期刊文献+

心肌显像剂^(99)Tc^mN(NOEt)_2放化纯度的分析

Measuring of Radiochemical Purity of Myocardial Imaging Agent ^(99)Tc^mN(NOEt)_2
下载PDF
导出
摘要 采用单层析与双层析法分别对新型心肌显像剂 99Tcm N(NOEt) 2 的放化纯度进行了分析。结果表明 ,双层析法对 99Tcm N(NOEt) 2 及杂质的分离较为迅速、准确 ,可作为 99Tcm N(NOEt) 2 放化纯度分析的较佳方法。 The radiochemical purity of 99 Tc mN(NOEt) 2 are measured by monolayer and bilayer chromatography respectively. The results show that the bilayer chromatography can seperate the impurity from 99 Tc m(NOEt) 2 more quickly and correctly. It is a better method for measuring the radiochemical purity of 99 Tc mN(NOEt) 2.
出处 《同位素》 CAS 2000年第1期25-27,共3页 Journal of Isotopes
关键词 心肌显像剂 双层析法 放化纯度 Tc99NCNOEth myocardial imaging agent 99 Tc m N(NOEt) 2 monolayer chromatography bilayer chromatography radiochemical purity
  • 相关文献

参考文献2

  • 1 Pasqualini R, Duatti A,BellandeE, et al. Bis(dithiocarbamato) Nitrido 99mTc Radiopharmaceuticals: a Class ofNeutral Myocardial Imaging Agents[J].J Nucl Med, 1994, 35:334.
  • 2 Abram U, Lorenz B, Kaden L, et al. Nitrido Complexes of Technetium With TertiaryPhosphines and Arsines[J]. Polyhedron, 1998, 7:285. 修回日期:1999-10-20