摘要 针对目前国外对非接触眼压测量技术的垄断,以及国内现存技术的不足,研制了一种新型的基于力平衡原理和空气静压技术的非接触式眼压测量系统。该系统以ARM9S3C2440A为核心处理器,主要包括气浮测量模块、控制模块、人机交互及通讯接口等模块,其中气浮式测量模块能够对CCD摄像头信号、压力传感器信号、角膜反射的光电信号进行采集和测量,控制模块实现中心气腔供气压力的控制。利用研制的系统对人体眼压进行测量,结果表明眼压测量的重复性误差≤6.01%,与进口仪器相比眼压测量的绝对误差≤1.3mmHg,表明该系统能够快速、方便地检测出人体的眼压值。 For the foreign monopoly of current non-contact tonometry technology and the disadvantage of domestic existing technology,we develop a new type of non-contact tonometry system base on equilibrium principle and aerostatic technology.The system adopts ARM9 S3C2440A as the core,mainly including flotation measurement module,control module,human-computer interaction and communication interface module.The flotation measurement module could finish the signal acquisition and measurement on CCD camera,pressure sensor,and optical signals of corneal reflection that the control module mainly controlled the supply gas pressure of centered gas chamber.When the developed system is used to measure human intraocular pressure,the measured values show that there is only less than 6.01% repeatability error.Compared with the imported non-contact tonometry,the absolute error between two equipments' measured IOP values is also no more than 1.3m mHg,which shows that the system could detect the intraocular pressure of people quickly and easily.
出处 《重庆大学学报(自然科学版)》 EI CAS CSCD 北大核心 2013年第6期53-59,共7页 Journal of Chongqing University
基金 国家自然科学基金资助项目(30970764)
关键词 眼压 非接触式 力平衡原理 空气静压技术 传感器 IOP non-contact measurement equilibrium principle aerostatic technology sensor
  • 相关文献

参考文献14

  • 1干江红,周莉萍,徐龙.眼内压力测量技术的研究[J].中国医疗器械杂志,2009,33(3):202-206. 被引量:4
  • 2Kanngiesser H E, Kniestedt C,Robert Y C. Dynamiccontour tonometry: presentation of a new tonometer[J]. Journal of Glaucoma?2005 ,14(5): 344-350.
  • 3Lam A K,Chan R,Lam C H. The validity of a newnoncontact tonometer and its comparison with theGoldmannn tonometer [J]. Optometry and VisionScience,2004,81(8):601-605.
  • 4Pepose J S,Feigenbaum S K,Qazi M A,et al. Changesin corneal biomechanics and intraocular pressurefollowing LASIK using static, dynamic and noncontacttonometry [j]. American Journal of Ophthalmology,2007,143(1):39-47.
  • 5Lam A, Chen D, Chiu R, et al. Comparison of IOPmeasurements between ORA and GAT in normalChinese[J]. Optometry and Vision Science,2007,84(9):909-914.
  • 6Kempf R,Kurita Y,Lida Y,et al. Understanding eyedeformation in non-contact tonometry[C] // Proceedingsof the 28th Annual International Conference of theIEEE Engineering Medicine and Biology Society,August 30-September 3,2006,New York,NY.Piscataway:IEEE Press,2006 ,1:5428-5431.
  • 7Han Y M,Bryanston-Cross P J,Lee W K,et al. Activenon-contact tonometer for glaucoma detection [ J].Optics in Health Care and Biomedical Optics:Diagnostics and Treatment ?2002,4916: 143-150.
  • 8Ogbuehi K C? Almubrad T M. Accuracy and reliabilityof the Keeler Pulsair EasyEye non-contact tonometer[J]. Optometry and Vision Science, 2008? 85( 1 ):61-66.
  • 9Gloster J, Perkins E S. The validity of the imbert'ficklaw as applied to applanation tonometry [ J].Experimental Eye Research, 1963 ,2(3): 274-283.
  • 10Hsu H. Non-contact high frequency tonmeter: US,4928697 A[P]. 1990-05-29.

二级参考文献26

  • 1段宣初,吴勤,蒋幼芹,卿国平,江冰,石晶明.中央角膜厚度对压平眼压计测量值的影响[J].中国实用眼科杂志,2004,22(10):778-782. 被引量:13
  • 2王溪,窦宏亮,高富军.昼夜眼压波动及其在青光眼治疗中的意义[J].国外医学(眼科学分册),2005,29(3):187-190. 被引量:6
  • 3Andrzej Hachol,Marta Grabowska.Investigation of dynamic parameters of ophthalmic tonometric transducers[J].Proc.SPIE 2002,4887:124-131.
  • 4Matteo Leonardi,Peter Leuenberger,Daniel Bertrand,et al.First Steps toward Noninvasive Intraocular Pressure Monitoring with a Sensing Contact Lens[J].Investigative Ophthalmology & Visual Science,Sep.2004,45,(9):3113-3117.
  • 5Schmidt Theo AF.The clinical Application of the Goldmann Applanation Tonometer[J].AMJ Ophthalmol,1960,49-967.
  • 6Marcio Marc Abreu.Apparatus and Method for Noninvasive Measurement of Analytes from the Conjunctiva using Mid-Infraded Radiation.United States Patent[J],US7137952B2,Jul.22,2008.
  • 7Mohammed K.EIMallah,Sanjay G.Asrani.New ways to measure intraocular pressure[Jl.Curr Opin Ophthalmol,2008,19:122-126.
  • 8Kalyan C.Katuri,Sanjay Asrani,Melur K.Ramasubramanian.Intraocular Pressure Monitoring Sensors[J].IEEE Sensor Journal,2008,8(1):13-19.
  • 9Matteo Leonardi,Peter Leuenberger,et al.Intraocular pressure recording system[J].United States Patent,US7137952B2,Nov.21,2006.
  • 10Wilfried Mokwa,Uwe Schnakenberg.Micro-Transponder Systems for Medical Applications[J].IEEE Transactions on Instrumentation and Measurement,Dec.2001,50(6):1551-1555.

共引文献28

同被引文献17

引证文献3

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部