期刊文献+

无源效率刻度在活度测量中的应用研究 被引量:12

Application of Sourceless Efficiency Calibration in Activity Measurement
下载PDF
导出
摘要 利用几种典型的实验室标准源,对无源效率刻度软件LabSOCS效率模拟的准确性和可靠性进行了检验;从射线级联符合相加效应原理出发,探讨了利用具有级联关系的X射线进行效率刻度的可能性,扩展了效率刻度的能区范围;利用该软件模拟计算了一定体积的圆柱形气体体源效率随几何尺寸的变化关系,优化了圆柱形气体源盒的几何尺寸。 The accuracy and reliability of source - less efficiency calibration software are verified by comparing efficiency values between experimental calibration and simulation. From the principle of cascade coincidence, the efficiency calibration method by X - ray is discussed in this article. The efficiency calibration scope is ex- panded. The geometry of cylinder gas source vessel is optimized by simulating efficiencies as a function of the diameter for a fixed volume vessel.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2013年第5期568-571,共4页 Nuclear Electronics & Detection Technology
关键词 无源效率刻度 LabSOCS 高纯锗Γ谱仪 source - less efficiency calibration LabSOCS HPGe γSpectroscopy
  • 相关文献

参考文献11

  • 1Canberra. User’s manual of model S573 ISOCS Cali-bration Software[Z]. USA,2002.
  • 2Canberra. User,s manual of model S574 LabSOCSCalibration Software [ Z ]. US A,2002 .
  • 3R. Plenteda. A Monte Carlo Based Virtual GammaSpectroscopy Laboratory[D]. Austria: Institute of A-tomic and Subatomic Physics, 2004.
  • 4袁肇云,魏义彬,岑韬,王志远,吴绍云.ORTEC高纯锗伽玛谱仪系统的功能与特点[J].核电子学与探测技术,2007,27(3):611-616. 被引量:7
  • 5马吉增,刘立业,潘红娟,等.髙纯锗探测器无源效率刻度的方法:中国,CN101038261[P]. 2007 - 09.
  • 6王世联,王军,李奇,常印忠,樊元庆,陈占营,刘蜀疆.炭纤维底衬不锈钢气体测量源盒的研制[J].核电子学与探测技术,2010,30(1):27-30. 被引量:4
  • 7Comprehensive Nuclear - Test - Ban Treaty. Text ofthe treaty and its protocol [ R]. New York, USA: [ s.n. ] ,1996.
  • 8P. karhu,R. Clawson. Radionuclide laboratories sup-porting the network of radionuclide stations in verifica-tion of the Comprehensive Nuclear - Test - Ban Treaty[Jj.KERNTECHNIK, 2001, 66(3) :126-128.
  • 9J. P. Fontaine, F. Pointurier, X. Blanchard,et al.Atmospheric xenon radioactive isotope monitoring[J]. J. Environ. Radioact.,2004, 72: 129-135.
  • 10A. Ringbomf T. Larson, A. Axelsson, et al. SAUNA-a system for automatic sampling, processing, and a-nalysis of radioactive xenon [ J ] . Nucl. Instr. Meth.A ,2003 , 508 : 542-553.

二级参考文献8

  • 1朱文凯,陈军,熊建平,陈坤,陈益军.ISOCS系统无源效率刻度测量方法的准确性检验[J].核电子学与探测技术,2005,25(3):287-290. 被引量:34
  • 2P. karhu and R. Clawsor. Raclionuclide laboratories supporting the network of radionuclide stations in verification of the Comprehensive Nuclear-Test-Ban Treaty [J]. KERNTECHNIK, 2001, 66(3) : 126 - 128.
  • 3M. Matthews and J. Schulze The radionuelide monitoring system of the Comprehensive Nuclear- Test-Ban Treaty Organisation: Irom sample to product [J].KERNTECHNIK, 2001, 66 (3), 102 -112.
  • 4J.-P. Fontaine, F. Pointurier, X. Blanehard, T. Taffary. Atmospheric xenon radioactive isotope monitoring [J]. J. Environ.Radioact. , 2004, 72: 129 -135.
  • 5A. Ringbom, T. I.arson, A. Axelsson, K. Elmegren, C. Johansson. SAUNA- a system for automatic sampling, processing, and analysis of radioactive xenon[J]. Nucl. Instr. Meth., 2003, A508:542 -553.
  • 6R. B.Firestone. Table of Isotopes CD-ROM, Eighth Edition. New York: John Wiley and Sons, 1998.
  • 7F. Bronson. Validation of the accuracy of the LabSOCS software for mathematical efficiency calibration of Ge detectors for typical laboratory samples[J]. 2003, 255(1) : 137 -141.
  • 8Canberra User's Manual of Model S574 LabSOCS Calibration Software [M]. USA, 2002.