Comparison of Anterior Segment Optical Coherence Tomography and Ultrasound Biomicroscopy for Iris Parameter Measurements in Patients with Primary Angle Closure Glaucoma 被引量:8

Comparison of Anterior Segment Optical Coherence Tomography and Ultrasound Biomicroscopy for Iris Parameter Measurements in Patients with Primary Angle Closure Glaucoma
下载PDF
导出
摘要 Purpose: To compare the repeatability and consistency of anterior segment optical coherence tomography (AS-OCT) and ultrasound biomicroscopy (UBM) in measuring iris parameters in patients with primary angle closure glaucoma. Methods: Twenty-two patients (38 eyes) with primary angle closure glaucoma,including 5 eyes with acute angle closure glaucoma,10 fellow eyes of acute angle closure glaucoma, and 23 eyes with chronic angle closure glaucoma, were recruited consecutively in our hospital. All subjects underwent anterior scanning by AS-OCT and UBM. Peripheral iris thickness (PIT) and iris curvature (IC) in the anterior segment image obtained by AS-OCT and UBM were measured twice. The reproducibility of these two scans was evaluated by the intraclass correlation coefficient (ICC). A paired t-test was used to compare the difference between the two scans and the 95% limits of agreement (LoA) were calculated. Results:The ICCs of PIT and IC measured by UBM were 0.892 and 0.936 respectively, while for AS-OCT these values were 0.629 and 0.859, respectively. UBM had a higher reproducibility in both PIT and IC measurements as compared with AS-OCT.Differences in PIT measurement between AS-OCT and UBM(P=0.331).were not statistically significant, the 95% LoA (-0.178~0.156) mm was 36.1~41.2% of the mean. The IC was 0.053 mm smaller when measured by UBM than by AS-OCT (P=0.017), with the 95% LoA (-0.100~0.206) mm, or 36.2~74.6% of the mean.Conclusion:UBM had a higher reproducibility in measuring iris parameters than AS-OCT. The consistency between AS- OCT and UBM in measuring iris parameters was low in primary angle closure glaucoma patients. (Eye Science 2013; 28:1-6) Purpose: To compare the repeatability and consistency of an- terior segment optical coherence tomography (AS-OCT) and ultrasound biomicroscopy (UBM) in measuring iris parame- ters in patients with primary angle closure glaucoma. Methods: Twenty-two patients (38 eyes) with primary angle closure glaucoma, including 5 eyes with acute angle closure glaucoma, 10 fellow eyes of acute angle closure glaucoma, and 23 eyes with chronic angle closure glaucoma, were re- cruited consecutively in our hospital. All subjects underwent anterior scanning by AS-OCT and UBM. Peripheral iris thick- ness (PIT) and iris curvature (IC) in the anterior segment image obtained by AS-OCT and UBM were measured twice. The reproducibility of these two scans was evaluated by the intraclass correlation coefficient (ICC). A paired t-test was used to compare the difference between the two scans and the 95% limits of agreement (LoA) were calculated. Results: The ICCs of PIT and IC measured by UBM were 0.892 and 0.936 respectively, while for AS-OCT these values were 0.629 and 0.859, respectively. UBM had a higher repro- ducibility in both PIT and IC measurements as compared with AS-OCT. Differences in PIT measurement between AS-OCT and UBM (P--0.331) were not statistically significant, the 95% LoA (-0.178-4).156) mm was 36.1--41.2% of the mean. The IC was 0.053 mm smaller when measured by UBM than by AS- OCT (P--0.017), with the 95% LoA (-0.100 - 0.206) mm, or 36.2-74.6% of the mean.Conclusion: UBM had a higher reproducibility in measuring iris parameters than AS-OCT. The consistency between AS- OCT and UBM in measuring iris parameters was low in primary angle closure glaucoma patients.
出处 《Eye Science》 CAS 2013年第1期1-6,共6页 眼科学报(英文版)
基金 Science and Technology Plan Program of Guangdong Province (2012A080204002 2012B031800294 2011B080701033)
关键词 断层扫描仪 参数测量 生物显微镜 光学相干 青光眼 原发性 患者 虹膜 iris parameter ultrasound biomicroscopy anterior segment optical coherence tomography glaucoma
  • 相关文献

参考文献1

二级参考文献10

共引文献12

同被引文献36

引证文献8

二级引证文献37

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部