γ辐射场中TEPC微剂量谱的测量与模拟 被引量:2

Measurement and Simulation of TEPC Microdosimetric Spectra in Gamma Radiation Field
下载PDF
导出
摘要 组织等效正比计数器(TEPC)所测中子、γ辐射的微剂量谱体现出不同的特征,可对中子、γ混合辐射场剂量当量份额进行区分。对中子、γ辐射剂量当量分辨方法进行研究,在60 Co、137 Csγ参考辐射场测量纯γ辐射的微剂量谱。采用FLUKA程序,模拟TEPC在多个单能点γ辐射场的微剂量谱。结果表明,60 Co、137 Csγ参考辐射场测量的微剂量谱与蒙特卡罗模拟的微剂量谱均体现了γ辐射的特征即线能基本低于10keV/μm,是进行中子、γ辐射剂量当量贡献分辨的基础。 According to the different characteristics of microdosimetric spectra measured by tissue equivalent proportional counter (TEPC), the neutron dose equivalent and γ dose e- quivalent could be distinguished in a unknown neutron and γ mixed radiation field. In order to discriminate the γ radiation dose equivalent from the total value,the pure γmicrodosimet- ric spectra was measured in60Co, 137Cs radionuclide radiation field with TEPC. TEPC mi- crodosimetric spectra in a series of monoenergy γ radiation field were simulated by FLUKA code. All the 3' radiation microdosimetric spectra, including measured spectrum in 60Co, 137Cs radiation field and that of simulation spectrum by FLUKA code, reveal a trait that the linear energy of γ radiation is basically lower than 10 keV/tam. This trait is the very foun- dation to discriminate the γ radiation from the mixed radiation.
出处 《同位素》 CAS 2013年第3期144-147,共4页 Journal of Isotopes
关键词 组织等效正比计数器 微剂量谱 Γ辐射 FLUKA程序 TEPC microdosimetric spectra γ radiation FLUKA code
  • 相关文献

参考文献9

  • 1Menzel HG, Schuhmacher H. Experimental Un- certainties and Applicability of the Proportional counter Method for the Dertermination of Photon Dose Fractions in Neutron Beams [R]//In.. Ion Chambers for Neutron Dosimetry. Broese: EUR6782EN, 1980: 337-343.
  • 2Kota C, Maughan RL, Tattam Da. Use of low- pressure tissue equivalent proportional counters for the dosimetry of neutron beams used in BNCT and BNCEFNT[J]. Med Phys, 2000,27(3).
  • 3Burmeister J, Kota C, Maughan RL, et al. Minia- ture tissue-equivalent proportional counters for BNCT and BNCEFNT dosimetry[J]. Med Phys, 2001,28(9).
  • 4Rossi HH, Zaider M. Microdosimetry and Its Ap- plications, Chapter Ⅱ[M]. Berlin.. Springer, Hei- dlberg, 1996.
  • 5Farahmand M. A novel Tissue-Equivalent Propor- tional Counter based on a Gas Electron Multiplier [D]. TU-Delft, 2004.
  • 6Dietze G, Menzel HG, Buhler G. Calibration of tissue-equivalent proportional counters used as ra- diation protection do dosemeters [J]. Radiation Protection Dosimetry, 1984, 9 (3) :245-249.
  • 7Schrewe UJ, Brede H J, Pihet P, et al. On the cal- ibration of tissue-equivalent proportional counters with buil tin a particle sources[J]. Radiation pro- tection dosimetry, 1988, 23 (1):249-252.
  • 8Rollet S, Beck P, Ferrari A, et al. Dosimetric considerations on tepc fluka-lation and measure- ments[J]. Radiation Protection Dosimetry, 2004, 110, (1) :833-837.
  • 9Srdoc D. Experimental technique of measurement of microscopic energy Distribution in irradiated matter using Rossi Counte[J]. Radiat Res, 1970, 43:302.

同被引文献10

  • 1ANJOMANI J, HANU A R, PRESTWICH W V, et al. Monte Carlo design study for thick gas electron multiplier-based multi-element microdo-simetric detector [ J ]. Nucl Instrum Methods Phys Res A: Accel Spectrum Detect Assoc Equip, 2014, 757(1): 67-74.
  • 2WAKER A J, SZORNEL K, NUNES J. TEPC performance in the CANDU workplace[J]. Radi- ation Protection Dosimetry, 1997, 70(1-4) : 197- 202.
  • 3BALLARINI F, BATTISTONI G, CAMPAN- ELLA M, et al. The FLUKA code: An over- view[J]. Journal of Physics: Conference Series, 2006, 41(1): 151-160.
  • 4ROLLET S, COLAUTTI P, GROSSWENDT B, et al. Monte Carlo simulation of mini TEPC microdosimetric spectra: Influence of low energy electrons[J]. Radiation Measurements, 2010, 45 (10): 1 330-1 333.
  • 5NUNOMIYA T, KIM E, KUROSAWA T, et al. Measurements of lineal-energy distributions for neutrons of 8 keV to 65 MeV by using a tissue equivalent proportional counter[J]. Radiation Pro- tection Dosimetry, 2002, 102(1): 49-59.
  • 6SI-YOUNG C, BONG-HWAN K. Understand- ing of the microdosimetric quantities obtained by a TEPC [J]. Journal of Nuclear Science and Technology, 2008(5): 213-216.
  • 7Recommendation of the International Commission on Radiological Protection, ICRP Publication 60 [R]. Sweden: International Commission on Ra- diological Protection, 1991.
  • 8FAWAZ A. Design and characterization of next- generation tissue equivalent proportional counters for use in low energy neutron fields[D]. Canada: Faculty of Energy Systems and Nuclear Science, University of Ontario Institute of Technology, 2014.
  • 9Conversion coefficients for use in radiologieal protection against external radiation, ICRP Pub- lication 74[R]. Sweden: International Commis- sion on Radiological Protection, 1996.
  • 10张伟华,王志强,肖雪夫,刘毅娜,李春娟,骆海龙.组织等效正比计数器结构优化设计[J].原子能科学技术,2014,48(5):930-933. 被引量:5

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部