期刊文献+

顶空气相色谱法测定一次性使用无菌导尿管中环氧乙烷残留量 被引量:2

Determination of Ethylene Oxide Residual Quantity in Single-Use Sterile Urethral Catheter by Headspace Gas Chromatography
下载PDF
导出
摘要 目的建立一次性使用无菌导尿管中环氧乙烷残留量的测定方法。方法采用顶空进样毛细管气相色谱法,FID检测器,应用P/NAgilent 19091J-413 HP-5毛细管柱(30.0 m×320μm,0.25μm),载气为氮气,柱温采取恒温80℃、保持4 min,测定一次性使用无菌导尿管中环氧乙烷的残留量。结果环氧乙烷质量浓度在0.523 4~16.748 2μg/mL范围内与峰面积呈良好的线性关系,回归方程Y=0.462 8 X-0.048 39,r=0.999 97(n=6),平均回收率为92.56%,RSD为0.72%(n=6)。结论该方法快速简便、结果准确、可靠、重现性好,可作为一次性使用无菌导尿管中环氧乙烷残留量的测定方法。 Objective To establish the method for the determination of ethylene oxide residual quantity in the single-use sterile urethral catheter.Methods The ethylene oxide residual quantity in the single-use sterile urethral catheter was tested by headspace gas chromatography with FID detector,the P / N Agilent 19091 J-413 HP-5 capillary column(30.0 m × 320 μm,0.25 μm) was adopted with N 2 as the carrier gas.The column temperature was kept at the constant temperature of 80 ℃ for 4 min.Results The mass concentration of ethylene oxide in the range of 0.523 4-16.748 2 μg / mL showed the good linear relation with the peak area.The regression equation was Y = 0.462 8 X-0.048 39,r = 0.999 97( n = 6).The average recovery rate was 92.56%,RSD was 0.72%( n = 6). Conclusion This established method is rapid,simple,accurate and reliable with good reproducibility,and can be used as the determination method for the ethylene oxide residual quantity in single-use sterile urethral catheter.
作者 张成 王刚林
出处 《中国药业》 CAS 2013年第14期61-62,共2页 China Pharmaceuticals
关键词 顶空气相色谱法 一次性使用无菌导尿管 环氧乙烷残 留量 headspace gas chromatography single-use sterile urethral catheter ethylene oxide residual quantity
  • 相关文献

参考文献4

  • 1GBl5980-1995:一次性使用医疗用品卫生标准[S].北京:中华人民共和国卫生部,1995.
  • 2GB15979-2002.一次性使用卫生用品卫生标准[S].[S].,..
  • 3GB/T16886.7-2001,医疗器械生物学评价第7部分:环氧乙烷灭菌残留量[S].
  • 4GB/T14233.1-2008.医用输液、输血、注射器具检验方法第1部分:化学分析方法[S].

共引文献42

同被引文献8

引证文献2

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部