摘要
采用滴涂法制备了壳聚糖修饰石墨圆盘电极,并用于L-抗坏血酸(LAA)的半微分伏安法测定。研究了电解液种类及酸度、膜厚度、富集电位及时间、扫速和温度等因素对伏安曲线的影响,获得了较为优化的测试条件。在0.20mol/L NaH2PO4-Na2HPO4(pH6.0)缓冲液中,LAA的浓度在 1. 0×10-5~ 6. 0 × 10-4 mol/L范围内与氧化峰电流呈良好线性关系( r=0.9997),检测限达6.7×10-7mol/L;生物体内常见的干扰物质如亚硝酸盐、多巴胺、葡萄糖等不干扰测定。 9次测得电极在含 5.0 ×10-5 mol/L LAA试液中峰电流的 RSD为3.6%,回收率为96.4%~102.8%( n=5)。已成功用于维生素C片剂和人血清中 LAA的测定。
Graphite disc electrode modified with chitosan by drop-coationg method was used to determine L-ascorbic acid (ICA) by semi-differential voltammetry. The effect of the type and acidity of electrolytes, the thickness of chitosan membrane, enrichment potential and time, scan late and temperature on the voltammetric curve of LAA were investigated to acquire optimun conditions.In 0. 20 mol/L NaH2PO4-Na2HPO4-HPO4 buffer (pH 6. 0), there was an excellent linearity between oxidation current and concentration of L-ascorbic acid in the range from 1.0 ×10-6 to 6.0 × 10-4 mol/L (r = 0.9997) with a detection limit of 6. 7 ×10-7 mol/L.Some ordinary biological substances such as nitrite, dopamine,glucose, etc. did not interfere with the determination of LAA. Relative standard deviation (RSD) Of the oxidation peak currents obtained from 9 determinations of the same solution containing 5 .0 ×10-5 LAA was 3.6%. Recoveries of 96.4% -102. 8% were obtained from five determinations. This method has been applied to the determination of LAA in tablets and human serum successfully.
出处
《分析化学》
SCIE
EI
CAS
CSCD
北大核心
2000年第10期1220-1223,共4页
Chinese Journal of Analytical Chemistry
关键词
壳聚糖
半微分伏安法
L-抗坏血酸
测定
石墨电极
Chitosan, edified electrode, Send-differential voltammetry, L-ascorbic acid