摘要
目的建立测定水中氰化物的流动注射分析法。方法应用SKALAR SAN++型全自动流动注射分析仪直接在线蒸馏测定饮用水中氰化物。结果该测定方法相关性好,r=0.999 4,线性范围1.25~50μg/L,相对偏差为1.06%~6.0%,检出限为1.2μg/L。结论方法简便,快速,干扰少,灵敏度高,有较高的实用价值,特别适合大批量样品的测定。
[ Objective ] To develop a flow injection analysis method for determination of cyanide in drinking water. [ Methods] By u-sing SKALAR SAN + + automatic flow injection analyzer, the content of cyanide in drinking water was determined with the straight distillation. [ Results] The linear relationship of this method was good (r = O. 999 4), the linearity range was 1.25-50 μg/L, the rela-tive deviation ranged from 1.06%-6.0%, and the detection limit was 1.2 μg/L. [ Conclusion] This method is simple and rapid, with less interference and high sensitivity, which has higher practical value and is suitable for the detection of large quantities of samples.
出处
《职业与健康》
CAS
2013年第9期1087-1088,共2页
Occupation and Health
基金
呼市饮用水检测平台与检测体系建设(项目编号:呼和浩特市科技局2010-重-农社-23)