
  • 9Biersteker HA, Andriessen TM, Horn J, et al. Factors in- fluencing intracranial pressure monitoring guideline compliance and outcome after severe traumatic brain injury. Crit Care Med 2012;40(6): 1914-22.
  • 10Bekar A, Dogan S, Abas F, et al. Risk factors and compli- cations of intracranial pressure monitoring with a fiberoptic device. J Clin Neurosci 2009; 16(2):236-40.

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部