3M和5M显示器对不同分辨率乳腺影像图中微钙化识读的影响 被引量:5

Effects of 3-megapixel and 5-megapixel Monitors on Detecting Micro-calcification in High-and Low-resolution Breast Images
原文传递
导出
摘要 通过使用高、低两种分辨率的乳腺影像图来比较3M与5M医用专业显示器对乳腺微钙化识读的影响。选择高、低两种分辨率的乳腺影像图各100例(各含微钙化病例40例、正常对照病例60例)。由1名高年资和1名低年资放射科医师评估两种显示器对乳腺微钙化的显示率,识别效能用ROC曲线判断,并使用Kappa分析检验两名医生的判读一致性。在低分辨率影像图组中,两名医生在3M与5M医用专业显示器上对微钙化的识别效能相同(P=0.451及0.559);在高分辨率影像图组中,高年资医生使用5M医用专业显示器对乳腺微钙化的识别率明显高于3M(P=0.022),低年资医生的识别率无显著差异(P=0.141)。两名医生在5M显示器上判读的一致性都好于3M显示器,在5M显示器上判读高分辨率影像图时,两名医生的判读有极好的一致性(K=0.862)。因此,对不同分辨率的乳腺影像图,应配套相应分辨率的显示器,高年资医师识读高分辨率影像图组时应配套5M显示器更利于微钙化的检出。 To evaluate the influence of 3- and 5-megapixel medical professional monitors in detecting the micro-calcifi- cations on high- and low-resolution breast images, we performed a retrospective study in low- (n= 100) and high- resolution (n= 100) data, including 40 micro-calcification patients in a group and 60 normal ones in control group re- spectively. Two doctors, one junior, and the other senior, reviewed all the images without knowing the clinical data and histology, and their observations of each image with different monitors were calculated. The areas under the ROC curves (Az) were compared. Finally, the interpretation consistency of the two doctors was assessed using Kap- pa analysis. In the low resolution data group, the two doctors' detection performance of breast micro-calcifications were very similar in the 3M and 5M medical professional monitors (P=0. 451 and 0. 559). In the high resolution group, however, the senior doctor's recognition rate on the 5M monitor was significantly higher than that on the 3M (P=0. 022), while the junior's recognition rate had no significant difference (P=0. 141) between the two readings. The two doctors' interpretatiorx consistency on 5M monitowas better than that on 3M monitor. For the high-resolu- tion images on the 5M monitor, the interpretation of the two doctors had extremely great consistency (K=0. 862). Therefore, different breast images of different resolutions should match corresponding resolution monitor. Interpre- tation of high-resolution images with 5M monitor has more advantages in the micro-calcification detection for senior doctors.
出处 《生物医学工程学杂志》 EI CAS CSCD 北大核心 2013年第2期245-248,共4页 Journal of Biomedical Engineering
关键词 显示器 分辨率 乳腺 影像识读 数字化X线摄影 Monitor Resolution Breast Imaging reading Digital radiography
  • 相关文献

参考文献9

  • 1MORGAN M P, COOKE M M, MCCARTHY G M. Micro-calcifications associated with breast cancer: an epiphenomenonor biologically signi. cant feature of selected tumours. [J]. JMammary Gland Biol Neoplasia, 2005, 10(2) : 181-187.
  • 2UEMATSU T,KASAMI M, UCHIDA Y. Soft-copy readingin digital mammography of microcalcifications : diagnostic per-formance of a 5-megapixel cathode ray tube monitor versus a 3-megapixel liquid crystal display monitor in a clinical setting[J].Acta Radiol, 2007, 48(7): 714-720.
  • 3KAMITANI T, YABUUCHI H, SOEDA H,et al. Detectionof masses and microcalcifications of breast screen on digitalmammograms: comparison among hard copy film. 3-megapixelliquid crystal display (LCD) monitors and 5-megapixel LCDmonitors: an observer performance study [J]. Eur Radiol,2007, 17(5): 1365 - 1371.
  • 4American College of Radiology (ACR). Breast imaging repor-ting and data system (BI-RADS) [M]. 4th ed. Reston: AmCollege Radial, 2003.
  • 5DORFMAN D D, BERBAUM K S, METZ C E. Receiver op-erating characteristic rating analysis. Generalization to thepopulation of readers and patients with the jackknife method[J], Invest Radiol, 1992, 27(9): 723-731.
  • 6SVANHOLM H, STARKLINT H, GUNDERSEN H J, et al.Reproducibility of histomorphologic diagnoses with special ref-erence to the kappa statistic[J]. APMIS, 1989, 97(8) : 689-698.
  • 7LANDIS J R, KOCH G G. The measurement of observer a-greement for categorical data[J]. Biometrics, 1977, 33 (1);159-174.
  • 8IDEGUCHI T,HIGASHIDA Y, KAWAJI Y, et al. New CRsystem with pixel size of 50 microm for digital mammography:physical imaging properties and detection of subtle microcalcifi-cations[J]. Radiat Med, 2004, 22(4): 218-224.
  • 9BACHER K,SMEETS P, DE HAUWERE A, et al. Imagequality performance of liquid crystal display systems: influenceof display resolution,magnification and window settings oncontrast-detail detection [J]. Eur J Radiol, 2006,58(3):471-479.

同被引文献16

  • 1何劲松,王先明,朱国献,刘月华,张家庭,李征毅,李泉水.高频超声引导下Mammotome旋切系统在乳腺微小钙化灶切取活检中的价值[J].中国微创外科杂志,2006,6(9):667-668. 被引量:16
  • 2曹厚德.“软阅读”及专业显示器[J].中国医学计算机成像杂志,2006,12(5):359-363. 被引量:9
  • 3包刚,杨德启,周波,刘鹏,王殊,程琳.新辅助化疗对乳腺癌组织中雌激素、孕激素受体及P53和CerbB-2表达的影响[J].中华医学杂志,2007,87(40):2843-2845. 被引量:36
  • 4BACHER K,SMEETS P,DE HAUWERE A. Image quality performance of liquid crystal display systems:influence of display resolution,magnification and window settings on contrast-detail detection[J].European Journal of Radiology,2006,(03):471-479.doi:10.1016/j.ejrad.2005.12.016.
  • 5CHIDA K,ZUGUCHI M,SAI M. Optimization of tube potential-filter combinations for film-screen mammography:a contrast detail phantom study[J].Clinical Imaging,2006,(04):246-250.
  • 6SCHARITZER M,PROKOP M,WEBER M. Detectability of catheters on bedside chest radiographs:comparison between liquid crystal display and highresolution cathode-ray tube monitors[J].Radiology,2005,(02):611-616.
  • 7BALASSY C,PROKOP M,WEBER M. Flatpanel display (LCD) versus high-resolution grayscale display (CRT) for chest radiography:an observer preference study[J].American Journal of Roentgenology,2005,(03):752-756.
  • 8UEMATSU T,KASAMI M,UCHIDA Y. Soft-copy reading in digital mammography of microcalcifications:diagnostic performance of a 5-megapixel cathode ray tube monitor versus a 3-megapixel liquid crystal display monitor in a clinical setting[J].Acta Radiologica,2007,(07):714-720.
  • 9KAMITANI T,YABUUCHI H,SOEDA H. Detection of masses and microcalcications of breast screen on digital mammograms:comparison among hard copy lm,3-megapixel liquid crystal display (LCD) monitors and 5-megapixel LCD monitors:an observer performance study[J].European Journal of Radiology,2007,(05):1365-1371.
  • 10钟晓绯,彭玉兰,刘吉斌.美国放射学会BI-RADS乳腺超声术语及评估[J].临床超声医学杂志,2009,11(10):719-720. 被引量:37

引证文献5

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部