应用锥形束CT研究盆腔肿瘤放射治疗分次间及分次内的摆位误差 被引量:28

A Study of Interfractional and Intrafractional Setup Errors in Radiotherapy by Cone Beam CT (CBCT) for Pelvic Cancer
原文传递
导出
摘要 [目的]研究盆腔肿瘤放疗分次间及分次内的摆位误差,计算临床靶区(CTV)到计划靶区(PTV)的外放边界(MPTV)。[方法]应用ELEKTA Synergy IGRT直线加速器系统治疗盆腔肿瘤24例,通过锥形束CT(CBCT)影像技术获得患者左右(X)、头脚(Y)、前后(Z)方向线性摆位误差以及分别以X、Y、Z轴旋转形成相应的U、V、W旋转摆位误差,分析分次间、分次内的摆位误差,计算MPTV。[结果]24例患者共行365次首次摆位后CBCT扫描,系统误差(均数)±随机误差(标准差)在X、Y、Z方向上分别为(0.73±1.67)mm、(0.11±4.69)mm、(-1.77±2.60)mm,U、V、W方向上分别为(0.81°±1.11°、-0.01°±1.18°、0.39°±0.88°;纠正后摆位误差显著低于首次摆位后摆位误差(P<0.05),治疗后摆位误差较纠正后显著增加(P<0.05);纠正前X、Y、Z方向的MPTV分别为4.93mm、12.63mm、7.06mm,纠正后X、Y、Z方向的MPTV分别为1.25mm、2.43mm、1.67mm。[结论]盆腔肿瘤放疗时Y方向摆位误差最大,Z方向次之,X方向最小,旋转误差一般不超过3°;应用CBCT实施IGRT,可在线实时纠正分次间的摆位误差,提高放疗的精确度;应用CBCT引导放疗时,MPTV可缩小至所有方向均为3mm。 [Purpose] To analyze the interfractional and intrafractional setup errors in radiotherapy for pelvic cancer and to estimate the margin required for clinical target volume (CTV) to planning target volume (PTV)(Mprv). [Methods] Twenty-four patients with pelvic cancer were treated with ELEKTA Synergy IGRT linear accelerator. Linear errors(X, Y, Z axis) and rotational errors(U, V, W) were obtained by cone beam CT (CBCT). Interfractional and intrafractional setup errors and MPTV were analyzed. [Results] Twenty-four patients received a total of 365 CBCT scans after initial setup. The systemic+random errors on X(left-right),Y(superior-inferior), Z(an- terior-posterior) directions were(0.73±1.67) mm, (0.11±4.69) mm, (-1.77±2.60)mm respectively, and those on U,V,W directions were (0.81°±1.11°),(-0.01°±1.18°,(0.39°±0.88°) respectively. All the setup errors were corrected. Comparing to the initial setup errors, the re-positioning set- up errors decreased significantly (P〈0.05). Setup errors after radiation delivery were increased significantly compared to the re-positioning setup errors (P〈0.05). MPTV before re-positioning was 4.93mm in X axis, 12.63mm in Y axis,7.06mm in Zaxis. After re-positioning MPTV was 1.25mm, 2.43mm, 1.67mm in X, Y, Z axis, respectively. [ Conclusion] The spatial distributions of setup error is maximum in Y direction,moderate in Z direction,and minimum in X direction,and the most of rotational error was less than 3°. Real-time online correction of setup errors using CBCT can reduce the interfractional errors and improve the precise of radiotherapy in patients with pelvic cancer. With the use of CBCT, MPTV may reduce to 3mm in all directions.
出处 《肿瘤学杂志》 CAS 2013年第4期245-248,244,共4页 Journal of Chinese Oncology
基金 福建省卫生教育联合攻关计划项目(WKJ2008-2-47-1) 福建省临床重点专科建设项目(2012年)
关键词 图像引导放射治疗 锥形束CT(CBCT) 盆腔肿瘤 靶区外放边界 image-guided radiotherapy (IGRT) cone beam computed tomography (CBCT) pelvic neoplams CTV-PTV margin
  • 相关文献

参考文献10

  • 1Stroom JC,Heijmen BJ. Geometrical uncertainties,radio- therapy planning mar- gins,and the ICRU-62 report [J]. Radiother Oncol, 2002,64(1): 75-83.
  • 2van Herk M. Errors and margins in madiotherapy [J]. Semin Radiat 0ncol,2004,14(1): 52-64.
  • 3McKenzie A,van Herk M,Mijnheer B. Margins for geo- metric uncertainty around organs at risk in radiotherapy [J]. Radiother Oncol, 2002,62(3):299-307.
  • 4Van Herk M, Remeijer P, Lebesque JV. Inclusion of geo- metric uncertainties in treatment plan evaluation [J]. Int J Radiat Oncol Biol Phys,2002,52(5): 1407-1422.
  • 5van Herk M,Remeijer P,Rasch C,et al. The probability of correct target dosage : dose-population histograms for de- riving treatment margins in radiotherapy [J].Int J Radiat Oncol Biol Phys, 2000 ;47(4): 1121-1135.
  • 6刘跃平,刘新帆,李晔雄,李涛.盆腔肿瘤三维适形放疗摆位重复性研究[J].中华放射肿瘤学杂志,2006,15(4):313-316. 被引量:44
  • 7Enmark M, Korreman S, NystrSm H. IGRT of prostate can- cer; is the margin reduction gained from daily IG time de- pendent?[J]. Acta Oncologica,2006,45(7): 907-914.
  • 8Wong JR,GrimmL,Uematsu M,et al. Image-guided radio- therapy for prostate cancer by CT linear accelerator combi- nation: prostate movements and dosimetrie considerations [J]. Int J Radiat Oncol Biol Phys,2005,61(2): 561-569.
  • 9Burridge N,Airier A,Marchant T,et al. Online adaptive radiotherapy of the bladder: small bowel irradiated-volume reduction [J].Int J Radiat Oncol Biol Phys,2006,66 (3): 892-897.
  • 10Palombarini M ,Mengoli S,Fantazzini P,et al. Analysis of interfraction setup errors and organ motion by daily kilo- voltage cone beam computed tomography in intensity modulated radiotherapy of prostate cancer[J].Radiat Oncol, 2012,7:56.

二级参考文献21

  • 1林承光,邓小武,黄峻,温志祥,肖达宜,林览峰.鼻咽癌适形放疗实施过程精确度和重复性研究[J].肿瘤学杂志,2004,10(4):208-210. 被引量:32
  • 2刘跃平,刘新帆,李晔雄,金晶,关莹.前列腺癌三维适形放疗中体位对靶区和正常组织体积及照射剂量的影响[J].中华放射肿瘤学杂志,2005,14(5):422-426. 被引量:12
  • 3胡逸民.调强适形放射治疗.见:殷蔚伯,谷铣之,主编.肿瘤放射治疗学.第3版.北京:中国协和医科大学出版社,2002.179-211.
  • 4Malone S,Szanto J,Perry G,et al.A prospective comparison of three systems of patient immobilization for prostate radiotherapy.Int J Radiat Oncol Biol Phys,2000,48:657-665.
  • 5Fiorino C,Reni M,Bolognesi A,et al.Set-up error in supinepositioned patients immobilized with two different modalities during conformal radiotherapy of prostate cancer.Radiother Oncol,1998,49:133-141.
  • 6Gilbeau L,Octave-Prignot M,Loncol T,et al.Comparison of setup accuracy of three different thermoplastic masks for the treatment of brain and head and neck tumors.Radiother Oncol,2001,58:155-162.
  • 7Ezz A,Munro P,Porter AT,et al.Daily monitoring and correction of radiation field placement using a video-based portal imaging system:a pilot study.Int J Radiat Oncol Biol Phys,1992,22:159-165.
  • 8Wang L,Jacob R,Chen LL,et al.Stereotactic IMRT for prostate cancer:Setup accuracy of a new stereotactic body localization system.J Appl Clin Med Phys,2004,5:18-28.
  • 9Hunt MA,Kutcher GJ,Burman C,et al.The effect of setup uncertainties on the treatment of nasopharynx cancer.Int J Radiat Oncol Biol Phys,1993,27:437-447.
  • 10Bijhold J,van Herk M,Vijlbrief R,et al.Fast evaluation of patient setup during radiotherapy by aligning features in portal and simulator images.Phys Med Biol,1991,36:1665-1679.

共引文献43

同被引文献192

引证文献28

二级引证文献114

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部