摘要 本文开发了一种用于检测脑卒中患者膝关节运动能力的多参数测试仪。该仪器利用Auto CAD设计膝关节角度测试装置,利用多功能数据采集卡实现对传感器数据的采集,由上位机进行关节角度、角速度等参数的定量检测与分析。软件系统采用Visual C++MFC框架开发,利用Access共享数据库实现患者信息存储,ADO技术实现数据库与软件系统的互联。经对20名正常人进行初步测试,各参数的组内相关系数(ICC)均大于0.8,表明仪器具有良好的可靠性。 This article is aimed to present a design of a testing device for knee joint kinematics parameters of patients with strokes. We used Auto CAD to design the knee joint sensor device, and applied multifunction data acquisition card to collect sensor data. Then we transferred the data to the computer to quantitatively detect and analyze the joint angle, angular velocity and other parameters. The software system used Visual CA--4-MFC frame and shared data- base Access, and used ADO technology to realize the collection between the software system and patient information system. After a preliminary test of 20 healthy subjects, the result showed that intraclass correlation coefficient (ICC) was more than 0. 8, indicating a good reliability of the instrument.
出处 《生物医学工程学杂志》 EI CAS CSCD 北大核心 2013年第2期320-326,共7页 Journal of Biomedical Engineering
基金 上海市教委科研创新项目资助(11cxy46) 上海市科委重点科技攻关项目资助(10441900700)
关键词 膝关节角度 运动参数测试 VC++ MFC框架 ADO技术 Knee joint angle Motion parameters test VC++ MFC frame ADO technology
  • 相关文献

参考文献10

二级参考文献47

  • 1郑秀媛.现代运动生物力学[M].北京:国防工业出版社,2002.10.
  • 2SCHNEIDER R,GAUTIER JC. Leg Weakness Due to Stroke: Site of Lesions, Weakness Pattems and Causes[J]. Brain(S0006-8950), 1994,117:347-354.
  • 3HAMMOND MC, FITTS SS, KRAFT GH, et al Co-contraction in the Hemiparetic Forearm: Quantitative EMG Evaluation [J]. Arch Phys Med Rehabil (SS0003-9993), 1988,69(5) :348-351.
  • 4CANNING CG,ADA L,O'DWYER N. Slowness to Develop Force Contributes to Weakness after Stroke[J].Arch Phys Med Rehabil ( SS0003 - 9993 ), 1999,80 ( 1 ) :66-70.
  • 5LEVIN MF, HUI-CHAN C. Ankle Spasticity is Inversely Correlated with Antagonist Voluntary Contraction in Hemiparetic Subjects[J]. Electromyogr Clin Neurophysiol(S0301-150X), 1994,34(7) :415-425.
  • 6HAGBARTH KE. Evaluation of and Methods to Change Muscle Tone [J]. Stand J Med Rehabil Med Suppl(S0036-5505), 1994,41 : 237-241.
  • 7WILK KE. Isokinetic Testing Goals. Standards and Knee Test Interpretation[M]. Biodex Clinical Protocol Manual. Biodex Cor Poration,1990.
  • 8CRAM JR, KASMAN GS, HOLTZ J. Introduction to Surface Electro-myography [M]. Maryland: An Aspen Publishers, 1998.
  • 9Kellis E. Quantification of quadriceps and hamstring antagonist activity[J]. Sports Med,1998,25(1) :37--42.
  • 10Ng GYF, Man VY. EMG analysis of vastus medialis obliquus and vastus lateralis during static knee extension with different hip and ankle positions[J]. NZ J Physiother, 1996,12 : 7--10.

共引文献62

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部