期刊文献+

流动注射-分光光度法测定水中氰化物 被引量:8

Determination of Cyanide in Water by Flow Injection-Spectrophotometry
下载PDF
导出
摘要 使用流动注射(FIA)-分光光度法测定水中的氰化物的含量,并与传统分光光度法的分析结果进行比对。实验证明流动注射(FIA)-分光光度法操作简便、线性好,灵敏度、精密度、准确度都能符合分析工作要求。检出限为0.2μg/L,适用于水中微量氰化物的检测。分析频率为每小时30个样品,特别适合大批样品的测定。 We used the flow injection -spectrophotometry to determine the amount of cyanide in water, and compared the results with that of the traditional spectrophotometry. By experiments, the method was proved to be easy for operating and have a good linearity. The sensitivity, precision and accuracy of this method can meet analysis requirements. Detection limit of the method was found to be 0.2μg/L,which is suitable for the determination of cyanide in water. The sampling rate is 30 samples per hour and especially suitable for analysis of a large number of water samples.
作者 付生录 魏燕
出处 《中国无机分析化学》 CAS 2013年第2期39-42,共4页 Chinese Journal of Inorganic Analytical Chemistry
关键词 流动注射分析 氰化物 分光光度法 方法研究 flow injection analysis cyanide spectrophotometry method study
  • 相关文献

参考文献5

  • 1国家环保局.水和废水K测分析方法[M].北京:中国环境科学出版社,2002:144-155.
  • 2李春艳,申贵隽,李建华.数码比色法测定环境及生物样品中的氰化物浓度[J].中国无机分析化学,2012,2(2):24-26. 被引量:7
  • 3方肇伦.流动注射分析法[M].北京:北京科学出版社,2003.
  • 4国家环境保护总局.HJ484-2009水质氰化物的测定[S].北京:国环境科学出版社,2009.
  • 5JOHANNES C L.Spectmphotomaric determination oftotal cyanide,iron-cyanide complexes,free cyanide andthiocyanate in water by a continuous-flow system[J].Analyst,1989,114:959-963.

二级参考文献7

共引文献7

同被引文献37

引证文献8

二级引证文献34

相关作者