搏动式心脏辅助装置体外模拟试验的建立及实验研究 被引量:4

Design and experimental testing of a heart assist device with pulsatile flow pump
下载PDF
导出
摘要 目的:在体外模拟循环实验台测试搏动式血泵的流体力学特性,并通过动物实验论证其可行性,为进一步的临床性研究提供依据。方法:分别利用水和林格氏液作为循环介质,模拟体外循环条件,观察体外模拟循环试验前、后负荷的变化对输出量的影响,并通过动物实验加以论证。结果:在固定前负荷不变的情况下,增大后负荷可使心输出量逐渐下降,而保持后负荷不变的情况下,逐渐加大前负荷,输出量增加不明显。设定前负荷的改变范围为10~50 mmHg,后负荷的改变范围为60~110 mmHg,测出相应输出量的变化为4.0~5.9 L。后期动物实验结果表明该搏动式血泵对心衰猪的心脏功能有一定的辅助作用。结论:搏动式血泵对后负荷增加表现出明显的流量抑制效应,而对前负荷反应则不明显,动物实验亦证实了其有一定的可行性。 Objective:To test the pulsatile blood pumps in a mock circulation loops, and then to prove its feasibility by animal experiments. Methods: In the mock circuit, the water and Ringerrs solution were used respectively as the circulating mediums to test the performance of the heart assist devices with pulsatile flow pumps, and to prove it by animal experiments. Results:When preload was fixed, cardiac output was gradually decreased with the increase in afterload. While afterload was kept unchanged, cardiac output was almost not affected by the increase in preload. When preload was set between 10 mmHg and 50 mmHg, and afterlood was between 60 mmHg and 110 mmHg, we observed that cardiac output was changed between 4.0 L and 5.9 L. Animal experiments showed that pulsatile blood pump had an supporting role in pigs with heart failure. Conclusion:The study demonstrated that cardiac output of the pulsatile blood pump is significantly restrained by afterload while preload exerts no influence on it. Further investigation is needed to explore the significance of the difference in the load response characteristics.
出处 《国际心血管病杂志》 2013年第3期182-184,共3页 International Journal of Cardiovascular Disease
基金 2010年国家自然基金专项基金(81027001) 上海交通大学医工(理)交叉研究基金(YG2011MS56)
关键词 搏动式血泵 心脏辅助装置 体外循环 Pulsatile blood pumps Heart assist devices In-vitro mock circulation
  • 相关文献

参考文献8

  • 1Stewart GC,Givertz MM. Mechanical circulatory support foradvanced heart failure: patients and technology in evolution[J]. Circulation,2012,125(10) : 1304-1315.
  • 2Bellumkonda L,Bonde P. Ventricular assist device therapy forheart failure—past, present, and future[J]. Int AnevsthesiolClin, 2012’ 50(3) : 123-145.
  • 3邬顺捷,杨明,黄欢,李虹磊.一种非血液接触式气动心室辅助装置的设计及体外测试[J].中国医疗器械杂志,2011,35(6):398-401. 被引量:8
  • 4Kilic A, Ailawadi G. Left ventricular assist devices in heartfailure [J]. Expert Rev Cardiovasc Ther,2012,10 (5):649-656.
  • 5Hsu KH, Huang SC,Chou NH, et al. Ventricular assist deviceapplication as a bridge to pediatric heart transplantation: asingle center's experienced]. Transplant Proc,2012,44(4):883-885.
  • 6Loor G, Gonzalez-Stawinski G. Pulsatile vs. continuous flowin ventricular assist device therapy[J]. Best Pract Res ClinAnaesthesiol, 2012,26(2): 105-115.
  • 7Nakatani T. Ventricular assist systems and heart transplan-tationQ]. Nihon Rinsho*2011,69(Suppl 9) : 4334-4340.
  • 8Silvestry S. New developments in the surgical management ofend-stage heart failure[J]. Mo Med, 2012,109(4) : 288-294.

二级参考文献8

  • 1郭静萱,李海燕.慢性心力衰竭的诊治进展[J].中国实用内科杂志,2007,27(1):11-13. 被引量:159
  • 2DeBakey ME, The Odyssey of the artificial heart [J], Artif. Organs, 2000, 3(24): 405-411.
  • 3Deng MC, De Meester JM, Smits JM, et al. Effect of receiving a heart transplant: analysis of a national cohort entered on to a waiting list, stratified by heart failure severity [J], Comparative Outcome and Clinical Profiles in Transplantation (COCPIT) Study Group. BMJ, 2000, 321(7260): 540-545.
  • 4杨明,刘一,李世阳,等.气动直接心室辅助装置[P].ZL200810037403.8[P],2008-09-24.
  • 5D. Timms, M. Hayne, K. McNeil, et al. A complete mock circulation loop for the evaluation of left, right, and biventricular assist devices [J], Artif. Organs, 2005, 29(7): 564-572.
  • 6Y. Liu, P. Allaire, Y. Wu, H. Wood, et al. Construction of an artificial heart pump performance test system [J], Cardiovasc. Eng., 2006, 6: 151-158.
  • 7刘一,杨明.直接心脏辅助装置的现状与未来[J].中国医疗器械杂志,2009,33(1):36-39. 被引量:3
  • 8李虹磊,杨明,李世阳.用于评估心室辅助装置的人体循环系统半实物仿真模型[J].中国医疗器械杂志,2010,34(1):16-20. 被引量:5

共引文献7

同被引文献21

引证文献4

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部