期刊文献+

一种X射线测厚校正方法的研究 被引量:1

Research of a Correction Method of Thickness Measurement Based on X-ray
下载PDF
导出
摘要 针对工业CT系统中所使用的X射线是一种多能谱射线,进而导致厚度测量不准确的现象,介绍了X射线测厚原理,分析了X射线在对物体进行厚度测量时发生射束硬化现象对测量结果的影响,推导出一种用于测厚的校正数学模型,并通过CD-300BXR型工业CT研究平台对该数学模型进行了验证。实验结果表明,当被测钢板厚度范围为5~50 mm时,利用该校正模型进行测厚的相对误差在±0.5%以内,能满足实际测厚需求。 The X-ray that used in industrial CT is a multi- energy ray, which leads to inaccurate thickness measurement. Considering this, the principle of X - ray thickness measurement and the effect of beam harden- ing phenomenon of X - ray are analyzed in the paper. Then, a mathematical model for correction is derived, and we verify the model through CD - 300 BXR research platform. The experimental results show that, the rel- ative error of thickness measurement based on the proposed model is less than 0.5 % when the measured steel plates range from 5 mm to 50 mm, which can satisfy the actual demand for thickness measurement.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2013年第5期621-623,646,共4页 Nuclear Electronics & Detection Technology
基金 中央高校基本科研业务费科研专项(0210005202035)
关键词 工业CT X射线 测厚 校正模型 实验 Industrial CT X - ray thickness measurement correction model experiment
  • 相关文献

参考文献4

  • 1黄悄音.X射线测厚仪的研制[D].北京:冶金自动化研究设计院,2009.
  • 2高上凯.医学成像系统[M].北京:淸华大学出版社,2003:13 -17.
  • 3王孝忠.核辐射物理[M].北京:原子能出版社,1986.
  • 4AganlV4 B K. X - Ray Spectroscopy(Springer Seriesin Optical Sciences) [ M]. 2nd ed. Berlin: Spriner,1993:36,144-150,201.