期刊文献+

新型头模型EIT实验装置的设计及应用

Design and Application of New Head Phantom EIT Experimental Device
下载PDF
导出
摘要 目的:研制能精确定位的头模型EIT实验装置,满足进一步研究的需要。方法:安装电极和支撑颅骨物理模型的半球状结构选用有机玻璃材料,电极用银电极,支撑悬挂结构材料选用硬度较大、方便取材的钢或铜材料。结果:新型头部物理模型在精度和稳定性方面均达到了要求,能够满足进一步实验及其相关分析的需要。结论:新型头部物理模型对扰动目标的水平位置和垂直位置定位均精确到mm,能够满足进一步实验及其相关分析的需要。 Objective To develop a head EIT phantom experimental device for precise localization.Methods Plexiglass material was used to make the electrode mounting assemble and semispherical support for the skull model,and steel or copper was applied to the support suspension,with the silver electrode adopted.Results The precision and stability of the new phantom could meet the requirements of further research and analysis.Conclusion The localization precision of the new head phantom is limited to 1 mm for the horizontal and vertical positions of the disturbance object,and thus the requirements of further research and analysis can be satisfied.
出处 《医疗卫生装备》 CAS 2013年第4期1-3,共3页 Chinese Medical Equipment Journal
基金 国家自然科学基金重点项目(50937005) 国家自然科学基金面上项目(61071033) 国家科技支撑计划项目(2012BAI19B01 2011BAI08B13)
关键词 电阻抗断层成像 新型头部物理模型 颅骨模型 非均匀电导率 均匀电导率 electrical impedance tomography(EIT) new head phantom skull phantom non-uniform conductivity uniform conductivity
  • 相关文献

参考文献15

二级参考文献34

  • 1任超世,池学东,崔云莉.生物电阻抗成像技术与仪器[J].国外医学(生物医学工程分册),1995,18(5):249-255. 被引量:3
  • 2白峰 董秀珍 等.用虚拟设备驱动程序实现Windows 95平台的实时数据采集系统[J].计算机应用研究,1999,16(2):146-146.
  • 3汤孟兴,董秀珍,秦明新,王志敏,付峰,史学涛.动态电阻抗断层成像反投影算法的仿真研究[J].第四军医大学学报,1997,18(1):42-44. 被引量:6
  • 4史学涛,Chin J Biomed Eng,1997年,6卷,3期,119页
  • 5史学涛,Chin J Biomed Eng,1995年,4卷,2期,100页
  • 6Liu R G,第四军医大学学报,2000年,21卷,1期,107页
  • 7Tang M,Med Biol Eng Comput,1998年,36卷,4期,395页
  • 8Woo E J,IEEE Trans Med Imaging,1993年,12卷,2期,137页
  • 9Hua P,学位论文,1990年
  • 10Pan Tinsu,IEEE Transactions Medical Imaging,1992年,11卷,4期

共引文献48

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部