CR在全身照射定位和肺部挡铅的摆位误差

Set-up error of CR in total body irradiation localization and lung shielding
原文传递
导出
摘要 目的探讨计算机x射线成像(CR)在全身照射(TBI)定位和肺部挡铅验证中的摆位误差。方法使用TOR18FG测量其相同条件下kV级和MV级x射线能量的图像质量,建立其临床应用流程,并分析摆位误差。结果在TBI治疗条件下,MV级的x射线图像的低对比度分辨率和空间分辨率远较kV级的差,但可以比较清晰地识别高对比度组织,并应用于TBI治疗。头脚方向误差为:腹背(AP)方向,左肺(0.50±1.65)cm,右肺(1.16±1.56)cm;(背腹)PA方向,左肺(1.12±2.22)cm,右肺(0.41±2.16)cm。左右两侧方向误差为:AP方向,左肺(0.81±1.19)cm,右肺(0.43±1.20)cm;PA方向,左肺(0.31±1.64)cm,右肺(0.55±1.49)cm。结论通过CR的应用,提高了TBI的治疗摆位精度。 Objective To investigate the set-up error of CR in total body irradiation localization and lung shielding. Methods TOR 18FG software was employed to measure the image quality of images at kV and MV levels. The clinical processes were established and the positioning error was analyzed. Results The low contrast resolution and spatial resolution of MV level images were much worse than those at kV level in the condition of total body irradiation, but the image at MV level could be used to identify the high contrast tissues and employed in total body irradiation. The longitudinal errors were (0. 50 ±1.65) cm for left lung and ( 1.16 ±1.56)cm for right lung in AP direction, while ( 1.12 ±.22)cm and (0. 41 ±. 16) cm respectively in PA direction. The errors of lateral were (0.81 ±1.19)cm for left lung and (0. 43 ±1.20)era for right lung in AP direction, while (0. 31 ±1.64)cm and (0. 55 ±1.49)cm respectively in PA direction. Conclusions Application of CR in total body irradiation could make positioning in treatment much easily and reduce the localization errors.
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2013年第3期303-305,共3页 Chinese Journal of Radiological Medicine and Protection
关键词 计算机成像技术 全身放射治疗 图像质量 摆位误差 Computed radiography Total body irradiation Image quality Set-up error
  • 相关文献

参考文献9

  • 1Van Dyk J, Galvin JM, Glasgow GP, et al. AAPM Report No. 17. The Physical aspects of total and half body irradiation. New York : the American Association of Physicists in Medicine, 1986.
  • 2Lyra ME, Kordolaimi SD, Salvara AL, et al. Presentation of digital radiographic systems and the quality control procedures that currently followed by various organizations worldwide. Recent Patents on Medical Imaging, 2010,2 : 5-21.
  • 3Marshall NW. An examination of automatic exposure control regimes for two digital radiography systems. Phys Med Biol, 2009, 54 ( 15 ) : 4645-4670.
  • 4Rampado O, Isoardi P, Ropolo R. Quantitative assessment of computed radiography quality control parameters. Phys Med Biol, 2006, 51(6) : 1577-1593.
  • 5Beyzadeoglu M, Oysul K, Dirican B, et al. Effect of dose-rate and lung dose in total body irradiation on interstitial pneumonitis after bone marrow transplantation. Tohoku J Exp Med, 2004, 202(4) :255-263.
  • 6彭振军,伍钢,王晔,徐宁,冯敢生.CR与屏(铜)-片成像在头颈部肿瘤放疗射野影像验证中的比较应用[J].临床放射学杂志,2006,25(10):963-965. 被引量:3
  • 7昝志生,秦维昌,亓恒涛,魏忠舜,冯德朝.CR、DR系统低密度信号检出能力比较[J].中华放射医学与防护杂志,2006,26(6):617-618. 被引量:4
  • 8Yao R, Bernard D, Turian J, et al. A simplified technique for delivering total body irradiation (TBI) with improved dose homogeneity. Med Phys,2012, 39 (4) :2239-2248.
  • 9Vano E, Geiger B, Schreiner A, et al. Dynamic flat panel detector versus image intensifier in cardiac imaging: dose and image quality. Phys Med Biol, 2005,50(23) : 5731-5742.

二级参考文献7

  • 1肖湘生.为维护医师的合法权益而努力[J].中华放射学杂志,2003,37:106-106.
  • 2European Commission.European guidelines on quality criteria for diagnostic radiographic images.EUR 16260 EN,1996,12
  • 3陈太平,彭振军,主编.X线摄影分册(医学影像技术丛书).武汉:湖北科学技术出版社,2000,1
  • 4Metz S,Damoser P,Hollweck R,et al.Chest radiography with a digital flat-panel digital detect:experimental receiver operating characteristic analysis.Radiology,2005,234:776-784.
  • 5Laundors JH,Cower AR,Bury RF.Towards image quality beam energy and effective dose optimization in digital thoracic radiography.Eur Radiol,2001,11:870-875.
  • 6Geijer H,Beckmann KW,Andersson T,et al.Image quality versus radiation dose for a flat-panel amorphous silicon detect:a phantom study.Eur Radiol,2001,11:1704-1709.
  • 7于兹喜,王昌元,徐跃,谢晋东,张梦龙,王键.计算机X线摄影的观察者操作特性曲线测试探讨[J].中华放射学杂志,2003,37(5):460-463. 被引量:8

共引文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部