摘要 从接种液种类、空白 BOD5大小、接种液最佳添加范围等方面对 BOD5测定进行探讨 ,提出以接纳工业废水的河水作为最佳接种液 ,雨季时以土壤浸取液更适合。探讨了接种液空白值对 BOD5标液和工业废水的影响。结果表明 ,河水空白应控制在 0 .1~0 .8mg/ L,土壤浸取液空白控制在 0 .0 1~ 0 .1 0 mg/ L为宜。 s:Effects of the inoculated solution type,the BOD 5 blank and the inoculated solution volume on BOD 5 determination was studied.The results showed that the river water,which took in industrial wastewater,was suitable inoculated solution in general,while the soil extraction was much better in raining time.The influences of inoculated solution blank on the BOD 5 standard solution and industrial wastewater were studied also.The data obtained indicated that the blank of river water and soil extraction should be controlled within 0 1~0 8 mg/L and 0 01~0 10 mg/L respectively.
作者 谢美云
出处 《环境污染与防治》 CAS CSCD 北大核心 2000年第5期39-40,共2页 Environmental Pollution & Control
关键词 接种稀释水 BOD5 废水监测 稀释法 样品 Inoculated diluting water BOD 5 Determination