ALLEGRO Oculyzer与IOL Master测量角膜曲率及前房深度的比较 被引量:3

Comparison of ALLEGRO Oculyzer and IOLMaster in Measuring Corneal Curvature and Anterior Chamber Depth
原文传递
导出
摘要 目的:比较ALLEGRO Oculyzer与IOLMaster两种非接触性检查仪器的角膜曲率及前房深度测量结果。方法:随机选取符合入选条件的门诊患者80例160眼,分别用ALLEGRO Oculyzer与IOLMaster进行角膜曲率及前房深度(anterior chamber depth,ACD)测量,并对测量结果进行配对t检验,分析两种检查方法测量结果的差异。结果:ALLEGRO Oculyzer测得的最陡峭径线曲率半径(R1)、最平坦径线曲率半径(R2)与平均曲率半径(Rm)分别为7.78±0.28 mm、7.51±0.27 mm与7.64±0.25 mm;IOLMaster相应的测量结果分别为7.78±0.29 mm、7.50±0.28 mm与7.64±0.26 mm。经配对t检验,两种方法测得R1、R2、Rm结果差异无统计学意义(P>0.05)。ALLEGRO Oculyzer与IOLMaster测得的ACD分别为3.46±0.38 mm与3.33±0.39 mm,两者测量结果差异有统计学意义(P<0.05)。结论:ALLEGRO Oculyzer与IOLMaster测得的角膜曲率在正常人群都具有较高的准确性,但ACD测量结果有一定差异,在使用ACD数据进行IOL计算、术前评估等临床应用时应注意。 Objective: To investigate the difference between ALLEGRO Oculyzer and IOL Master in measuring corneal curvature and anterior chamber depth (ACD). Methods: A total of 80 cases (160 eyes) were measured with the ALLEGRO Oculyzer and IOL Master. The steepest corneal radius (R1), flattest corneal radius (R2), mean corneal radius (Rm), anterior chamber depth (ACD) were recorded and compared. Results: The R1, R2, Rm was respectively 7.78± 0.28 mm, 7.51± 0.27 ram, 7.64± 0.25 mm when measured by ALLEGRO Oculyzer and 7.78± 0.29 ram, 7.50± 0.28 mm, 7.64± 0.26 mm by IOL Master. There was no statistical difference between the two methods (P 〉0.05). The ACD measured by ALLEGRO Oculyzer was 3.46± 0.38mm and was 3.33± 0.39mm by IOL Master, with statistical difference (P〉0.05). Conclusion: Both ALLEGRO Oculyzer and IOL Master showed high accuracy in measuring corneal curvature, but had different results in measuring ACD, so it should be paid attention when ACD data were applied in IOL calculation and preoperative assessment.
出处 《现代生物医学进展》 CAS 2013年第6期1115-1117,共3页 Progress in Modern Biomedicine
关键词 Oculyzer IOL MASTER 角膜曲率 前房深度 Oculyzer IOLMaster Corneal curvature Anterior chamber depth
  • 相关文献

参考文献16

  • 1姚春艳,翟芳,刘超,郑玲,孙守彬,温泉,孙伟英,盛毅,朱佳.Q值调整个体化准分子激光原位角膜磨镶术术后角膜非球面变化[J].现代生物医学进展,2012,12(3):469-472. 被引量:1
  • 2Biermann J, Bredow L, Boehringer D, et al. Evaluation of ciliary sul- cus diameter using ultrasound biomicroscopy in emmetropic eyes and myopic eyes. [J]. Cataract Refract Surg, 2011, 37(9):1686-1693.
  • 3Salouti R, Nowroozzadeh MH, Zamani M, et al. Comparison of anterior chamber depth measurements using Galilei, HR Pentacam, and Orb- scan II[J]. Optometry, 2010, 81(1):35-39.
  • 4Buzzonetti L, Petrocelli G, Laborante A, et al. A new transepithelial phototherapeutic keratectomy mode using the NIDEK C XIII excimer laser[J]. Refract Surg, 2009, 25(1):S122-S124.
  • 5Meyer CH, Krohne TU, Holz FG. Intraocular pharmacokinetics after a single intravitreal injection of 1.5 mg versus 3.0 mg of bevacizumab in humans. [J]. Retina, 2011, 31 (9): 1877-1884.
  • 6Chela YA, Hirnschall N, Findl O. Evaluation of 2 new optical biometry devices and comparison with the current gold standard biometer[J]. Cataract Refract Surg, 2011, 37(3):513-517.
  • 7Savini G, Carbonelli M, Barboni P, et al. Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes[J]. Journal of Cataract and Refractive Surgery, 2011, 37(2):302-309.
  • 8Olsen T. Prediction of the effective postoperative (intraocular lens) an- terior chamber depth. Journal of Cataract and Refractive Surgery[J]. 2006, 32(3):419-424.
  • 9Himschall N, Murphy S, Pimenides D, et al. Assessment of a new av- eraging algorithm to increase the sensitivity of axial eye length mea- surement with optical biometry in eyes with dense cataract [J]. Cataract Refract Surg, 2011, 37(1):45-49.
  • 10Roh YR, Lee SM, Han YK, et al. Intraocular lens power calculation using IOLMaster and various formulas in short eyes[J]. Korean J Oph- thalmol, 201 I, 25(3):151-155.

二级参考文献31

  • 1胡军,赖巧红,项楠,李贵刚.眼部疾病的B超诊断分析[J].中国现代医学杂志,2006,16(19):2998-3000. 被引量:2
  • 2姚春艳,路宏,唐伟.近视眼LASIK术后眼高阶像差变化[J].现代生物医学进展,2007,7(4):593-595. 被引量:3
  • 3刘家琦,李凤鸣主编.实用眼科学[M].人民卫生出版社,2009,843.
  • 4钱蕴秋主编.实用超声诊断手册[M].人民军医出版社,2006,94.
  • 5周永昌,郭万学,主编.超声医科[M]科学技术文献出版社,1988,4:304-305.
  • 6Chmait R, Pretorius D, Moore T, et al. Prenatal detection of associated anomalies in fetuses diagnosed with cleft lip with Or without cleft palate in utero[J]. Ultrasound Obstet Gynecol, 2006, 27 (2) : 173-176.
  • 7Parravano M, Oddone F, Sampalmieri M,et al. Reliability of the IOL Master in axial length evaluation in silicone oil-filled eyes [J]. Eye, 2007, 21(7): 909-911.
  • 8Byme SF, Green RL. Ultrasound of the eye and orbit [M]. St,Louis, Mosby-Year Book lnc, 1992:243-246.
  • 9Henderson J W. Orbital tumors [M].3rd ed. New York, Ravea Press, 1994:228.
  • 10Fujikado T,Kuroda T,Ninomiya S,et al.Age-related changes in ocul-ar and corneal aberrations[J].Am Ophthalmol,2004,138:143-146.

共引文献4

同被引文献36

引证文献3

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部