期刊文献+

CT自适应心率变速扫描研究 被引量:1

Study of CT Variable Scanning Velocity with Adaptive to Heart Rate
下载PDF
导出
摘要 针对心脏运动导致CT图像的模糊,提出了自适应心率的CT变速扫描方法。该方法可根据心率和所需要的时间分辨率自适应选择最优的扫描速度,获得理想的投影数据。为了验证该模型,通过分段线性和非线性变形模型仿真心脏运动,利用滤波反投影重建算法进行了心脏图像重建。重建结果表明:根据心率选择合适的扫描速度可得到理想的投影数据,有效地提高了心脏图像质量,从而为CT心脏成像提供新的模式。 In order to aim at CT image blur due to heart motion,the variable velocity scanning model based on the heart rate and temporal resolution was proposed.The optimal scanning velocity could be obtained by the above model to achieve ideal projection data.For verifying this approach,the heart motion pattern with piecewise linear and the non-linear deformation model was given to imitate the heart motion.The cardiac image was reconstructed by filtered back-projection algorithm.The reconstruction indicates that ideal projection data can be obtained by selected scanning velocity so as to achieve high quality cardiac image and provide new mode for cardiac CT imaging.
作者 刘杰
出处 《系统仿真学报》 CAS CSCD 北大核心 2013年第6期1290-1295,共6页 Journal of System Simulation
基金 国家自然科学基金(30970777/H1804)
关键词 心脏CT 心率 扫描速度 重建 cardiac computerized tomography heart rate scanning velocity reconstruction
  • 相关文献

参考文献1

二级参考文献8

  • 1毛海鹏,张定华,梁亮,赵歆波,李山.一种基于PC的快速三维图像重建方法[J].系统仿真学报,2004,16(11):2486-2489. 被引量:11
  • 2孙丰荣,刘泽,李艳玲,曲怀敬,张梅,张运.基于模型的CT三维医学图像重建仿真[J].系统仿真学报,2006,18(3):781-784. 被引量:14
  • 3Robert M. Cardiac Cone-beam CT [J]. Medical Physic (S00942405), 2005, 3200): 3227-3229.
  • 4Nielsen T, Manzke R, Proksa R. Cardiac Cone-beam CT Volume Reconstruction Using ART [J]. Medical Physics (S00942405), 2005, 32(4): 851-854.
  • 5Manzke R, Grass M, Nielson T. Adaptive Temporal Resolution Optimization in Helical Cardiac Cone Beam CT Reconstruction [J]. Medical Physics (S00942405), 2003, 30(12): 3072-3080.
  • 6Wang G, Zhao S Y, Heuscher D. A Knowledge-based Cone-beam X-ray CT Algorithm for Dynamic Volumetric Cardiac Imaging [J]. Medical Physics (S00942405), 2002, 29(8): 807-810.
  • 7Hui H, Tinsu P, Yun S. Multi-slice Helical CT: Image Temporal Resolution [J]. IEEE Transaction on Medical Imaging (S0278-0062), 2000, 19(5): 384-486.
  • 8高尚凯.医学成像系统[M].北京:清华大学出版社,2000.

同被引文献26

  • 1高文洁,霍健伟,窦秀梅,穆锦巍,李霞.螺旋CT冠状动脉造影心率控制方法探讨[J].中国心血管病研究,2006,4(10):754-755. 被引量:4
  • 2Schaller S, Folhr T, Klingenbeck K, et al. Spiral interpolation algorithm for muhislice spiral CT-part Ⅰ: theory [ J]. IEEE Trans Med Imaging,2000,19 (9) :822-834.
  • 3Goua H, Varenne O, Trinquart L, et al. Coronary artey stenosis in high-risk patients: 64- section CT and coronary angiography- Prospective study and analysis of discordance [ J ]. Radiology, 2010,252(2) :377-385.
  • 4Mark DB, Berman DS, Budoff MJ, et al. ACCF/ACR/AHA./ NASCL/SAIP/SCAL/SCCT 2010 expert consensus document on coronary computed tomographic angiography: a report of the american college of cardiology foundation task force on expert consensus documents [ J ]. Catheter Cardiovasc Interv, 2010,76 (2) :E1-42.
  • 5Uetani T, Amano T, Kunimura A,et al. The association between plaque characterization by CT angiography and post-procedural myocardial infarction in patients with elective stent implantation [J]. JACC Cardiovasc Imaging,2010,3( 1 ) :19-28.
  • 6Carrascosa PM, Capunay CM, Garcia-Merletti P, et al. Characterization of coronary athemsclemtic plaques by multidetector computed tomography [ J ]. Am J Cardiol, 2006,97(5) :598-602.
  • 7Steigner ML, Otero HI, Cai T,, et al. Narrowing the phase window width in prospectively ECC/:gated single heart beat 320 - detector row coronary CT angiagraphy [ J ]. Int J Cardiovase Imaging,2009,25 ( 1 ) :85-90.
  • 8Rybicki FJ, Otero HJ, Steigner ML, et al. Initial evaluation of coronary image from 320-detector row computed tomography [ J ]. Int J Cardiovasc Imaging,2008,.24(5) :535-546.
  • 9Sabarudin A, Sun Z, Ng KH. Radiation dose in coronary CT angiography associated with prospective ECG-triggefing technique: comparisons with different CT generations [ J]. Radiat Prot Dnsimetry ,2013,154 ( 3 ) :301-307.
  • 10Jiang B, Gai L, Sun Z, et al. The combination of 64 multislice CT angiograpby and optical coherence tomography optimally characterizes coronary plaques [ J ]. Acta Cardiol, 2011,66 ( 2 ) : 213-218.

引证文献1