摘要 采用水杨酸-次氯酸盐光度法,研究了pH对KS2301氨氮在线监测仪测量结果的影响。研究结果表明,酸碱度对仪器测量结果的准确性均有影响,其中,酸度的影响较碱度的影响要大。当PH为11.6~11.7时,显色剂的显色效果最好,仪器在这个值附近测量结果最稳定。 This paper is about the influence of pH value in the measuring results of ammonia nitrogen online monitoring instruments with method of Salicylic acid-hypochlorite.The results show that the pH value does have an effect on the accuracy of measurement results,from which we can conclude that the impact of acidity is greater than that of alkalinity.When the pH value is between 11.6 and 11.7,the colored effect of the reagent is best,and the measuring results of instruments are most stable.
作者 杨凯 李秀芳
出处 《中国环境监测》 CAS CSCD 北大核心 2013年第3期123-127,共5页 Environmental Monitoring in China
基金 国家水环境监测技术体系研究与示范基金资助项目(2009ZX07527-007)
关键词 酸碱度 氨氮在线监测仪 影响 pH ammonia nitrogen online monitoring instruments impact
  • 相关文献

参考文献15

  • 1HJ536-2009水质.氨氮的测定:水杨酸分光光度法[S].
  • 2胡小玲,吴鹏.水杨酸—次氯酸盐测定水中氨氮方法的改进[J].干旱环境监测,2005,19(3):184-185. 被引量:21
  • 3温丽云,范朝,袁倬斌.我国环境监测中的氨氮分析方法[J].中国环境监测,2005,21(4):28-32. 被引量:75
  • 4罗澍,严少红,黄远峰,侯化明.四种氨氮自动监测仪与实验室国标方法测定氨氮的比较分析[J].中国环境监测,2010,26(3):32-36. 被引量:27
  • 5Hansen, H.P. and Grasshoff, K.. Automated determination of ammonia. In: K. Grasshoff, M. Ehrhardt and F. Kremling(Editors) , Methods of Seawater Analysis [ J ]. Verlag Chemie, Weinheim, 1983,2 : 363- 365.
  • 6蒋岳文,陈淑梅,马英.靛酚蓝分光光度法测定海水中氨氮最佳条件的选择[J].海洋环境科学,1997,16(4):43-47. 被引量:15
  • 7A. Aminot, D. S. Kirkwood, R. Keroue. Determination of ammonia in seawater by the indophenol-blue method: Evaluation of the ICES NUTS I/C 5 question-naire[J]. Marine Chemistry , 1997,56:59-75.
  • 8Abarnou, A. and Miossec, L. Chlorinated waters discharged to the marine environment chemistry and environmental impact - An overview [ J ]. Sci. Total Environ., 1992,126: 173-197.
  • 9Koroleff, F.. Direct determination of ammonia in natural waters as indophenol blue. In: ICES, Information on Techniques and Methods for Seawater Analysis, An Interlaboratory Report [ R]. ICES C. M. , 1970, 3: 19-22.
  • 10Gravitz, N. and Gleye, L. ,A photochemical side reaction that interferes with the phenolhypochlorite assay for ammonia [ J]. Limnol. Oceanogr. , 1975, 20 (6) : 1 015-1 017.

二级参考文献46

共引文献123

同被引文献69

引证文献6

二级引证文献23

相关作者