低剂量扫描在多层螺旋CT下肢动脉成像中的临床研究 被引量:14

The clinical research of low dose scanning protocol for aorto-iliac and lower extremity arteries CT angiography
原文传递
导出
摘要 目的探讨多层螺旋CT(MSCT)下肢动脉成像中应用低管电压(kVp,100kV)设置和管电流调制减低辐射剂量的可行性。方法连续选取需行下肢动脉造影的患者61例,按随机数表随机分为A组(120kV)、B组(100kV)、c组(100kV+管电流调制)。比较3组患者的图像客观评价指标、图像质量主观评分以及辐射剂量。c组进-步按体重指数(BMI)分为c1组(BMI〈24.9kg/m2)及c2组(BMI≥24.9kg/m2),比较2组患者间的辐射剂量。结果3组患者的主.髂部轴位、容积再现(VR)、最大强度投影(MIP)图像的主观质量评分均在良好以上,且3组图像的信噪比、对比噪声比差异无统计学意义(P〉0.05)。A、B、c3组的有效吸收剂量分别为8.20、5.36和7.48mSv;B组较A组有效吸收剂量减低约34.6%,C组较B组有效吸收剂量增加约39.5%,且与A组比较差异无统计学意义(P〉0.05)。c1组和c2组有效吸收剂量为7.11和9.69mSv,两组比较差异有统计学意义(t=-3.163,P〈0.05),c1组较c2组有效剂量减低约26.6%。C1组较A组有效剂量减低13.3%(Z=-2.822,P〈0.05),但较B组有效剂量增加了32.6%(Z=-3.900,P〈0.05);C2组较A组有效吸收剂量增加18.2%(P〉0.05),较B组有效剂量增加了80.8%(Z=-3.426,P〈0.05)。结论下肢动脉MSCTA运用低kVp设置(100kV)能在保证图像质量的同时降低辐射剂量,其应用是可行有效的;低kVp设置(100kV)结合管电流调制技术仅对特定人群(体重指数〈24.9kg/m2)具有减低辐射剂量的价值。 Objective To study the feasibility of reducing radiation dose in aorto-iliac and lower extremity arteries CT angiography (CTA) with low tube voltage (100 kV) and automatic tube current modulation(ATCM). Methods Totally 61 patients requiring aorto-iliac and lower extremity arteries CTA for clinical reasons were prospectively enrolled in study. The patients were randomly assigned to 3 groups: Group A( 120 kV), Group B( 100 kV) and Group C( 100 kV with automatic tube current modulation). Both quantitative and qualitative analysis were included in this research. Group C was divided into obese group(BMI≥ 24. 9 kg/m2 ) and normal group ( BMI 〈 24.9 kg/m2 ). The radiation doses were analyzed respectively among two groups. Results The subjective evaluations of image quality for axial, MIP and VR were good. There were no significant differences among group A, B and C in SNR and CNR (P 〉 0. 05). The effective dose of group A, B and C was 8.20 mSv, 5.36 mSv, and was 7.48 mSv, respectively Group B was 34.6% less than group A. Group C was 39.5% more than group B, and there was no significant differences between group A and C (P 〉 0. 05 ). The effective dose of group C1 was 7.11 mSv, group C2 was 9.69 mSv, the E with group C1 were significantly less than group C2 ( t = - 3. 163 ,P 〈 0.05), the effective dose of group C1 was 13.3% less than group A(Z = -2.822,P 〈0.05), but the group C2 was more than group A(P 〉 0. 05) and group B (Z = - 3. 426,P 〈 0.05 ). Conclusions Low-kilovohage (100 kV) CT scanning protocol is feasible in multi-detector CT angiography for aortoiliac and lower extremity arteries. Automatic low tube voltage ( 100 kV) with automatic tube current modulation (ATCM) scanning protocol can be used for someone with BMI less than 24.9 kg/m2.
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2013年第3期318-322,共5页 Chinese Journal of Radiological Medicine and Protection
基金 广东省自然科学基金(8151008901000210)
关键词 多层螺旋CT 下肢动脉 血管造影术 低剂量 M SCT Lower extremity artery Angiography Low radiation dose
  • 相关文献

参考文献11

  • 1Met R, Bipat S, Legemate DA, et al. Diagnostic performance of computed tomography angiography in peripheral arterial disease : a systematic review and meta-analysis. JAMA, 2009, 301 ( 4 ) : 415-424.
  • 2Fraioli F, Catalano C, Napoli A, et al. Low-dose muhidetector- row CT angiography of the infra-renal aorta and lower extremity vessels: image quality and diagnostic accuracy in comparison with standard DSA. Eur Radiol,2006, 16( 1 ) : 137-146.
  • 3Wintersperger B, Jakobs T, Herzog P, et al. Aorto-iliac muhidetector-row CT angiography with low kV settings: improved vessel enhancement and simultaneous reduction of radiation dose. Eur Radiol, 2005, 15(2) : 334-341.
  • 4周淑琴,彭振鹏,周旭辉,严超贵,高樱,林少春,李树荣.64层螺旋CT下肢动脉血管造影:月国动脉监测触发对图像质量的影响[J].中国医学计算机成像杂志,2011,17(2):150-154. 被引量:13
  • 5Laswed T, Rizzo E, Guntern D, et al. Assessment of occlusive arterial disease of abdominal aorta and lower extremities arteries: value of multidetector CT angiography using an adaptive acquisition method. Eur Radiol, 2008, 18 (2) : 263-272.
  • 6Schemthaner R, Stadler A, Lomoschitz F, et al. Multi-detector CT angiography in the assessment of peripheral arterial occlusive disease: accuracy in detecting the severity, number, and length of stenoscs. Eur Radiol, 2008, 18(4) : 665-671.
  • 7Brenner D J, Hall EJ. Computed tomography--an increasing source of radiation exposure. N Engl J Med, 2007, 357 (22) :2277-2284.
  • 8Huda W, Scalzetti E, Levin G, et al. Technique factors and image quality as functions of patient weight at abdominal CT. Radiology, 2000, 217 (2) :430-435.
  • 9Kalra MK, Maher MM, Blake MA, et al. Multidetector CT scanning of abdomen and pelvis: a study for optimization of automatic tube current modulation technique in 120 subjects (abstr). Radiological Society of North America Scientific Assembly and Annual Meeting Program. Oak Brook : Radioi Soci North Am, 2003:294.
  • 10McCollough C, Primak AN, Braun N, et al. Strategies for reducing radiation dose in CT. Radiol Clin North Am, 2009,47 ( 1 ) :27-40.

二级参考文献13

  • 1Rubin GD, Schmidt AJ, Logan L, et al. Multidirector row CT angiography of low extremity arterial inflow and runoff: initial experience. Radiology,2001,221: 146- 158.
  • 2Chin AS, Rubin GD. CT angiography of peripheral arterial disease. Tech Vasc Interv Radiol, 2006,34: 143- 149.
  • 3Heijenbrok- Kal MH, Kock MC, Hunink MG. Lower extremity arterial disease: multidetector CT angiography meta analysis. Radiology, 2007, 245: 433 - 439.
  • 4Fleischmann D, Rubin GD. Quantification of intravenously administered contrast medium transit through the peripheral arteries: implications for CT angiography. Radiology, 2005, 236:1076 - 1082.
  • 5Fraioli F, Catalano C, Napoli A, et al. Low - dose multidetector - row CT angiography of the infra- renal aorta and lower extremity vessels: image quality and diagnostic accuracy in comparison with standard DSA. Eur Radiol, 2006,16:137 - 146.
  • 6Romano M, Mainenti P P, Imbriaco M, et al. Multidetector row CT angiography of the abdominal aorta and lower extremities in patients with peripheral arterial occlusive disease: diagnostic accuracy and inter observer agreement. Eur J Radiol, 2004, 50: 303- 308.
  • 7Nakaya Y, Kim T, Hori M, et al. 64- Slice multidetector row computed tomography angiography of aortoiliac and lower extremity arteries: efficacy of testing injection using a monitoring scan at knee level. J Comput Assist Tomogr, 2009,33: 20- 25.
  • 8Rieker O, Duber C, Neufang A, et al. CT Angiography versus intraarterial digital subtraction angiography for assessment of aortoiliac occlusive disease. AJR, 1997,169:1133 - 1138.
  • 9Lawrence JA, Kim D, Kent KC, et al. Lower extremity spiral CT angiography versus catheter angiography. Radiology, 1995, 194: 903- 908.
  • 10Qanadli SD, Chiappori V, Kelekis A, et al. Multislice computed tomography of peripheral arterial disease: new approach to optimize vascular opacification with 16- row platform. Eur Radiol, 2004,14(suppl 2) :b806.

共引文献12

同被引文献124

  • 1Patricia Carrascosa,Gastón A Rodriguez-Granillo,Carlos Capuay,Alejandro Deviggiano.Low-dose CT coronary angiography using iterative reconstruction with a 256-slice CT scanner[J].World Journal of Cardiology,2013,5(10):382-386. 被引量:2
  • 2刘少东,常春丽,李兰涛.16层螺旋CT下肢动脉造影扫描参数的优化[J].山东医药,2013,53(45):61-63. 被引量:1
  • 3仲海,邵广瑞,徐卓东,柳澄.CT血管成像在下肢动脉疾病中的应用价值[J].实用放射学杂志,2007,23(2):157-159. 被引量:30
  • 4张龙江,祁吉.对比剂肾病:一个值得关注的问题[J].中华放射学杂志,2007,41(8):882-884. 被引量:69
  • 5Fleischmann D, Rubin G D. Quantification of intravenously ad ministered contrast medium transit through the peripheral arteries: implications for CT angiography[J]. Radiology, 2005, 236 (3):1076-1082.
  • 6Nakaya Y,Kim T, Hori M,et al. 64-slice multidetector row com- puted tomographic angiography of aortoiliac and lower extremity arteries:efficacy of test injection using a monitoring scan at knee level[J].J Cornput Assist Tomogr,2009,33(1):20 -25.
  • 7Nakaya Y, Kim T, Hori M, et al. Correlations between aorto pop liteal bolus transit speed and aortic and popliteal bolus transit time during CT angiography of aortoiliac and lower extremity arteries [J]. Eur J Radiol,2011,79(2) :272-276.
  • 8Nakaura T,Awai K, Yauaga Y, et al. Contrast injection protocols for coronary computed tomography angiography using a 64-detec- tor scanner: comparison between patient weight adjusted-and fixed iodine-dose protoeols[J]. Invest Radiol,2008,43(7):512- 519.
  • 9Kim T,Murakami T,Takahashi S,et al. Effects of injection rates of contrast material on arterial phase hepatic CT[J]. AJR, 1998, 171(2) :429-432.
  • 10Albrecht T,Meyer B C. MDCT angiography of peripheral arter- ies: technical considerations and impact on patient managerrlent [J]. Eur Radiol,2007,17(Suppl 6) ,F5-F15.

引证文献14

二级引证文献85

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部