老年人埋藏式心律转复除颤器误放电原因分析及处理

Analysis of causes of inappropriate discharge of an implantable cardioverter defibrillator in elderly patients
原文传递
导出
摘要 目的 分析老年患者埋藏式心律转复除颤器(ICD)误放电发生原因并探讨相应的处理方法. 方法 对植入ICD的患者进行随访,应用体外程控仪记录误放电发生时的事件情况并进行分析. 结果 95例患者共随访235例次/年,出现ICD误放电16例,误放电原因包括心室误感知7例(43.7%),室上速误认为室性心动过速4例(25.0%),其他误放电5例(31.3%),包括ICD电极磨损2例(12.5%),肌电干扰、肌肉电刺激仪、电极磨损合并肌电干扰各1例(各占6.3%).经程控及无创处理后可减少及避免再次误放电11例,其中1例经程控后出现感知低下,未能正确识别心室颤动,经体外除颤后转复;余5例经有创处理好转. 结论 植入ICD患者可出现多种类型的误放电,可根据不同ICD进行相应参数调整,在保证ICD感知灵敏度的同时减少误放电. Objective To analyze the causes of inappropriate discharge of implantable cardioverter defibrillator (ICD) in elderly patients and to discuss the corresponding solutions.Methods Totally 95 elderly patients with ICD were collected.They were followed up and the inappropriate ICD discharges were recorded and analyzed.Results All patients were followed up for 235 times/year totally.There were 16 patients with inappropriate ICD discharges.The causes of inappropriate ICD discharges included ventricular oversensing in 7 patients (43.7%),supraventricular tachycardia misclassified into ventricular tachycardia in 4 patients (25.0%),electrode fracture in 2 patients (12.5%),and electromyogram (EMG) interference,electromagnetic interference and electrode fracture with EMG interference in 1 patients respectively (6.3% each).11 inappropriate ICD discharges were reduced and avoided by reprogrammed and noninvasive treatment in 16 patients.1 patient with the underdetected ventricular fibrillation after reprogramming and the problem was solved by external defibrillation.The other 5 patients got better by the invasive treatment.Conclusions Different causes account for inappropriate discharges in patients with ICD.Reprogramming the corresponding parameters in different ICD can reduce or avoid inappropriate discharges without affecting the sensitivity of ICD.
出处 《中华老年医学杂志》 CAS CSCD 北大核心 2013年第5期465-468,共4页 Chinese Journal of Geriatrics
基金 国家自然科学基金资助项目(81100067)
关键词 除颤器 植入型 心律失常 Defibrillators,implantable Arrhythmias
  • 相关文献

参考文献9

二级参考文献46

  • 1Gregoratos G, Abrams J, Epstein AE, et al. ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices (ACC/AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines). Circulation ,2002,106:2145-2161.
  • 2Swedberg K, Cleland J, Dargie H, et al. Guidelines for the Diagnosis and Treatment of Chronic Heart Failure: executive summary (update 2005 ) :The task force for the diagnosis and treatment of chronic heart failure of the european society of cardiology. Eur Heart J ,2005,26 : 1115-1140.
  • 3Connolly S J, Gent M, Roberts RS, et al. Canadian Implantable Defibrillator Study(CIDS) :A randomized trial of the implantable cardioverter defibrillator against amiodarone. Circulation, 2000, 101 : 1297-1301.
  • 4The Antiarrhythmics Verus Implantable Defibrillaton (AVID)Investigators. A comparison of Antiarrhythmic2drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. N Engl J Med, 1997,337:1576-1583.
  • 5Kuck KH, Cappao R, Siebels J, et al. Randomized comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resudcitated from cardiac arrest. The cardiac arrest study Hamburg ( CASH ). Circulation, 2000,102 : 748 -753.
  • 6Bardy GH, Lee KL, Mark DB, et al. Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) Investigators. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med ,2005,352:225-237.
  • 7Feldman AM, de Lissovoy G, Bristow MR, et al. Cost effectiveness of cardiac resynchronization therapy in the Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure( COMPANION ) trial. J Am Coil Cardiol,2005 ,46 :2311-2321.
  • 8ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: Executive Summary. J Am Coil Cardiol, 2008,51 : 2085 -2105,.
  • 9Brown DW, Croft JB, Greenlund KJ, et al. Trends in hospitalization for the implantation of cardioverter-defibrillators in the Uniled States, 1990-2005. Am J Cardiol,2008,101 : 1753-1755.
  • 10Birnie D,Tung S,Simpson C,et al.Complications associated with defibrillation threshold testing:the Canadian experience.Heart Rhythm,2008,5:387-390.