摘要 目的分析基层医院纯音听阈测定结果错误的特点及原因。方法对基层医院转诊的患者进行纯音听阈测定、耳科检查及声导抗测试,对基层医院纯音听阈测定的误诊进行分析。结果 124例患者(145耳)中有18耳纯音听阈测定结果错误,其中传导性聋准确率较低,错误掩蔽(包括掩蔽不足和未掩蔽)比有效掩蔽的准确率低。结论基层医院纯音听阈测定存在一定的误诊,需要加强检测人员对于掩蔽的培训。 Objective To analyse the accuracy of pure tone audiometry in primary hospital. Methods Review the pure tone audiometry of the patients from primary hospital, retested pure tone audiometry, otology examination and tympanometry. Results 18 ears had wrong pure tone audiometry tests, conductive deafness had lower accuracy, inappropriate masking had lower accuracy. Conclusion There was an error of the pure tone audiometry in primary hospital, testing personnel needs training of masking.
出处 《中国实用医药》 2013年第12期33-34,共2页 China Practical Medicine
关键词 纯音听阈测定 掩蔽 Pure tone audiometry Masking
  • 相关文献

参考文献5

  • 1Hood J. The principles and practices of bone-conduction audiome- try. Laryngoscope, 1960,70 : 1211-8.
  • 2Fernandes KC, Russo IC. Clinical masking: applicability of plateau and optimized methods in hearing thresholds testing. Pro Fono, 2009,21 (4) :333-8.
  • 3Turner R. Masking redux I: an optimized method. Journal Ameri- can Academy of Audiology, 2004,15 : 17-28.
  • 4Turner R. Masking redux II: a recommended masking protocol. Journal American Academy of Audiology, 2004,15:29-46.
  • 5Tadashi Uchiyama, Sayaka Kanazawa, Ikuyo Ishizaka, et al. Use of insert earphone with an open ear tip and reference level in the bone conduction noise method as a test of bone conduction thresh- old. Audiology Japan, 2012,55: 223-229.